    Variation in speech perception scores among children with cochlear implants
    Sarant, JZ ; Blamey, PJ ; Dowell, RC ; Clark, GM ; Gibson, WPR (LIPPINCOTT WILLIAMS & WILKINS, 2001-02)
    OBJECTIVE: The objective of this study was to identify common factors affecting speech perception scores in children with cochlear implants. DESIGN: Speech perception data for 167 implanted children were collected at two cochlear implant centres in Melbourne and Sydney. The data comprised audition-alone scores on open-set word and sentence tests. Children were selected on the basis that they had a Nucleus 22-electrode cochlear implant. The average age of the children was 5 yr. Information was also collected about 12 factors that may have influenced speech perception scores for each child. Analysis of covariance was used to identify factors that significantly affected speech perception scores. Pearson pairwise correlation coefficients were also calculated for all factors analyzed. RESULTS: The analyses in this study identified factors that accounted for 51%, 34%, and 45% of the variance in phoneme, word and sentence perception scores. Scores decreased by 1.4 to 2.4% per year of profound deafness prior to implantation. Children who normally use oral communication scored significantly higher than children normally using sign or simultaneous oral and sign communication. Children implanted in Sydney scored higher on average than children implanted in Melbourne. CONCLUSIONS: The results show that a significant part of the variation in speech perception scores is systematically related to audiological and environmental factors for each child. The reasons for significant differences between children using different communication modes or from different clinics were not identified.
  • Item
    Thumbnail Image
    Electrode discrimination and speech perception in young children using cochlear implants
    Dawson, P. W. ; McKay, C. M. ; Busby, P. A. ; Grayden, D. B. ; Clark, Graeme M. ( 2000)
    Objective: The aim was to determine the efficacy of a child-appropriate procedure to assess electrode discrimination ability in young children using cochlear implants and to investigate the relationship of electrode discrimination ability and speech perception performance in children implanted at a young age. Design: An adaptation of the play audiometry procedure was used to assess electrode discrimination in seventeen 4- to 10-yr-old children. The children were required to respond with a game-like motor response when a repeating stimulation on a reference electrode “changed” to a different electrode. They were also assessed on a speech feature discrimination test, a closed-set word recognition test and a nonverbal intelligence task. Results: Sixty-five percent of subjects demonstrated ability to discriminate adjacent electrodes in mid and apical regions of the cochlea, whilst the remaining subjects needed electrode separations of between two and nine electrodes for successful discrimination. In a forward stepwise regression analysis electrode discrimination ability was found to be the strongest factor in accounting for variance in the speech perception scores. Subject variables such as duration of deafness, nonverbal intelligence and implant experience did not significantly account for further variance in the speech perception scores for this group of children. Conclusions: Electrode discrimination ability was the strongest factor in predicting performance on speech perception measures in a group of children using cochlear implants.
  • Item
    Thumbnail Image
    The development of the Melbourne/Cochlear multiple-channel cochlear implant for profoundly deaf children
    Clark, Graeme M. ; Busby, Peter A. ; Dowell, Richard C. ; Dawson, Pamella W. ; Pyman, Brian C. ; Webb, Robert L. ; Staller, Steven J. ; Beiter, Anne L. ; Brimacombe, Judith A. ( 1992)
    In 1978-79, a speech processing strategy which extracted the voicing (FO) and second formant (F2) frequencies and presented these as rate and place of stimulation respectively to residual auditory nerve fibres was developed for the University of Melbourne's prototype multiple-channel receiver-stimulator (Clark et aI1977, Clark et a11978, Tong et aI1980). This speech processing strategy was shown to provide post linguistically deaf adults with some open-set speech comprehension using electrical stimulation alone, and considerable help when used in combination with lipreading (Clark et al 1981).
  • Item
    Thumbnail Image
    Speech perception, production and language results in a group of children using the 22-electrode cochlear implant
    Blamey, P. J. ; Dawson, P. W. ; Dettman, S. J. ; Rowland, L. C. ; Brown, A. M. ; Busby, P. A. ; Dowell, R. C. ; Rickards, F. W. ; Clark, Graeme M. ( 1992)
    Five children out of a group of nine (aged 5.5 to 19.9 years) implanted with the 22-electrode cochlear implant (Cochlear Ply. Ltd.) have achieved substantial scores on open-set speech tests using hearing without lipreading. Phoneme scores for monosyllabic words ranged from 40% to 72%. Word scores in sentences ranged from 26% to 74%. Four of these five children were implanted during preadolescence. The fifth child, who had a progressive loss and was implanted during adolescence after a short period of very profound deafness, scored highest on all speech perception tests. The remaining four children who did not demonstrate open-set recognition were implanted during adolescence after a long duration of profound deafness. Post-operative performance on closed-set speech perception tests was better than pre-operative performance for all children. Improvements in speech and language assessments were also noted. These improvements tended to be greater for the younger children. The results are discussed with reference to variables which may contribute to successful implant use: such as age at onset, duration of profound hearing loss, age at implantation, aetiology, educational program, and the type of training provided.
