Biomedical Engineering - Research Publications

Permanent URI for this collection

Search Results

Now showing 1 - 10 of 16
  • Item
    Thumbnail Image
    Subject-specific musculoskeletal modeling in the evaluation of shoulder muscle and joint function
    Wu, W ; Lee, PVS ; Bryant, AL ; Galea, M ; Ackland, DC (ELSEVIER SCI LTD, 2016-11-07)
    Upper limb muscle force estimation using Hill-type muscle models depends on musculotendon parameter values, which cannot be readily measured non-invasively. Generic and scaled-generic parameters may be quickly and easily employed, but these approaches do not account for an individual subject's joint torque capacity. The objective of the present study was to develop a subject-specific experimental testing and modeling framework to evaluate shoulder muscle and joint function during activities of daily living, and to assess the capacity of generic and scaled-generic musculotendon parameters to predict muscle and joint function. Three-dimensional musculoskeletal models of the shoulders of 6 healthy subjects were developed to calculate muscle and glenohumeral joint loading during abduction, flexion, horizontal flexion, nose touching and reaching using subject-specific, scaled-generic and generic musculotendon parameters. Muscle and glenohumeral joint forces calculated using generic and scaled-generic models were significantly different to those of subject-specific models (p<0.05), and task dependent; however, scaled-generic model calculations of shoulder glenohumeral joint force demonstrated better agreement with those of subject-specific models during abduction and flexion. Muscles in generic musculoskeletal models operated further from the plateau of their force-length curves than those of scaled-generic and subject-specific models, while muscles in subject-specific models operated over a wider region of their force length curves than those of the generic or scaled-generic models, reflecting diversity of subject shoulder strength. The findings of this study suggest that generic and scaled-generic musculotendon parameters may not provide sufficient accuracy in prediction of shoulder muscle and joint loading when compared to models that employ subject-specific parameter-estimation approaches.
  • Item
    Thumbnail Image
    Unicortical and bicortical plating in the fixation of comminuted fractures of the clavicle: a biomechanical study
    Looft, JM ; Correa, L ; Patel, M ; Rawlings, M ; Ackland, DC (WILEY, 2017-11)
    BACKGROUND: Intraoperative neurovascular complications with clavicle fracture fixation are often due to far cortex penetration by drills and screws, but could be avoided using a unicortical construct. The objective of this study was to compare the bending and torsional strength of a unicortical locking screw plate construct and a hybrid (with central locked and outer non-locked long oblique screws) unicortical plate construct for clavicle fracture fixation with that of a conventional bicortical locking screw construct of plate fixation. METHODS: Twenty-four human clavicle specimens were harvested and fractured in a comminuted mid-shaft butterfly configuration. Clavicles were randomly allocated to three surgical fixation groups: unicortical locking screw, bicortical locking screw and hybrid unicortical screw fixation. Clavicles were tested in torsion and cantilever bending. Construct bending and torsional stiffness were measured, as well as ultimate strength in bending. RESULTS: There were no significant differences in bending stiffness or ultimate bending moment between all three plating techniques. The unicortical locked construct had similar torsional stiffness compared with the bicortical locked construct; however, the hybrid technique was found to have significantly lower torsional stiffness to that of the bicortical locking screw construct (mean difference: 87.5 Nmm/degree, P = 0.028). CONCLUSIONS: Unicortical locked screw plate fixation and hybrid unicortical plating fixation with centrally locked screws and outer long, oblique screws may alleviate far cortex penetration, protecting nearby anatomical structures, and may ease implant removal and conversion to bicortical fixation for revision surgery; however, use of long oblique screws may increase the risk of early loosening under torsion.
  • Item
    Thumbnail Image
    Microstructure Variations in the Soft-Hard Tissue Junction of the Human Anterior Cruciate Ligament
    Zhao, L ; Lee, PVS ; Ackland, DC ; Broom, ND ; Thambyah, A (WILEY, 2017-09)
  • Item
    Thumbnail Image
    Load response and gap formation in a single-row cruciate suture rotator cuff repair
    Huntington, L ; Richardson, M ; Sobol, T ; Caldow, J ; Ackland, DC (WILEY, 2017-06)
    BACKGROUND: Double-row rotator cuff tendon repair techniques may provide superior contact area and strength compared with single-row repairs, but are associated with higher material expenses and prolonged operating time. The purpose of this study was to evaluate gap formation, ultimate tensile strength and stiffness of a single-row cruciate suture rotator cuff repair construct, and to compare these results with those of the Mason-Allen and SutureBridge repair constructs. METHODS: Infraspinatus tendons from 24 spring lamb shoulders were harvested and allocated to cruciate suture, Mason-Allen and SutureBridge repair groups. Specimens were loaded cyclically between 10 and 62 N for 200 cycles, and gap formation simultaneously measured using a high-speed digital camera. Specimens were then loaded in uniaxial tension to failure, and construct stiffness and repair strength were evaluated. RESULTS: Gap formation in the cruciate suture repair was significantly lower than that of the Mason-Allen repair (mean difference = 0.6 mm, P = 0.009) and no different from that of the SutureBridge repair (P > 0.05). Both the cruciate suture repair (mean difference = 15.7 N/mm, P = 0.002) and SutureBridge repair (mean difference = 15.8 N/mm, P = 0.034) were significantly stiffer than that of the Mason-Allen repair; however, no significant differences in ultimate tensile strength between repair groups were discerned (P > 0.05). CONCLUSION: The cruciate suture repair construct, which may represent a simple and cost-effective alternative to double-row and double-row equivalent rotator cuff repairs, has comparable biomechanical strength and integrity with that of the SutureBridge repair, and may result in improved construct longevity and tendon healing compared with the Mason-Allen repair.
