Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 10 of 14
  • Item
    Thumbnail Image
    High rate electrical stimulation of the auditory nerve: physiological and pathological results [Abstract]
    Shepherd, Robert K. ; XU, JIN ; TYKOCINSKI, MICHAEL ; Millard, Rodney, E. ; Clark, Graeme M. ( 1995)
    Previous experimental studies have shown that chronic electrical stimulation of the auditory nerve using charge balanced biphasic current pulses at rates of up to 500 pulses per second (pps) do not adversely affect the adjacent spiral ganglion population. More recently, a number of clinical trials have indicated that speech processing strategies based on high pulse rates (1000 pps and more), can further improve speech perception. In this paper we summarize our results following acute and chronic electrical stimulation of the auditory nerve using high pulse rates.
  • Item
    Thumbnail Image
    Speech perception benefits for implanted children with preoperative residual hearing [Abstract]
    Hollow, R. ; Rance, G. ; Dowell, R.C. ; Pyman, B. ; Clark, Graeme M. ; Cowan, R. S. C. ; Galvin, K. L. ; Barker, E. J. ; Sarant, J. Z. ; Dettman, S. ( 1995)
    Since the implantation of the first children with the Nucleus 22-channel cochlear prosthesis in Melbourne in 1985, there has been rapid expansion in the number of implanted children world-wide. Improved surgical technique and experience in paediatric assessment and management have contributed to a trend to implant very young children. At the same time there has also been continuing development of improved speech processing strategies resulting in greater speech perception benefits. In the Melbourne program, over 60% of children obtain significant scores on open-set word and sentence tests using their cochlear implant alone without the aid of lipreading. As parents and professionals have become aware of these improved benefits to speech perception benefits in profoundly deaf children, there have been requests to consider implanting severely-to-profoundly deaf children. In these children with higher levels of residual hearing, only those children with poorer-than-expected performance on speech perception tests using hearing aids have been considered for surgery. A number of such cases have now been implanted in the Melbourne program. The speech perception benefits for this group are reported and are being compared with benefits for the profoundly deaf group of children.
  • Item
    Thumbnail Image
    Speech perception benefits for children using the Speak speech processing strategy in quiet and noise [Abstract]
    Whitford, L.A. ; Dowell, R.C. ; Brown, C. ; Gibson, W.P.R. ; Clark, Graeme M. ; Cowan, R. S. C. ; Galvin, K. L. ; Barker, E. J. ; Sarant, J. Z. ; Shaw, S. ; Everingham, C. ( 1995)
    The Speak speech processing strategy, based on the Spectral Maxima Speech Processor (SMSP) developed at the University of Melbourne, has now been implemented in the Spectra 22 speech processor developed by Cochlear Pty Limited, and clinical trials of both patients changing from the previous Multipeak strategy to Speak and patients starting up with. Speak have been conducted. Results in adult patients changing to Speak have shown significant improvements in speech perception in quiet and particularly in background noise as compared with Multipeak. Preliminary studies with children changing from Multipeak to Speak strategy, measured over a 10 month period, have also shown significant benefits from use of the Speak scheme in both quiet and noisy test situations. Results of follow up studies of these children after more than one year experience with the Speak processing strategy are presented. Statistical analysis of performance over time suggests that an increase in benefit is observed in children after additional experience with the Speak processing strategy. In addition, results for children who have used only the Speak processing strategy from the time of implantation are also presented. The results confirm that the Speak processing strategy provides significant benefits in quiet, and particularly in the presence of background noise for both groups of patients.
  • Item
    Thumbnail Image
    Results of multichannel cochlear implantation in very young children [Abstract]
    Galvin, K. ; Clark, Graeme M. ; Dettman, S. ; Dowell, R. ; Barker, E. ; Rance, G. ; Hollow, R. ; Cowan, R. ( 1995)
    Most researchers and clinicians working in the cochlear implant field have assumed that profoundly deaf children will have a better prognosis in terms of speech perception, speech production and language development. if implanted at as young an age as possible. However, it has been difficult to gather direct evidence for this hypothesis due to the problems in assessing children under the age of five years with formal tests. Recent results with older children have supported the view that early implantation may provide the optimal outcome in most cases. The implantation of very young children raises two areas of concern that do not apply in adults and older children: accurate assessment of degree of hearing loss and auditory potential; and postoperative assessment of outcomes. This paper will describe research results from the University of Melbourne which address these issues and present results for children implanted as young as eighteen months of age.
