Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 10 of 13
  • Item
    Thumbnail Image
    Speech perception in implanted children: influence of preoperative residual hearing on outcomes [Abstract]
    Cowan, R. S. C. ; Barker, E. J. ; Pegg, P. ; Dettman, S. ; Rennie, M. ; Galvin, K. ; Meskin, T. ; Rance, G. ; Cody, K. ; Sarant, J. ; Larratt, M. ; Latus, K. ; HOLLOW, RODNEY ; Rehn, C. ; Dowell, R. C. ; Pyman, B. ; Gibson, W. P. R. ; Clark, Graeme M. ( 1998)
    Since the first child was implanted with the Nucleus 22-channel prosthesis in Melbourne in 1985, several thousand children world-wide have now benefitted from this technology. More effective paediatric assessment and management procedures have now been developed, allowing cochlear implants to be offered to children under the age of 2 years. Improvements in speech processing strategy have also been implemented in the Nucleus implant system, resulting in increased mean speech perception benefits for implanted adults. Although a range of performance on formal measures of hearing, speech or language has been reported for children using implants, results from the first decade of implant experience consistently show that significant benefits are available to children receiving their implant at an early age. Reported speech perception results for implanted children show that a considerable proportion (60%) of paediatric patients in the Melbourne and Sydney clinics are able to understand some open-set speech using electrical stimulation alone. These results, and the upward trend of mean speech perception benefits shown for postlinguistically deafened adults have raised questions as to whether severely, or severely-to-profoundly deaf children currently using hearing aids would in fact benefit more from a cochlear implant. To investigate the potential influence of the degree of preoperative residual hearing on postoperative speech perception, results for all implanted children in the Melbourne and Sydney cochlear implant programs were analysed. Results showed that as a group, children with higher levels of preoperative residual hearing were consistently more likely to achieve open-set speech perception benefits. Potential factors in this finding could be higher levels of ganglion cell survival or greater patterning of the auditory pathways using conventional hearing aids prior to implantation. Conversely, children with the least preoperative residual hearing were less predictable, with some children achieving open-set perception, and others showing more limited closed-set benefits to perception. For these children, it is likely that preoperative residual hearing is of less significance than other factors in outcomes.
  • Item
    Thumbnail Image
    Acquisition of a tactile-alone vocabulary by normally hearing users of the Tickle Talker
    Galvin, Karyn L. ; Blamey, Peter J. ; Oerlemans, Michael ; Cowan, Robert S. C. ; Clark, Graeme M. ( 1999)
    Abstract not available due to copyright.
  • Item
    Thumbnail Image
    A modification of play audiometry to assess speech discrimination ability in severe-profoundly deaf 2- to 4-year-old children
    Dawson, P. W. ; Nott, P. E. ; Clark, Graeme M. ; Cowan, Robert S. C. ( 1998)
    Objective: The aim was to develop an assessment procedure that was independent of language and speech production ability, to test speech feature discrimination severe-profoundly deaf children 2 to 4 yr of age. Design: The procedure being trialled was adapted from existing procedures. The child was required to respond with a game-like motor response to a “change” in a speech stimulus that was being presented repeatedly through a speaker. The change occurred at randomly determined times, and false alarm responses were measured during the waiting periods (while the child waited for the change). Two- to four-yr-old normally hearing children and hearing-impaired children using hearing aids and a group of 4-yr-old hearing-impaired children using cochlear implants were assessed on the task. Results: More than 82% of the 3- and 4-yr-old normally hearing and hearing-impaired children were able to complete the testing for the eight speech sound contrasts within three 20 minute sessions. Fifty percent of the 2-yr-old normally hearing and hearing-impaired children were able to condition and complete the task. All of the normally hearing children who completed the task successfully discriminated all speech sound contrasts. The performance of the hearing-impaired children using hearing aids was influenced by the degree of hearing loss and the type of speech contrast being tested. Similarly, the average performance of the children using cochlear implants was better for easier contrasts such as /ba/bi/ with contrasting vowel formant cues. Conclusions: This procedure has potential for use as a reliable clinical and research tool for assessing the development of auditory discrimination ability in 2- to 4-yr-old severe-profoundly deaf children.
