Medical Bionics - Research Publications

Permanent URI for this collection

Search Results

Now showing 1 - 10 of 14
  • Item
    Thumbnail Image
    Platinum dissolution and tissue response following long-term electrical stimulation at high charge densities
    Shepherd, RK ; Carter, PM ; Dalrymple, AN ; Enke, YL ; Wise, AK ; Nguyen, T ; Firth, J ; Thompson, A ; Fallon, JB (IOP Publishing Ltd, 2021-04)
    Objective. Established guidelines for safe levels of electrical stimulation for neural prostheses are based on a limited range of the stimulus parameters used clinically. Recent studies have reported particulate platinum (Pt) associated with long-term clinical use of these devices, highlighting the need for more carefully defined safety limits. We previously reported no adverse effects of Pt corrosion products in the cochleae of guinea pigs following 4 weeks of electrical stimulation using charge densities far greater than the published safe limits for cochlear implants. The present study examines the histopathological effects of Pt within the cochlea following continuous stimulation at a charge density well above the defined safe limits for periods up to 6 months.Approach. Six cats were bilaterally implanted with Pt electrode arrays and unilaterally stimulated using charge balanced current pulses at a charge density of 267μC cm-2phase-1using a tripolar electrode configuration. Electrochemical measurements were made throughout the implant duration and evoked potentials recorded at the outset and on completion of the stimulation program. Cochleae were examined histologically for particulate Pt, tissue response, and auditory nerve survival; electrodes were examined for surface corrosion; and cochlea, brain, kidney, and liver tissue analysed for trace levels of Pt.Main results. Chronic stimulation resulted in both a significant increase in tissue response and particulate Pt within the tissue capsule surrounding the electrode array compared with implanted, unstimulated control cochleae. Importantly, there was no stimulus-induced loss of auditory neurons (ANs) or increase in evoked potential thresholds. Stimulated electrodes were significantly more corroded compared with unstimulated electrodes. Trace analysis revealed Pt in both stimulated and control cochleae although significantly greater levels were detected within stimulated cochleae. There was no evidence of Pt in brain or liver; however, trace levels of Pt were recorded in the kidneys of two animals. Finally, increased charge storage capacity and charge injection limit reflected the more extensive electrode corrosion associated with stimulated electrodes.Significance. Long-term electrical stimulation of Pt electrodes at a charge density well above existing safety limits and nearly an order of magnitude higher than levels used clinically, does not adversely affect the AN population or reduce neural function, despite a stimulus-induced tissue response and the accumulation of Pt corrosion product. The mechanism resulting in Pt within the unstimulated cochlea is unclear, while the level of Pt observed systemically following stimulation at these very high charge densities does not appear to be of clinical significance.
  • Item
    Thumbnail Image
    An objective in vivo diagnostic method for inflammatory bowel disease
    Payne, SC ; Shepherd, RK ; Sedo, A ; Fallon, JB ; Furness, JB (ROYAL SOC, 2018-03)
    Inflammatory damage to the bowel, as occurs in inflammatory bowel disease (IBD), is debilitating to patients. In both patients and animal experimental models, histological analyses of biopsies and endoscopic examinations are used to evaluate the disease state. However, such measurements often have delays and are invasive, while endoscopy is not quantitatively objective. Therefore, a real-time quantitative method to assess compromised mucosal barrier function is advantageous. We investigated the correlation of in vivo changes in electrical transmural impedance with histological measures of inflammation. Four platinum (Pt) ball electrodes were placed in the lumen of the rat small intestine, with a return electrode under the skin. Electrodes placed within the non-inflamed intestine generated stable impedances during the 3 h testing period. Following an intraluminal injection of 2,4,6-trinitrobenzene sulfonic acid (TNBS), an established animal model of IBD, impedances in the inflamed region significantly decreased relative to a region not exposed to TNBS (p < 0.05). Changes in intestinal transmural impedance were correlated (p < 0.05) with histologically assessed damage to the mucosa and increases in neutrophil, eosinophil and T-cell populations at 3 h compared with tissue from control regions. This quantitative, real-time assay may have application in the diagnosis and clinical management of IBD.