  • Item
    Thumbnail Image
    Speech perception benefits for children using an advanced cochlear implant speech processing strategy in quiet and in noise [Abstract]
    Dettman, Shani J. ; Skok, Marissa ; Dowell, Richard C. ; Clark, Graeme M. ; Cowan, Robert S. C. ; Whitford, Lesley A. ; Sarant, Julia Z. ; Galvin, Karyn L. ; Barker, Elizabeth J. ; King, Alison ( 1994)
    A new speech processing strategy (SPEAK) has been developed by the University of Melbourne and Cochlear Pty Ltd for use with the Nucleus 22-channel electrode array. In this strategy, 20 programmable filters are repetitively scanned at an average rate of 250Hz and the largest spectral components or maxima are selected from the incoming speech signal. This new speech processing strategy has been shown to provide significantly improved benefits in adult implant patients, particularly in the presence of background noise. This report presents data of a preliminary paediatric clinical trial of the new SPEAK speech processing strategy.
  • Item
    Thumbnail Image
    Physiological and histopathological response of the cochlea to chronic electrical stimulation of the auditory nerve at high stimulus rates [Abstract]
    Shepherd, R. K. ; Xu, J. ; Clark, Graeme M. ( 1994)
    Previous research has shown that chronic electrical stimulation of the auditory nerve using charge balanced biphasic current pulses at rates of up to 500 pulses per second (pps) does not adversely affect the adjacent spiral ganglion population. More recently, a number of clinical trials have suggested that speech processing strategies based on high pulse rates (e.g. 1000 pps), can further improve speech perception. In the present study we evaluated the physiological and histopathological response of the cochlea following long-term stimulation using rates of 1000 pps. Thirteen normal hearing cats were bilaterally implanted with scala tympani electrodes and unilaterally stimulated using 25-50 �s per phase charge balanced biphasic current pulses presented at 1000 pps. Additional charge balance was achieved by shorting the electrodes between current pulses. Each animal was stimulated for periods ranging from 700 - 2100 hours at current levels within its dynamic range. Auditory brainstem responses to both acoustic (ABR) and electrical (EABR) stimuli were periodically recorded throughout the chronic stimulation program. At completion of the program the cochleas were prepared for histological examination. While all animals exhibited an increase in acoustic thresholds following surgery, click evoked ABR's returned to near normal levels in half the animals. Frequency specific stimuli indicated that the most extensive hearing loss occurred adjacent to the array (>12 kHz) while lower frequency thresholds appeared at or near normal Our EABR data showed that the majority of animals exhibited slight increases in threshold, although response amplitudes remained very stable for the duration of the stimulus program. The physiological data reported here will be correlated with cochlear histopathology. These initial findings suggest that chronic intracochlear electrical stimulation at high pulse rates, using a carefully designed charge balanced stimulator, does not appear to adversely affect the implanted cochlea.
  • Item
    Thumbnail Image
    Clinical comparison of open-set speech perception with MSP and WSPIII speech processors and preliminary results for the new SPEAK processor [Abstracts]
    Whitford, Lesley A. ; Partick, James F. ; Clark, Graeme M. ; Dowell, Richard C. ; Marsh, Michael A. ; HOLLOW, RODNEY ; Blamey, Peter J. ; Pyman, Brian C. ; Seligman, Peter M. ( 1993)
    There are several studies which compare the WSP III (FOIF11F2) and MSP (Multipeak) speech processors for the Nucleus multiple-channel cochlear implant in small, controlled groups of patients. In the present study we were interested in the benefits of open set speech perception provided by the MSP over the prior WSP III speech processor in a large, unselected clinical population.