  • Item
    Thumbnail Image
    A comparison of Tactaid II+ and Tactaid 7 use by adults with a profound hearing impairment
    Galvin, Karyn L. ; Mavrias, Gina ; Moore, Alessandra ; Cowan, Robert S. C. ; Blamery, Peter J. ; Clark, Graeme M. ( 1999)
    Objective: To evaluate and compare use of the Tactaid II+ and the Tactaid 7, in terms of speech perception, by adults with a hearing impairment. Design: Eight adults used one device daily for approximately 10 wk and attended seven training sessions. Performance was measured with tests of phonetic contrast perception, closed-set vowel and consonant identification, word and phoneme recognition in monosyllabic word lists, word recognition in sentences and speechtracking rate. A questionnaire was also administered. The protocol was repeated with the alternative device. Results: With each device, the group discriminated most phonetic contrasts at better-than-chance levels and demonstrated somewhat enhanced visual or auditory-visual perception when measured in terms of vowel identification, monosyllabic word recognition and speechtracking rate. An increase in speechtracking rate was also demonstrated for some individuals. Subjects generally reported little subjective improvement in speech perception and production, but were satisfied with the physical attributes of each device. Five of six subjects preferred the Tactaid 7. Conclusions: The Tactaid II+ and the Tactaid 7 provided suprasegmental and segmental information, enabling the group to discriminate phonetic contrasts and improve their perception of some speech materials. No consistent advantage was found for either device, thought most subjects preferred the Tactaid 7. Alternatives likely to provide a greater benefit to communication should be considered before Tactaid fitting.
  • Item
    Thumbnail Image
    Improved sound processing for cochlear implants
    James, C.J. ; Just, Y. ; Knight, M.R. ; Martin, L.F.A. ; McKay, C.M. ; Plant, K.L. ; Tari, S. ; Vandali, A.E. ; Clark, Graeme M. ; Cowan, R.S.C. ; McDermott, H. J. ; Blamey, P. J. ; Dawson, P. ; Fearn, R. A. ; Grayden, D. B. ; Henshall, K. R. ( 2002)
    Four signal processing schemes currently under development aim to improve the perception of sounds/ especially speech, for children and adults using the Nucleus cochlear implant system. The schemes are (1) fast-acting input-signal compression, (2) Adaptive Dynamic Range Optimisation (ADRO), (3) TESM, a scheme that emphasises transients in signals, and (4) DRSP, a strategy that applies different stimulation rates to selected sets of electrodes.
  • Item
    Thumbnail Image
    Application of advanced radiographic technology in cochlear implant research
    XU, JIN ; TYKOCINSKI, MICHAEL ; Saunders, E. ; Clark, Graeme M. ; Cowan, R. ( 2001)
    The effective development of peri-modiolar or other advanced electrode arrays for cochlear implants requires detailed analysis of the insertion procedure and electrode positioning in the cochlea. Routine x-ray techniques cannot provide sufficient detail to meet this need. A new micro-focus x-ray imaging system has been built for our research. The system consists of a x-ray tube with a sub 10-micron focal spot mounted below an adjustable work surface and an image intensifier placed approximately 100 cm above the x-ray aperture. A variety of intracochlear electrode arrays and human temporal bones were studied using this system. The micro-focus x-ray imaging system allows for micro-fluoroscopy to visualise the real time implantation procedure. It also enables capturing of images onto reusable phosphor imaging plates or films for subsequent viewing or analysis. Images are produced at up to 95 times magnification with superior resolution and enhanced contrast. This new radiographic technology plays an important role in development of safe and effective advanced intracochlear electrode arrays.
