Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 4 of 4
  • Item
    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve at high stimulus rates: a physiological and histopathological study
    XU, JIN ; Shepherd, Robert K. ; Millard, Rodney E. ; Clark, Graeme M. ( 1997)
    A major factor associated with recent improvements in the clinical performance of cochlear implant patients has been the development of speech-processing strategies based on high stimulation rates. While these processing strategies show clear clinical advantage, we know little of their long-term safety implications. The present study was designed to evaluate the physiological and histopathological effects of long-term intracochlear electrical stimulation using these high rates. Thirteen normal-hearing adult cats were bilaterally implanted with scala tympani electrode arrays and unilaterally stimulated for periods of up to 2100 h using either two pairs of bipolar or three monopolar stimulating electrodes. Stimuli consisted of short duration (25-50 µs/phase) charge-balanced biphasic current pulses presented at 1000 pulses per second (pps) per channel for monopolar stimulation, and 2000 pps/channel for bipolar stimulation. The electrodes were shorted between current pulses to minimize any residual direct current, and the pulse trains were presented using a 50% duty cycle (500 ms on; 500 ms oft) in order to simulate speech. Both acoustic (ABR) and electrical (EABR) auditory brainstem responses were recorded periodically during the chronic stimulation program, All cochleas showed an increase in the click-evoked ABR threshold following implant surgery; however, recovery to near-normal levels occurred in approximately half of the stimulated cochleas 1 month post-operatively. The use of frequency-specific stimuli indicated that the most extensive hearing loss generally occurred in the high-frequency basal region of the cochlea (12 and 24 kHz) adjacent to the stimulating electrode. However, thresholds at lower frequencies (2, 4 and 8 kHz), appeared at near-normal levels despite long-term electrode implantation and electrical stimulation. Our longitudinal EABR results showed a statistically significant increase in threshold in nearly 40% of the chronically stimulated electrodes evaluated; however, the gradient of the EABR input/output (I/O) function (evoked potential response amplitude versus stimulus current) generally remained quite stable throughout the chronic stimulation period. Histopathological examination of the cochleas showed no statistically significant difference in ganglion cell densities between cochleas using monopolar and bipolar electrode configurations (P = 0.67), and no evidence of cochlear damage caused by high-rate electrical stimulation when compared with control cochleas. Indeed, there was no statistically significant relationship between spiral ganglion cell density and electrical stimulation (P = 0.459), or between the extent of loss of inner (IHC, P = 0.86) or outer (OHC, P=0.30) hair cells and electrical stimulation. Spiral ganglion cell loss was, however, influenced by the degree of inflammation (P=0.016) and electrode insertion trauma. These histopathological findings were consistent with the physiological data. Finally, electrode impedance, measured at completion of the chronic stimulation program, showed close correlation with the degree of tissue response adjacent to the electrode array. These results indicated that chronic intracochlear electrical stimulation, using carefully controlled charge-balanced biphasic current pulses at stimulus rates of up to 2000 pps/channel, does not appear to adversely affect residual auditory nerve elements or the cochlea in general. This study provides an important basis for the safe application of improved speech-processing strategies based on high-rate electrical stimulation.
  • Item
    Thumbnail Image
    Cochlear implantation in young children: histological studies on head growth, leadwire design, and electrode fixation in the monkey model
    Burton, M. J. ; Shepherd, R. K. ; Xu, S. A. ; Xu, J. ; Franz, B. K-H. G. ; Clark, Graeme M. ( 1994)
    For safe cochlear implantation in children under 2 years of age, the implant assembly must not adversely affect adjacent tissues or compromise head growth. Furthermore, growth changes and tissue responses should not impair the function of the device. Dummy receiver-stimulators, interconnect plugs, and leadwire-lengthening systems were implanted for periods of 36 months in the young monkey to effectively model the implantation of the young child. The results show that implanting a receiver-stimulator package has no adverse effects on skull growth or the underlying central nervous system. The system for fixing the electrode at the fossa incudis proved effective. There was marked osteoneogenesis in the mastoid cavity, resulting in the fixation of the leadwire outside the cochlea. This study provides evidence for the safety of cochlear implantation in young subjects.
  • Item
    Thumbnail Image
    Comparison of half-band and full-band electrodes for intracochlear electrical stimulation
    Xu, Shi-Ang ; McAnally, Ken I. ; Xu, Jin. ; Clark, Graeme M. ( 1993)
    It was hypothesized that intracochlear stimulating electrodes oriented toward the modiolus would require a lower stimulus current to elicit a threshold neural response than longitudinal band electrodes. Electrically evoked auditory brain stem responses (EABRs) and electrode impedances were recorded with full-band and oriented half-band scala tympani electrodes in anesthetized, deafened cats. To elicit a threshold EABR, the stimulus current required for stimulation through half-band electrodes oriented toward the modiolus was not significantly different from the current required for stimulation through full-band electrodes. The impedances of full-band electrodes were significantly lower than those of half-band electrodes. Considering the significantly higher impedance and current density of half band electrodes in comparison to full-band electrodes, as well as the critical orientation of half-band electrodes during implantation, we believe that full-band electrodes have advantages over half-band electrodes for scala tympani implantation.
  • Item
    Thumbnail Image
    Radiologic evaluation of multichannel intracochlear implant insertion depth
    Marsh, Michael A. ; XU, JIN ; Blamey, Peter J. ; Whitford, Lesley A. ; Xu, Shi-Ang ; Silverman, Julianna M. ; Clark, Graeme, M. ( 1993)
    Postoperative plain film x-rays are necessary in all multichannel cochlear implant patients to confirm intracochlear position, detect possible electrode kinking, and provide a reference if postoperative slippage occurs. In addition, precise documentation of multichannel intracochlear electrode insertion depths is necessary for comparison of speech recognition results among patients and may be of use for future speech processing strategies. In the present study, a method has been devised, using a modified Stenver's view, to more accurately document insertion depths of the electrode array and location of individual electrodes on 50 multichannel cochlear implant patients. Surgical estimates of insertion depth are shown to have great variability in regard to distance along the basilar membrane when compared with x-ray documentation. Additionally, there is preliminary evidence that insertion depth, as determined by x-ray studies, has a strong correlation with open-set speech discrimination.