Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 3 of 3
  • Item
    Thumbnail Image
    Insertion study using new peri-modiolar electrode array designs [Abstract]
    Treaba, Claudiu ; Clark, Graeme M. ; Cowan, Robert S. ; Tykocinski, Michael J. ; Cohen, Lawrence T. ; Saunders, Elaine ; Pyman, Brian C. ; Briggs, Robert S. ; Dahm, Markus C. ( 1999)
    Intracochlear multi-channel cochlear implants have been shown to successfully provide auditory information for profoundly deaf patients by electrically stimulating discrete populations of auditory nerve fibers via a scala tympani (ST) electrode array. Histological and radiological examination of implanted human temporal bones showed that the current straight Nucleus® array is usually positioned against the outer wall of the ST. An electrode array close to the modiolus could be expected to reduce stimulation thresholds and result in a more localized neural excitation pattern.
  • Item
    Thumbnail Image
    The development of a precurved cochlear implant electrode array and its preliminary psychophysical evaluation [Abstract]
    Cohen, L. T. ; Saunders, E. ; Treaba, C. ; Pyman, B. C. ; Clark, Graeme M. ( 1998)
    A precurved banded electrode array may provide a better interface with the auditory neural pathways for cochlear implants, and provide better speech perception. A prototype arrray has been further evaluated for ease of insertion, siting within the cochlea and induction of any cochlear trauma. The arrays were inserted into the human cochlear under simulated surgery. The bones were embedded in Araldite, X-rayed and sectioned. X-ray reconstruction analyses of the position of the implanted array showed its insertion to be favourable. Cochlear implants with precurved arrays have been implanted in three patients. Psychophysical evaluation and X-ray analyses have shown that as electrode distance from the modiolus decreased: threshold current decreased; dynamic range increased; current spread as measured by forward masking studies, was more focused; electrode discrimination with loudness jitter (being abetter representation of the dynamic speech signal) improved; JNDs for loudness, expressed as a function of dynamic range, decreased.
  • Item
    Thumbnail Image
    Histopathology following electrode insertion and chronic electrical stimulation
    Shepherd, R. K. ; Clark, Graeme M. ; Pyman, B. C. ; Webb, R. L. ; Murray, M. T. ; Houghton, M. E. (Raven Press, 1985)
    We have examined a number of safety issues associated with cochlear implants. This work has been primarily designed to evaluate the histopathological effects of intracochlear electrode implantation and chronic electrical stimulation. The results of these studies may be summarized as follows: 1) The insertion of the banded free-fit scala tympani array into human cadaver temporal bones produces minimal damage, occurring primarily to a localized region of the spiral ligament. This damage would not result in significant neural degeneration and thus, would not compromise the efficacy of the multiple channel device; 2) chronic intracochlear electrical stimulation for continuous periods of 500 to 2000 hours, using charge balanced biphasic current pulses developing charge densities of 18-32 }?C/cm2. geom./phase, does not adversely affect the spiral ganglion cell population; 3) labyrinthine infection severely reduces the viable spiral ganglion cell population; 4) the formation of new bone present in approximately half of the animals we have implanted --is not associated with electrical stimulation per se; 5) scanning electron microscope studies of electrodes subjected to long periods of intracochlear electrical stimulation reveals minimal platinum dissolution when compared with unstimulated control electrodes, and electrodes that have been stimulated for similar periods in inorganic saline.