  • Item
    Thumbnail Image
    Evaluation of expandable leadwires for paediatric cochlear implants
    Xu, Shi-Ang. ; Shepherd, Robert K. ; Clark, Graeme M. ; Tong, Yit C. ; Williams, John F. ( 1993)
    The development of cochlear implants for use in very young children (1-2 years old) will require techniques designed to accommodate temporal bone growth. Previous anatomic studies have shown that the leadwire of a cochlear implant must be capable of expanding up to 20 mm between the round window and the implanted receiver-stimulator in response to skull growth. In the present study morphologic and biomechanical evaluation of five expandable leadwire designs was conducted following their implantation in young cats. Two helical shaped leadwire designs frequently exhibited extensive fibrous tissue adhesions and broke during long-term implantation. In contrast, thin, flexible Silastic envelopes were effective in minimizing tissue adhesions. Residual V- and Z-shaped leadwires, placed in these envelopes, showed little evidence of fibrous tissue adhesions following implantation periods of up to 2 years. Moreover, these leadwires readily expanded both during the growth of the animal and when biomechanical expansion studies performed at the completion of the implant period. These expandable leadwire designs appear to be appropriate candidates for use in pediatric cochlear implants.
  • Item
    Thumbnail Image
    Paediatric cochlear implantation: radiologic observations of skull growth
    XU, JIN ; Shepherd, Robert K. ; Xu, Shi-Ang ; Seldon, H. Lee ; Clark, Graeme M. ( 1993)
    We investigated the effects of long-term implantation of auditory prostheses on skull growth in young animals. Four monkeys were implanted with dummy cochlear implants at 6 months of age. To simulate implantation in children, the bed for the receiver-stimulator or interconnecting plug was drilled across a calvarial suture down to the underlying dura. Plain skull oentgenograms were periodically taken to monitor head growth for up to 3 years after implantation. These longitudinal measurements revealed no significant asymmetric skull growth. Postmortem measurements using computed tomographic scans confirmed these results and showed no significant difference in the intracranial volumes between the implanted and control sides of each animal or between experimental and nonimplanted control monkeys. These results suggest that long-term cochlear implantation in very young children will not cause any significant deformity of the skull.
  • Item
    Thumbnail Image
    The postnatal growth of the temporal bone and its implications for cochlear implantation in children
    Dahm, Markus C. ; Shepherd, Robert K. ; Clark, Graeme M. ( 1993)
    The postnatal growth of the human temporal bone was examined by direct anatomical measurements on 60 cadaver specimens of all ages. The bones were dissected as one would perform cochlear implant surgery using a posterior tympanotomy approach. Nineteen anatomical /surgical landmarks with implications for cochlear implant surgery were identified on each bone and the distance between these points measured. The temporal hone was found to be a complex structure, phylogenetically, anatomically and functionally consisting of four different parts with independent postnatal development. The inner and middle cars were adult size at birth. The external auditory canal and most parts of the temporal hone were subject to significant lateral growth. The size of the pneumatised mastoid increased in all directions. In the facial recess, however, no postnatal growth was observed. Between birth and adulthood an average of 12 mm (SD 5 mm) of growth was seen directly between the sino-dural angle and the round window, the landmarks approximating the Implantation site for the receiver-stimulator and the electrode entry point into the inner car. However, if an electrode leadwire is fixed at a cortical fixation site such as the posterosuperior point of Macewen's triangle, the leadwire would be subject to approximately 20 mm of growth. These results indicate that a paediatric cochlear implant design incorporating an expandable leadwire to accommodate this growth should allow up to 25 mm of leadwire lengthening. The fossa incudis showed no growth relative to the round window and was found to be a convenient fixation site for the electrode array close to the cochlea. From an anatomical and surgical point of view, cochlear implantation in very young children is feasible, provided the electrode array is secured and the design accommodates for controlled leadwire lengthening.