Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 10 of 10
  • Item
    Thumbnail Image
    A speech processing strategy for multiple-electrode cochlear implant prostheses
    Tong, Y. C. ; Clark, Graeme M. (Monash University Press, 1983)
    Speech studies in a number of research centres have shown that useful speech information could be presented to deaf patients using single or multiple electrode cochlear implant prostheses (Parking & Anderson, 1983). In our laboratory, speech processing strategies were formulated on the basis of psychophysical results. This paper examines the psychophysical characteristics of the hearing sensations produced by electrical stimulation using scala tympani electrodes in postlingually deaf patients; a speech processing strategy is then discussed on the basis of these characteristics.
  • Item
    Thumbnail Image
    Selection of speech processing for cochlear implant prostheses
    Millar, J. B. ; Tong, Y. C. ; Clark, Graeme M. (Monash University Press, 1983)
    In this paper we consider a framework against which to discuss strategies for the design of speech processors for cochlear implant prostheses. We hope to encourage discussion of the bases for such a framework even though it may seem a distant objective owing to the large gaps in our understanding of several component parts of cochlear implant systems. The existence of such a framework would provide a background against which to view the current diverse cochlear implant systems and to evaluate their performance.
  • Item
    Thumbnail Image
    The implanted round window membrane in the cat [Abstract]
    Franz, B. ; Clark, Graeme M. ; Ng, J. ; Bloom, D. (Monash University Press, 1983)
    In cochlear implants the round window is convenient for the electrode insertion into the scala tympani because the surgical approach is reasonably easy and the inserted electrode lies close to systematically organised nerve fibres in the spiral lamina. However, complications might occur when a poor seal, extensive tissue damage or surgical asepsis are present that lead to a reduction in the nerve fibre population which is needed for electrical stimulation. Published articles available do not describe the role of the window membrane in cochlear implants. Probably this can be referred to the finidng of abundant scar tissue in the window niche and around the electrode giving the impression of a safely implanted electrode. This study performed on seven cats over 5 months was concerned with morphological properties of the implanted window membrane at different stages after implantation. In addition, horseradishperoxydase was used as a tracersubstance to give data concerning the sealing properties of the implanted round window membrane.
  • Item
    Thumbnail Image
    Initial results for six patients with a multiple-channel cochlear prosthesis
    Dowell, R. C. ; Brown, A. M. ; Seligman, P. M. ; Clark, Graeme M. (Monash University Press, 1983)
    A total of eight patients have been assessed with the multi-channel cochlear prosthesis at the University of Melbourne. The first two patients were implanted with a prototype device in 1978 and 1979, and their results with various speech evaluation procedures have been reported and summarized in detail elsewhere (Clark & Tong, 1982). Briefly, these results indicated that some very significant benefit could be obtained for these patients when using the cochlear prosthesis with external speech processing, particularly when using the device in conjunction with lipreading. It was also shown that some significant understanding of speech was possible without lipreading (open-set) for both patients, although this was fairly limited.
  • Item
    Thumbnail Image
    The auditory brainstem response in hearing and deaf cats evoked by intracochlear electrical stimulation
    Black, R. C. ; Clark, Graeme M. ; O'Leary, S. J. ; Walters, C. (Monash University Press, 1983)
    This study was performed to investigate in detail the auditory brainstem response (ABR) for intracochlear electrical stimulation. Brainstem response audiometry is a simple, noninvasive procedure with the responses under many stimulus conditions being readily understood in terms of single auditory nerve discharge properties. The amplitude and latency behaviour of the Nl brainstem response correlates well with that recorded directly from the auditory nerve (Huang & Buchwald, 1978). In addition, the brainstem response can be divided into frequency-specific components corresponding to tonotopical locations in the cochlea, as exhibited in the method of derived responses (e.g. Parker &Thornton, 1978). It is therefore well suited to both physiological and clinical investigation of auditory function and therefore should be useful in evaluating auditory function under conditions of electrical stimulation of the cochlea.
  • Item
    Thumbnail Image
    Electrical stimulation of the human cochlea: psychophysical and speech studies
    Clark, Graeme M. (Plenum Publishing Corporation, 1981)
    This report describes psychophysical and speech studies conducted on two of our post-lingually deaf patients implanted with the nature of the hearing sensations produced by the individual electrodes, and to investigate the feasibility of the transmission of speech information to higher centres by means of cadences of stimulation using on electrode at a time. Two totally deaf patients (MC1 and MC2) participated in these studies.