  • Item
    Thumbnail Image
    Biomechanical performance of an intramedullary Echidna pin for fixation of comminuted mid-shaft clavicle fractures
    Sidhu, N ; Huntington, LS ; Richardson, M ; Ackland, DC (Wiley, 2019-10-01)
    Surgical fixation of comminuted mid‐shaft clavicle fractures commonly employs intramedullary devices; however, pins with smooth surfaces are prone to migration, whilst threaded pins can be challenging to remove post‐operatively. The aim of this study was to evaluate the biomechanical performance of fractured clavicles repaired using a novel intramedullary Echidna pin device and a non‐threaded Knowles pin. The Echidna pin features retractable spines that engage with the bone to minimize migration and facilitate ease of device removal.
  • Item
    Thumbnail Image
    Muscle and Joint Function After Anatomic and Reverse Total Shoulder Arthroplasty Using a Modular Shoulder Prosthesis
    Ackland, DC ; Wu, W ; Thomas, R ; Patel, M ; Page, R ; Sangeux, M ; Richardson, M (John Wiley & Sons, Inc., 2019-09-01)
    Changes in joint architecture and muscle loading resulting from total shoulder arthroplasty (TSA) and reverse total shoulder arthroplasty (RSA) are known to influence joint stability and prosthesis survivorship. This study aimed to measure changes in muscle moment arms, muscle lines of action, as well as muscle and joint loading following TSA and RSA using a metal‐backed uncemented modular shoulder prosthesis. Eight cadaveric upper extremities were assessed using a customized testing rig. Abduction, flexion, and axial rotation muscle moment arms were quantified using the tendon‐excursion method, and muscle line‐of‐force directions evaluated radiographically pre‐operatively, and after TSA and revision RSA. Specimen‐specific musculoskeletal models were used to estimate muscle and joint loading pre‐ and post‐operatively. TSA lateralized the glenohumeral joint center by 4.3 ± 3.2 mm, resulting in small but significant increases in middle deltoid force (2.0%BW) and joint compression during flexion (2.1%BW) (p < 0.05). Revision RSA significantly increased the moment arms of the major abductors, flexors, adductors, and extensors, and reduced their peak forces (p < 0.05). The superior inclination of the deltoid significantly increased while the inferior inclination of the rotator cuff muscles decreased (p < 0.05). TSA using an uncemented metal‐backed modular shoulder prosthesis effectively restores native joint function; however, lateralization of the glenoid component should be minimized intra‐operatively to mitigate increased glenohumeral joint loading and polyethylene liner contact stresses. Revision RSA reduces muscle forces required during shoulder function but produces greater superior joint shear force and less joint compression. The findings may help to guide component selection and placement to mitigate joint instability after arthroplasty.
  • Item
    Thumbnail Image
    The moment arms of the muscles spanning the glenohumeral joint: a systematic review
    Hik, F ; Ackland, DC (WILEY, 2019-01)
    The moment arm of a muscle represents its leverage or torque-producing capacity, and is indicative of the role of the muscle in joint actuation. The objective of this study was to undertake a systematic review of the moment arms of the major muscles spanning the glenohumeral joint during abduction, flexion and axial rotation. Moment arm data for the deltoid, pectoralis major, latissimus dorsi, teres major, supraspinatus, infraspinatus, subscapularis and teres minor were reported when measured using the geometric and tendon excursion methods. The anterior and middle sub-regions of the deltoid had the largest humeral elevator moment arm values of all muscles during coronal- and scapular-plane abduction, as well as during flexion. The pectoralis major, latissimus dorsi and teres major had the largest depressor moment arms, with each of these muscles exhibiting prominent leverage in shoulder adduction, and the latissimus dorsi and teres major also in extension. The rotator cuff muscles had the largest axial rotation moment arms regardless of the axial position of the humerus. The supraspinatus had the most prominent elevator moment arms during early abduction in both the coronal and scapular planes as well as in flexion. This systematic review shows that the rotator cuff muscles function as humeral rotators and weak humeral depressors or elevators, while the three sub-regions of the deltoid behave as substantial humeral elevators throughout the range of humeral motion. The pectoralis major, latissimus dorsi and teres major are significant shoulder depressors, particularly during abduction. This study provides muscle moment arm data on functionally relevant shoulder movements that are involved in tasks of daily living, including lifting and pushing. The results may be useful in quantifying shoulder muscle function during specific planes of movement, in designing and validating computational models of the shoulder, and in planning surgical procedures such as tendon transfer surgery.
