Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 5 of 5
  • Item
    Thumbnail Image
    The spectral maxima sound processor: recent findings in speech perception and psychophysics
    McKay, Colette M. ; McDermott, Hugh J. ; Vandali, Andrew E. ; Clark, Graeme M. (Wien, 1994)
    The Spectral Maxima Sound Processor (SMSP) was developed at the University of Melbourne for use with the Mini System 22 implant manufactured by Cochlear Pty Ltd. The SMSP has been shown in recent studies to provide improved speech perception to implantees when compared to the currently commercially available processor for this implant (the MSP (MULTIPEAK) processor). In the first of three experiments, the effect on speech perception of increasing the rate of stimulation of the SMSP and of increasing the number of electrodes activated in each stimulation cycle was studied. It was found that these parameter changes made little difference to speech perception in quiet but both changes were advantageous for some subjects when listening in noise. The second and third experiments investigated psychophysically the effects of two aspects of the SMSP strategy which differ from previous processors for this implant. In the second experiment, it was found that concurrent stimulation of two adjacent or nearby electrodes evoked a pitch which was intermediate to that of either electrode. This may explain, in part, the better discrimination of vowel formants by users of the SMSP. In the third experiment, it was found that a pitch related to the modulation frequency was evoked by amplitude-modulating a constant rate stimulus, provided that the rate of stimulation was sufficiently high (four times the modulation frequency) or a multiple of the modulation frequency. This result may explain the equal ability of SMSP and MSP users to perceive speaker differences and intonation patterns, even though the rate of stimulation is constant In the SMSP.
  • Item
    Thumbnail Image
    Speech processing for multichannel cochlear implants: variations of the Spectral Maxima Sound Processor strategy
    McKay, Colette M. ; Vandali, Andrew E. ; McDermott, Hugh J. ; Clark, Graeme M. ( 1994)
    The Spectral Maxima Sound Processor (SMSP) incorporates a bank of 16 band-pass filters which are assigned to 16 intracochlear electrodes. In each stimulation period 6 electrodes are activated, based on the outputs of the filters with the largest amplitudes. The SMSP has previously been compared with the present MSP (MULTIPEAK) processor and found to improve speech comprehension results. The SMSP speech processing scheme has recently been implemented successfully in a new speech processor, also developed at the University of Melbourne, which utilises digital signal processing techniques. The programming flexibility of this processor has facilitated the investigation of variations of the SMSP strategy which might provide further enhancement of speech perception. Three variations have been investigated: firstly, increasing the constant pulse rate from the usual 250 Hz to 400 Hz; secondly, changing the number of electrodes selected in each stimulation period from 6 to numbers between 4 and 8; thirdly, sharpening the spectral peaks prior to selection of the active electrodes. The results of these studies showed that all three variations had minimal effect on speech perception in quiet, but that increasing the number of electrodes selected for stimulation to 8, or increasing the rate of stimulation, was advantageous for some subjects when listening in background noise.
  • Item
    Thumbnail Image
    Comparison of current speech coding strategies
    Whitford, L. A. ; Seligman, P. M. ; Blamey, Peter J. ; McDermott, H. J. ; Patrick, J. F. ( 1993)
    This paper reports on two studies carried out at the University of Melbourne jointly with Cochlear Pty Ltd. The studies demonstrated substantial speech perception improvements over the current Multipeak strategy in background noise.
  • Item
    Thumbnail Image
    A new portable sound processor for the University of Melbourne/ Nucleus Limited multielectrode cochlear implant
    McDermott, Hugh J. ; McKay, Colette M. ; Vandali, Andrew E. ( 1992)
    Abstract not available due to copyright.
  • Item
    Thumbnail Image
    An extension of the Multipeak speech processing strategy for the MSP/MINI 22 cochlear implant system
    Jones, P. A. ; McDermott, H. J. ; Sellgman, P. M. ; Millar, J. B. ( 1992)
    The speech perception of three post-linguistically deaf adults using the Nucleus MSP/Mini System 22 cochlear implant system programmed with a new speech processing strategy, MPEAK+AO. was evaluated. The MPEAK+AO strategy retains all the information of the standard Multipeak speech processing strategy and additionally presents acoustic components below 400Hz to the most-apical electrode. This extra spectral Information may help implantees understand speech, particularly in noise. Since the estimated fundamental frequency is presented as the rate of stimulation at a fixed intracochlear site and is thereby potentially perceived more easily. and the amplitude of the stimulation on the apical electrode, associated with the voice fundamental, Is directly determined from the estimated energy in the relevant spectral region. these coding factors may provide a better representation ot the prosodic information in speech and a more complete auditory feedback signal. The comparison between Multipeak and MPEAK+AO included tests of vowel, consonant and CNC word recognition. Speech materials were presented with both a male and female speaker. Sentence material. presented with background masking noise (four-speaker babble) was also used. The results showed that the new strategy significantly improved the ability of these MSP users to recognise words in open-set sentences in noisy conditions.