  • Item
    Thumbnail Image
    Transmural impedance detects graded changes of inflammation in experimental colitis
    Payne, SC ; Alexandrovics, J ; Thomas, R ; Shepherd, RK ; Furness, JB ; Fallon, JB (ROYAL SOC, 2020-02-19)
    Ulcerative colitis is a chronic disease in which the mucosa of the colon or rectum becomes inflamed. An objective biomarker of inflammation will provide quantitative measures to support qualitative assessment during an endoscopic examination. Previous studies show that transmural electrical impedance is a quantifiable biomarker of inflammation. Here, we hypothesize that impedance detects spatially restricted areas of inflammation, thereby allowing the distinction between regions that differ in their severity of inflammation. A platinum ball electrode was placed into minimally inflamed (i.e. normal) or 2,4,6-trinitrobenzene sulphonic acid (TNBS)-inflamed colonic regions of rats and impedance measurements obtained by passing current between the intraluminal and subcutaneous return electrode. Histology of the colon was correlated with impedance measurements. The impedance of minimally inflamed (normal) tissue was 1.5-1.9 kΩ. Following TNBS injection, impedance significantly decreased within the inflammatory penumbra (p < 0.05), and decreased more in the inflammatory epicentre (p = 0.02). Histological damage correlated with impedance values (p < 0.05). Thus, impedance values of 1.5-1.9, 1.3-1.4 and 0.9-1.1 kΩ corresponded to minimally inflamed, mildly inflamed and moderately inflamed tissue, respectively. In conclusion, transmural impedance is an objective, spatially localized biomarker of mucosal integrity, and distinguishes between severities of intestinal inflammation.
  • Item
    Thumbnail Image
    Anti-inflammatory Effects of Abdominal Vagus Nerve Stimulation on Experimental Intestinal Inflammation
    Payne, SC ; Furness, JB ; Burns, O ; Sedo, A ; Flyakumurat, T ; Shepherd, RK ; Fallon, JB (FRONTIERS MEDIA SA, 2019-05-08)
    Electrical stimulation of the cervical vagus nerve is an emerging treatment for inflammatory bowel disease (IBD). However, side effects from cervical vagal nerve stimulation (VNS) are often reported by patients. Here we hypothesized that stimulating the vagus nerve closer to the end organ will have fewer off-target effects and will effectively reduce intestinal inflammation. Specifically, we aimed to: (i) compare off-target effects during abdominal and cervical VNS; (ii) verify that VNS levels were suprathreshold; and (iii) determine whether abdominal VNS reduces chemically-induced intestinal inflammation in rats. An electrode array was developed in-house to stimulate and record vagal neural responses. In a non-recovery experiment, stimulation-induced off-target effects were measured by implanting the cervical and abdominal vagus nerves of anaesthetized rats (n = 5) and recording changes to heart rate, respiration and blood pressure during stimulation (10 Hz; symmetric biphasic current pulse; 320 nC per phase). In a chronic experiment, the efficacy of VNS treatment was assessed by implanting an electrode array onto the abdominal vagus nerve and recording in vivo electrically-evoked neural responses during the implantation period. After 14 days, the intestine was inflamed with TNBS (2.5% 2,4,6-trinitrobenzene sulphonic acid) and rats received therapeutic VNS (n = 7; 10 Hz; 320 nC per phase; 3 h/day) or no stimulation (n = 8) for 4.5 days. Stool quality, plasma C-reactive protein and histology of the inflamed intestine were assessed. Data show that abdominal VNS had no effect (two-way RM-ANOVA: P ≥ 0.05) on cardiac, respiratory and blood pressure parameters. However, during cervical VNS heart rate decreased by 31 ± 9 beats/minute (P ≥ 0.05), respiration was inhibited and blood pressure decreased. Data addressing efficacy of VNS treatment show that electrically-evoked neural response thresholds remained stable (one-way RM ANOVA: P ≥ 0.05) and therapeutic stimulation remained above threshold. Chronically stimulated rats, compared to unstimulated rats, had improved stool quality (two-way RM ANOVA: P < 0.0001), no blood in feces (P < 0.0001), reduced plasma C-reactive protein (two-way RM ANOVA: P < 0.05) and a reduction in resident inflammatory cell populations within the intestine (Kruskal-Wallis: P < 0.05). In conclusion, abdominal VNS did not evoke off-target effects, is an effective treatment of TNBS-induced inflammation, and may be an effective treatment of IBD in humans.