  • Item
    Thumbnail Image
    Development of safe and effective electrodes: a risk management approach [Abstract]
    Clark, Graeme M. ; COWAN, ROBERT ; Saunders, Elaine ; TYKOCINSKI, MICHAEL ; Cohen, Lawrence ; Treaba, Claudiu ; Briggs, Robert S. ; GIBSON, PETER ( 1999)
    Results from studies on experimental animals, computer modelling and preliminary psychophysical studies with three patients, have confirmed the potential for subjective improvement with electrode arrays which lie closer to the modiolus than does the Nucleus straight array. Results of psychophysical studies with three cochlear implant patients, using developmental pre-curved arrays, confirm the feasibility of improving patients' performance through improvements in electrode design. In particular, it was found in psychophysical tests, with patients using a developmental pre-curved electrode array, that both maximum comfortable level and threshold reduced with decreasing distance of a stimulated electrode from the modiolus, and that the dynamic range increased. More intense neural excitation patterns were obtained with the closer electrodes. From this it is inferred that the development of more sophisticated electrode arrays, positioned closer to the modiolus than is currently the case with the Nucleus standard array, will enable the development of improved speech processing strategies. There are technical constraints in the design of a peri-modiolar array, and currently a number of approaches to this problem have been investigated. Whilst the goal of the design is that it be effective for sophisticated and variable manners of stimulus delivery, a primary constraint is safety.
  • Item
    Thumbnail Image
    Insertion study using new peri-modiolar electrode array designs [Abstract]
    Treaba, Claudiu ; Clark, Graeme M. ; Cowan, Robert S. ; Tykocinski, Michael J. ; Cohen, Lawrence T. ; Saunders, Elaine ; Pyman, Brian C. ; Briggs, Robert S. ; Dahm, Markus C. ( 1999)
    Intracochlear multi-channel cochlear implants have been shown to successfully provide auditory information for profoundly deaf patients by electrically stimulating discrete populations of auditory nerve fibers via a scala tympani (ST) electrode array. Histological and radiological examination of implanted human temporal bones showed that the current straight Nucleus® array is usually positioned against the outer wall of the ST. An electrode array close to the modiolus could be expected to reduce stimulation thresholds and result in a more localized neural excitation pattern.
  • Item
    Thumbnail Image
    Studies in tactile speech perception using the University of Melbourne Tickle Talker [Abstract]
    Galvin, K. L. ; Cowan, R. S. C. ; Blamey, P. J. ; Oerlemans, M. ; Ginis, J. ; Mavrias, G. ; Moore, A. ; Lu, A. ; Millard, R. ; Clark, Graeme M. ( 1998)
    During the past four years, a series of studies have been undertaken investigating the perception of speech presented through the tactile modality. These studies formed part of a project to develop an effective and safe electro tactile speech perception device for those children and adults requiring additional communication assistance to that provided by hearing aids, but unable to benefit from cochlear implants.
  • Item
    Thumbnail Image
    Benefits of pre-curved electrode arrays for the Nucleus multichannel cochlear implant [Abstract]
    Pyman, B. ; Clark, Graeme M. ; Saunders, Elaine ; Cohen, Lawrence T. ; Cowan, Robert S. C. ; Treaba, C. ; Dahm, M. ; Tykocinski, M. ( 1998)
    The purpose of the study was to evaluate whether placement of an electrode closer to the modiolus would provide a more efficient interface with the auditory nerve. A pre-curved, banded 22-electrode array has been developed. Following comprehensive safety studies, the array was inserted in three adult patients. Detailed x-ray analyses have been conducted to specify the positions of the electrode bands. Psychophysical studies have been conducted, within each subject, using electrodes which lie at differing distances from the modiolus. These studies have shown that as the distance from the modiolus decreased, electrode threshold currents decreased and dynamic ranges increased. Thresholds were lower than in a comparison group of standard array users. Forward masking studies showed current spread to be more focused at the electrodes lying closer to the modiolus. Electrode discrimination was also better at electrodes closer to the modiolus. JND for loudness, expressed as a function of dynamic range also decreased in this condition. Thus, a preliminary profile of the parameters of interest, for an improved implant design, suggest that an array which lies closer to the modiolus can result in lower current requirements and a more localised pattern of neural excitation.