  • Item
    Thumbnail Image
    The influence of rotator cuff tears on muscle and joint-contact loading after reverse total shoulder arthroplasty
    Ackland, DC ; Robinson, DL ; Wilkosz, A ; Wu, W ; Richardson, M ; Lee, P ; Tse, KM (WILEY, 2019-01)
  • Item
    Thumbnail Image
    Prosthesis design and placement in reverse total shoulder arthroplasty
    Ackland, DC ; Patel, M ; Knox, D (BIOMED CENTRAL LTD, 2015-07-02)
    The management of irreparable rotator cuff tears associated with osteoarthritis of the glenohumeral joint has long been challenging. Reverse total shoulder arthroplasty (RSA) was designed to provide pain relief and improve shoulder function in patients with severe rotator cuff tear arthropathy. While this procedure has been known to reduce pain, improve strength and increase range of motion in shoulder elevation, scapular notching, rotation deficiency, early implant loosening and dislocation have attributed to complication rates as high as 62%. Patient selection, surgical approach and post-operative management are factors vital to successful outcome of RSA, with implant design and component positioning having a significant influence on the ability of the shoulder muscles to elevate, axially rotate and stabilise the humerus. Clinical and biomechanical studies have revealed that component design and placement affects the location of the joint centre of rotation and therefore the force-generating capacity of the muscles and overall joint mobility and stability. Furthermore, surgical technique has also been shown to have an important influence on clinical outcome of RSA, as it can affect intra-operative joint exposure as well as post-operative muscle function. This review discusses the behaviour of the shoulder after RSA and the influence of implant design, component positioning and surgical technique on post-operative joint function and clinical outcome.
  • Item
    Thumbnail Image
    An intramedullary Echidna pin for fixation of comminuted clavicle fractures: a biomechanical study
    Ackland, D ; Griggs, I ; Hislop, P ; Wu, W ; Patel, M ; Richardson, M (BMC, 2017-08-11)
    BACKGROUND: Intramedullary fixation of comminuted mid-shaft clavicle fractures has traditionally been employed with satisfactory clinical outcomes; however, pins with smooth surfaces may protrude from the bone and are prone to migration, while some threaded pins are difficult to remove post-operatively. The aim of this proof-of-concept study was to develop and evaluate the biomechanical strength of a novel intramedullary Echidna pin device designed to maintain fracture reduction, resist migration and facilitate ease of post-operative removal. METHODS: Thirty human clavicle specimens were harvested and fractured in a comminuted mid-shaft butterfly configuration. Each specimen was randomly allocated to three surgical repair groups including intramedullary fixation using the Echidna pin and Herbert Cannulated Bone Screw System, as well as plate fixation using bi-cortical locking screws. Using a biomechanical testing apparatus, construct bending and torsional stiffness were measured, as well as ultimate bending strength. RESULTS: There was no significant difference in torsional stiffness and ultimate bending moment between the Echidna pin and Herbert screw repair constructs (p > 0.05); however, the Echidna pin construct demonstrated a significantly greater bending stiffness compared to that of the Herbert screw construct (mean difference 0.55 Nm/deg., p = 0.001). The plate construct demonstrated significantly greater torsional stiffness, bending stiffness and ultimate bending moment compared to those of the Herbert screw and Echidna pin (p < 0.05). CONCLUSIONS: An intramedullary Echidna pin device was designed to stabilize comminuted fractures of the clavicle, maintain fracture compression and provide ease of removal post-operatively. Since the results suggest equivalent or superior torsional and bending stability in the Echidna pin compared to that of the Herbert screw, the Echidna pin concept may represent an alternative fixation device to conventional intramedullary screws, nails and pins; however, superior plating using bi-cortical locking screws provides substantially higher construct structural rigidity than intramedullary devices, and may therefore be useful in cases of osteoporotic bone, or where high fracture stability is required.