  • Item
    Thumbnail Image
    Speech perception benefits for children using the 22-channel Melbourne/cochlear hearing prosthesis [Abstract]
    Sarant, J.Z. ; Hollow, P.W. ; Clark, Graeme M. ; Dowell, Richard C. ; Cowan, Robert S.C. ; Pyman, B. C. ; Dettman, S. J. ; RANCE, GARY ; Barker, Elizabeth J. ( 1993)
    In 1985; the first child was implanted with the Cochlear 22-channel cochlear prosthesis at the University of Melbourne Royal Victorian Eye & Ear Hospital Cochlear Implant Clinic. There are now 42 children who have received the device in Melbourne. Analysis of patient details for these children show a very heterogeneous group, with a wide range in age, hearing thresholds, duration of deafness and aetiology. The major aetiologies found were either a congenital profound deafness.; or a hearing loss due to meningitis. In all but 3 cases, the children are using 15 or more electrodes in the array. Speech perception benefits have been analyzed according to a six-level hierarchical classification scheme. All of-the children achieved a minimum benefit of discrimination of suprasegmental information (Category 2), and 59% of the children achieved open-set understanding of unfamiliar speech material without the aid of lip-reading (Categories 5 & 6). Detailed analysis suggests that the majority of children achieving open-set speech perception benefits had more than one year of experience with their implant. and less than seven years of profound deafness prior to implantation.
  • Item
    Thumbnail Image
    Factors associated with open-set speech perception in children using the Cochlear multiple-channel prosthesis [Abstract]
    Yaremko, R. ; Rance, G. ; Sarant, Julia Z. ; Dawson, Pam W. ; Gibson, William P.R. ; Clark, Graeme M. ; Dowell, Richard C. ; Cowan, Robert S.C. ; Brown, Catherine D. ; Dettman, Shani J. ; Barker, Jane ; Barker, Elizabeth J. ( 1993)
    Since 1985, nearly 100 children have received the 22-channel cochlear prosthesis from the Melbourne and Sydney cochlear implant clinics. These two clinics account for the bulk of casesin Australia, and have similar management philosophies and selection criteria. The patient population represents a variety of etiologies, and ranges in age from 2 - 18 years of age. Bothcongenital and postlinguistic hearing losses are included. In order to assess benefit to speech perception in such a diverse group, the children's results have been tabulated according to a six level hierarchical scale of speech perception achievement. The scale ranges from category I,detection of sound only, to category 6, which includes significant perception scores for open-setwords and sentences. Analysis of the results shows that the majority of the children are achieving open-set speech perception benefits, and that results continue to improve with additional experience with their devices. There are a number of contributing factors to these open-set speech� perception results which have impact both on selection issues and on habilitation with different age ranges �of patients.
  • Item
    Thumbnail Image
    Future developments in speech processing for multichannel cochlear implants in children [Abstract]
    Dowell, Richard C. ; Dooley, G. ; McDermott, H. D. ; Blamey, P. ; McKay, C. ; Clark, Graeme M. ( 1992)
    The average speech perception score for adult implant patients is now about 60% on an open-set sentence test without lipreading. This is higher than the scores obtained by many profound and severe-to-profoundly impaired hearing aid users. This suggests that some hearing aid users, particularly those who use a hearing aid in one ear only, could benefit from a cochlear implant. As neither the implant nor the hearing aid will provide perfect speech recognition it is to be expected that this group should obtain maximum benefit by using the hearing aid in one ear together with the implant in the other ear. However, experience with this group of patients has shown that many people find the use of two independent devices unacceptable. Furthermore, perceptual interaction of the acoustic and electrical signals makes it desirable to be able to control the two outputs in a more co-ordinated way than is possible with two independent devices. Consequently, a "bimodal" speech processor has been developed with both acoustic and electrical outputs controlled from the same speech processing unit. Feature coding aspects of the implant processing have been applied to the acoustic signal in such a way as to enhance speech perception with the hearing aid and improve compatibility with the implant. Initial testing with the bimodal aid shows promise to help severely-to-profoundly impaired individuals. The device has also been useful as a research tool to investigate the complex interactions of simultaneous acoustic and electrical stimulation. The Spectral Maxima Sound Processor (SMSP) has also been developed at the University of Melbourne for use with the Nucleus cochlear implant. Studies with adult subjects have shown improved perception of vowels, consonants, words and sentences in quiet and sentences in background noise with the SMSP as compared with the MSP(MULTIPEAK) which is currently supplied for use with this implant. Results for four subjects showed mean scores for open set sentences at a 10 dB signal-to-noise ratio of 78.7% for the SMSP and 50.0% for the MSP. Mean scores for the same group on open set monosyllabic words in quiet were 57.4% for SMSP and 39.9% for MSP. These results suggest that future improvements in speech perception will be possible for children using the Nucleus cochlear implant.