    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve using non-charge-balanced stimuli
    Shepherd, Robert K. ; Linahan, N. ; Xu, J. ; Clark, Graeme M. ; Araki, S. ( 1999)
    This study was designed to evaluate the pathophysiological response of the cochlea following long-term intracochlear electrical stimulation using a poorly charge-balanced stimulus regime, leading to direct current (DC) levels >0.1 µA. Four normal-hearing adult cats were bilaterally implanted with scala tympani electrode arrays and unilaterally stimulated for periods up to 2200 h. Stimuli consisted of 50 µs monophasic current pulses presented at 2000 pulses per second (pps) per channel, and resulted in DC levels of 0.4-2.8 µA. Both acoustic and electrical (EABR) evoked potentials were periodically recorded during the stimulation program. Frequency-specific stimuli indicated that an extensive and widespread hearing loss occurred over the 4-24 KHz region in all stimulated cochleae, although the 2 KHz region exhibited thresholds close to normal in some animals, despite long-term implantation and chronic stimulation. Longitudinal EABRs showed a statistically significant increase in threshold for three of the four animals. Histopathological evaluation of the cochleae revealed a highly significant reduction in ganglion cell density in stimulated cochleae compared with their controls. Spiral ganglion cell loss was significantly correlated with the degree of inflammation, duration of electrical stimulation, and the level of DC. In conclusion, the present study highlights the potential for neural damage following stimulation using poorly charge-balanced stimuli.
  • Item
    Thumbnail Image
    Cochlear implantation in young children: histological studies on head growth, leadwire design, and electrode fixation in the monkey model
    Burton, M. J. ; Shepherd, R. K. ; Xu, S. A. ; Xu, J. ; Franz, B. K-H. G. ; Clark, Graeme M. ( 1994)
    For safe cochlear implantation in children under 2 years of age, the implant assembly must not adversely affect adjacent tissues or compromise head growth. Furthermore, growth changes and tissue responses should not impair the function of the device. Dummy receiver-stimulators, interconnect plugs, and leadwire-lengthening systems were implanted for periods of 36 months in the young monkey to effectively model the implantation of the young child. The results show that implanting a receiver-stimulator package has no adverse effects on skull growth or the underlying central nervous system. The system for fixing the electrode at the fossa incudis proved effective. There was marked osteoneogenesis in the mastoid cavity, resulting in the fixation of the leadwire outside the cochlea. This study provides evidence for the safety of cochlear implantation in young subjects.
  • Item
    Thumbnail Image
    Paediatric cochlear implantation: radiologic observations of skull growth
    XU, JIN ; Shepherd, Robert K. ; Xu, Shi-Ang ; Seldon, H. Lee ; Clark, Graeme M. ( 1993)
    We investigated the effects of long-term implantation of auditory prostheses on skull growth in young animals. Four monkeys were implanted with dummy cochlear implants at 6 months of age. To simulate implantation in children, the bed for the receiver-stimulator or interconnecting plug was drilled across a calvarial suture down to the underlying dura. Plain skull oentgenograms were periodically taken to monitor head growth for up to 3 years after implantation. These longitudinal measurements revealed no significant asymmetric skull growth. Postmortem measurements using computed tomographic scans confirmed these results and showed no significant difference in the intracranial volumes between the implanted and control sides of each animal or between experimental and nonimplanted control monkeys. These results suggest that long-term cochlear implantation in very young children will not cause any significant deformity of the skull.