  • Item
    Thumbnail Image
    A partial hearing animal model for chronic electro-acoustic stimulation
    Irving, S ; Wise, AK ; Millard, RE ; Shepherd, RK ; Fallon, JB (IOP PUBLISHING LTD, 2014-08)
    OBJECTIVE: Cochlear implants (CIs) have provided some auditory function to hundreds of thousands of people around the world. Although traditionally carried out only in profoundly deaf patients, the eligibility criteria for implantation have recently been relaxed to include many partially-deaf patients with useful levels of hearing. These patients receive both electrical stimulation from their implant and acoustic stimulation via their residual hearing (electro-acoustic stimulation; EAS) and perform very well. It is unclear how EAS improves speech perception over electrical stimulation alone, and little evidence exists about the nature of the interactions between electric and acoustic stimuli. Furthermore, clinical results suggest that some patients that undergo cochlear implantation lose some, if not all, of their residual hearing, reducing the advantages of EAS over electrical stimulation alone. A reliable animal model with clinically-relevant partial deafness combined with clinical CIs is important to enable these issues to be studied. This paper outlines such a model that has been successfully used in our laboratory. APPROACH: This paper outlines a battery of techniques used in our laboratory to generate, validate and examine an animal model of partial deafness and chronic CI use. MAIN RESULTS: Ototoxic deafening produced bilaterally symmetrical hearing thresholds in neonatal and adult animals. Electrical activation of the auditory system was confirmed, and all animals were chronically stimulated via adapted clinical CIs. Acoustic compound action potentials (CAPs) were obtained from partially-hearing cochleae, using the CI amplifier. Immunohistochemical analysis allows the effects of deafness and electrical stimulation on cell survival to be studied. SIGNIFICANCE: This animal model has applications in EAS research, including investigating the functional interactions between electric and acoustic stimulation, and the development of techniques to maintain residual hearing following cochlear implantation. The ability to record CAPs via the CI has clinical direct relevance for obtaining objective measures of residual hearing.
  • Item
    No Preview Available
    Evaluation of focused multipolar stimulation for cochlear implants in long-term deafened cats
    George, SS ; Wise, AK ; Fallon, JB ; Shepherd, RK (IOP PUBLISHING LTD, 2015-06)
    OBJECTIVE: Focused multipolar (FMP) stimulation has been shown to produce restricted neural activation using intracochlear stimulation in animals with a normal population of spiral ganglion neurons (SGNs). However, in a clinical setting, the widespread loss of SGNs and peripheral fibres following deafness is expected to influence the effectiveness of FMP. APPROACH: We compared the efficacy of FMP stimulation to both monopolar (MP) and tripolar (TP) stimulation in long-term deafened cat cochleae (n = 8). Unlike our previous study, these cochleae contained <10% of the normal SGN population adjacent to the electrode array. We also evaluated the effect of electrode position on stimulation modes by using either modiolar facing or lateral wall facing half-band electrodes. The spread of neural activity across the inferior colliculus, a major nucleus within the central auditory pathway, was used as a measure of spatial selectivity. MAIN RESULTS: In cochleae with significant SGN degeneration, we observed that FMP and TP stimulation resulted in greater spatial selectivity than MP stimulation (p < 0.001). However, thresholds were significantly higher for FMP and TP stimulation compared to MP stimulation (p < 0.001). No difference between FMP and TP stimulation was found in any measures. The high threshold levels for FMP stimulation was significantly reduced without compromising spatial selectivity by varying the degree of current focusing (referred as 'partial-FMP' stimulation). Spatial selectivity of all stimulation modes was unaffected by the electrode position. Finally, spatial selectivity in long-term deafened cochleae was significantly less than that of cochleae with normal SGN population (George S S et al 2014 J. Neural Eng. 11 065003). SIGNIFICANCE: The present results indicate that the greater spatial selectivity of FMP and TP stimulation over MP stimulation is maintained in cochleae with significant neural degeneration and is not adversely affected by electrode position. The greater spatial selectivity of FMP and TP stimulation would be expected to result in improved clinical performance.
  • Item
    Thumbnail Image
    Effects of chronic cochlear electrical stimulation after an extended period of profound deafness on primary auditory cortex organization in cats
    Fallon, JB ; Shepherd, RK ; Irvine, DRF (WILEY, 2014-03)
    Extended periods of deafness have profound effects on central auditory system function and organization. Neonatal deafening results in loss of the normal cochleotopic organization of the primary auditory cortex (AI), but environmentally-derived intracochlear electrical stimulation, via a cochlear implant, initiated shortly after deafening, can prevent this loss. We investigated whether such stimulation initiated after an extended period of deafness can restore cochleotopy. In two groups of neonatally-deafened cats, a multi-channel intracochlear electrode array was implanted at 8 weeks of age. One group received only minimal stimulation, associated with brief recordings at 4-6-week intervals, over the following 6 months to check the efficacy of the implant. In the other group, this 6-month period was followed by 6 months of near-continuous intracochlear electrical stimulation from a modified clinical cochlear implant system. We recorded multi-unit clusters in the auditory cortex and used two different methods to define the region of interest in the putative AI. There was no evidence of cochleotopy in any of the minimally stimulated animals, confirming our earlier finding. In three of six chronically stimulated cats there was clear evidence of AI cochleotopy, and in a fourth cat in which the majority of penetrations were in the anterior auditory field there was clear evidence of cochleotopy in that field. The finding that chronic intracochlear electrical stimulation after an extended period of deafness is able to restore cochleotopy in some (but not all) cases has implications for the performance of patients implanted after an extended period of deafness.