  • Item
    Thumbnail Image
    Signal processing in quiet and noise
    Dowell, R. C. ; Patrick, J. F. ; Blamey, P. J. ; Seligman, P. M. ; Money, D. K. ; Clark, Graeme M. ( 1987)
    It has been shown that many profoundly deaf patients using multichannel cochlear implants are able to understand significant amounts of conversational speech using the prosthesis without the aid of lipreading. These results are usually obtained under ideal acoustic conditions but, unfortunately, the environments in which the prostheses are most often used are rarely perfect. Some form of competing signal is always present in the urban setting, from other conversations, radio and television, appliances, traffic noise and so on. As might be expected, implant users in general find background noise to be the largest detrimental factor in their understanding of speech, both with and without the aid of lipreading. Recently, some assessment of implant patient performance with competing noise has been attempted using a four-alternative forced-choice spondee test (1) at Iowa University. Similar testing has been carried out at the University of Melbourne with a group of patients using the Nucleus multichannel cochlear prosthesis. This study formed part of an assessment of a two formant (F0/FI/F2) speech coding strategy (2). Results suggested that the new scheme provided improved speech recognition both in quiet and with competing noise. This paper reports on some more detailed investigations into the effects of background noise on speech recognition for multichannel cochlear implant users.
  • Item
    Thumbnail Image
    A formant-estimating speech processor for cochlear implant patients
    Blamey, P. J. ; Dowell, R. C. ; Brown, A. M. ; Seligman, P. M. ; Clark, Graeme M. (Speech Science and Technology Conference, 1986)
    A simple formant-estimating speech processor has been developed to make use of the "hearing" produced by electrical stimulation of the auditory nerve with a multiple-channel cochlear implant. Thirteen implant patients were trained and evaluated with a processor that presented the second formant frequency, fundamental frequency, and amplitude envelope of the speech. Nine patients were trained and evaluated with a processor that presented the first formant frequency and amplitude as well. The second group performed significantly better in discrimination tasks and word and sentence recognition through hearing alone. The second group also showed a significantly greater improvement when hearing and lipreading was compared with lipreading alone in a speech tracking task.
  • Item
    Thumbnail Image
    A model of auditory-visual speech perception
    Blamey, P. J. ; Clark, Graeme M. (Speech Science and Technology Conference, 1986)
    A mathematical model relating the probabilities of correctly recognizing speech features, phonemes, and words was tested using data from the clinical trial of a multiple-channel cochlear implant. A monosyllabic word test was presented to the patients in the conditions hearing alone (H), lipreading alone (L), and hearing plus lipreading (HL). The model described the data quite well in each condition. The model was extended to predict the HL scores from the feature recognition probabilities in the H and L conditions. The model may be useful for the evaluation of automatic speech recognition devices as well as hearing impaired people.
  • Item
    Thumbnail Image
    Histopathology following electrode insertion and chronic electrical stimulation
    Shepherd, R. K. ; Clark, Graeme M. ; Pyman, B. C. ; Webb, R. L. ; Murray, M. T. ; Houghton, M. E. (Raven Press, 1985)
    We have examined a number of safety issues associated with cochlear implants. This work has been primarily designed to evaluate the histopathological effects of intracochlear electrode implantation and chronic electrical stimulation. The results of these studies may be summarized as follows: 1) The insertion of the banded free-fit scala tympani array into human cadaver temporal bones produces minimal damage, occurring primarily to a localized region of the spiral ligament. This damage would not result in significant neural degeneration and thus, would not compromise the efficacy of the multiple channel device; 2) chronic intracochlear electrical stimulation for continuous periods of 500 to 2000 hours, using charge balanced biphasic current pulses developing charge densities of 18-32 }?C/cm2. geom./phase, does not adversely affect the spiral ganglion cell population; 3) labyrinthine infection severely reduces the viable spiral ganglion cell population; 4) the formation of new bone present in approximately half of the animals we have implanted --is not associated with electrical stimulation per se; 5) scanning electron microscope studies of electrodes subjected to long periods of intracochlear electrical stimulation reveals minimal platinum dissolution when compared with unstimulated control electrodes, and electrodes that have been stimulated for similar periods in inorganic saline.