Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 4 of 4
  • Item
  • Item
    Thumbnail Image
    Effects of chronic electrical stimulation on spiral ganglion neuron survival and size in deafened kittens
    Araki, Susumu ; Kawano, Atsushi ; Seldon, H. Lee ; Shepherd, Robert K. ; Funasaka, Sotaro ; Clark, Graeme M. ( 1998)
    We have studied spiral ganglion cell (SGC) survival and soma size in neonatally pharmacologically deafened kittens. They were implanted with a four-electrode array in the left cochlea at 100 to 180 or more days of age. Eight animals were chronically stimulated approximately 1000 hours over approximately 60 days with charge-balanced, biphasic current pulses; three were unstimulated controls. Using three-dimensional computer-aided reconstruction of the cochlea, the SGC position and cross-sectional area were stored. SGC position was mapped to the organ of Corti by perpendicular projections, starting from the basal end. The basal region of the cochlea was divided into three 4-mm segments. SGC survival (number per 0.1 mm of the length of the organ of Corti) and soma size for stimulated cochleae were compared statistically with implanted but unstimulated cochleae. There was no evidence of an effect of electrical stimulation on SGC survival under this protocol and with this duration. On the other hand, the cell size on the stimulated side was significantly larger than the control side in the middle segment (4 to 8 mm from the basal end). SGCs undergo a reduction in size after prolonged auditory deprivation; however, these changes may be partially moderated after chronic intracochlear electrical stimulation.
  • Item
    Thumbnail Image
    Cochlear implants in children: the value of cochleostomy seals in the prevention of labyrinthitis following pneumococcal otitis media
    Dahm, M. C. ; Webb, R. L. ; Clark, Graeme M. ; Franz, B. K-H. ; Shepherd, R. K. ; Burton, M. J. ; ROBINS-BROWNE, R. ( 1995)
    Cochlea implantation at an early age is important in rehabilitating profoundly hearing impaired children. Given the incidence of pneumococcal otitis media in young children, there has been concern that cochlear implantation could increase the possibility of otitis media, leading to labyrinthitis in this age group. Clinical experience has not indicated an increase in the frequency of otitis media and labyrinthitis in implanted adults or children over two years. However, labyrinthitis has occurred in implanted animals with otitis media. In order to assess the impact of cochlear implants on the occurrence of labyrinthitis, pneumococcal otitis media was induced in 21 kittens. Thirty-two kitten cochleas were implanted, of which 9 had a fascial graft and 9 a Gelfoam® graft. Nine control cochleas were unimplanted. Labyrinthitis occurred in 44% of unimplanted controls. 50% of implanted ungrafted cochleas, and 6% of implanted grafted cochleas. There was no statistically significant difference between the incidence of labyrinthitis in the implanted cochleas and the unimplanted controls. However there was a statistically significant difference between the ungrafted and grafted cochleas, but not between the two types of graft.
  • Item
    Thumbnail Image
    Cochlear implantation in young children: histological studies on head growth, leadwire design, and electrode fixation in the monkey model
    Burton, M. J. ; Shepherd, R. K. ; Xu, S. A. ; Xu, J. ; Franz, B. K-H. G. ; Clark, Graeme M. ( 1994)
    For safe cochlear implantation in children under 2 years of age, the implant assembly must not adversely affect adjacent tissues or compromise head growth. Furthermore, growth changes and tissue responses should not impair the function of the device. Dummy receiver-stimulators, interconnect plugs, and leadwire-lengthening systems were implanted for periods of 36 months in the young monkey to effectively model the implantation of the young child. The results show that implanting a receiver-stimulator package has no adverse effects on skull growth or the underlying central nervous system. The system for fixing the electrode at the fossa incudis proved effective. There was marked osteoneogenesis in the mastoid cavity, resulting in the fixation of the leadwire outside the cochlea. This study provides evidence for the safety of cochlear implantation in young subjects.