  • Item
    Thumbnail Image
    Effects of deafness and cochlear implant use on temporal response characteristics in cat primary auditory cortex
    Fallon, JB ; Shepherd, RK ; Nayagam, DAX ; Wise, AK ; Heifer, LF ; Landry, TG ; Irvine, DRF (ELSEVIER SCIENCE BV, 2014-09)
    We have previously shown that neonatal deafness of 7-13 months duration leads to loss of cochleotopy in the primary auditory cortex (AI) that can be reversed by cochlear implant use. Here we describe the effects of a similar duration of deafness and cochlear implant use on temporal processing. Specifically, we compared the temporal resolution of neurons in AI of young adult normal-hearing cats that were acutely deafened and implanted immediately prior to recording with that in three groups of neonatally deafened cats. One group of neonatally deafened cats received no chronic stimulation. The other two groups received up to 8 months of either low- or high-rate (50 or 500 pulses per second per electrode, respectively) stimulation from a clinical cochlear implant, initiated at 10 weeks of age. Deafness of 7-13 months duration had no effect on the duration of post-onset response suppression, latency, latency jitter, or the stimulus repetition rate at which units responded maximally (best repetition rate), but resulted in a statistically significant reduction in the ability of units to respond to every stimulus in a train (maximum following rate). None of the temporal response characteristics of the low-rate group differed from those in acutely deafened controls. In contrast, high-rate stimulation had diverse effects: it resulted in decreased suppression duration, longer latency and greater jitter relative to all other groups, and an increase in best repetition rate and cut-off rate relative to acutely deafened controls. The minimal effects of moderate-duration deafness on temporal processing in the present study are in contrast to its previously-reported pronounced effects on cochleotopy. Much longer periods of deafness have been reported to result in significant changes in temporal processing, in accord with the fact that duration of deafness is a major factor influencing outcome in human cochlear implantees.
  • Item
    Thumbnail Image
    Cochlear implantation for chronic electrical stimulation in the mouse
    Irving, S ; Trotter, MI ; Fallon, JB ; Millard, RE ; Shepherd, RK ; Wise, AK (ELSEVIER SCIENCE BV, 2013-12)
    The mouse is becoming an increasingly attractive model for auditory research due to the number of genetic deafness models available. These genetic models offer the researcher an array of congenital causes of hearing impairment, and are therefore of high clinical relevance. To date, the use of mice in cochlear implant research has not been possible due to the lack of an intracochlear electrode array and stimulator small enough for murine use, coupled with the difficulty of the surgery in this species. Here, we present a fully-implantable intracochlear electrode stimulator assembly designed for chronic implantation in the mouse. We describe the surgical approach for implantation, as well as presenting the first functional data obtained from intracochlear electrical stimulation in the mouse.
  • Item
    Thumbnail Image
    Chronic neurotrophin delivery promotes ectopic neurite growth from the spiral ganglion of deafened cochleae without compromising the spatial selectivity of cochlear implants
    Landry, TG ; Fallon, JB ; Wise, AK ; Shepherd, RK (WILEY, 2013)
    Cochlear implants restore hearing cues in the severe-profoundly deaf by electrically stimulating spiral ganglion neurons (SGNs). However, SGNs degenerate following loss of cochlear hair cells, due at least in part to a reduction in the endogenous neurotrophin (NT) supply, normally provided by hair cells and supporting cells of the organ of Corti. Delivering exogenous NTs to the cochlea can rescue SGNs from degeneration and can also promote the ectopic growth of SGN neurites. This resprouting may disrupt the cochleotopic organization upon which cochlear implants rely to impart pitch cues. Using retrograde labeling and confocal imaging of SGNs, we determined the extent of neurite growth following 28 days of exogenous NT treatment in deafened guinea pigs with and without chronic electrical stimulation (ES). On completion of this treatment, we measured the spread of neural activation to intracochlear ES by recording neural responses across the cochleotopically organized inferior colliculus using multichannel recording techniques. Although NT treatment significantly increased both the length and the lateral extent of growth of neurites along the cochlea compared with deafened controls, these anatomical changes did not affect the spread of neural activation when examined immediately after 28 days of NT treatment. NT treatment did, however, result in lower excitation thresholds compared with deafened controls. These data support the application of NTs for improved clinical outcomes for cochlear implant patients.