    Evaluation of a new Spectral Peak coding strategy for the Nucleus 22 channel cochlear implant system
    Skinner, Margaret W. ; Clark, Graeme M. ; Whitford, Lesley A. ; Seligman, Peter M. ; Staller, Steven J. ; Shipp, David B. ; Shallop, Jon K. ; Everingham, Colleen ; Menapace, Christine M. ; Arndt, Patti L. ; Antogenelli, Trisha ; Brimacombe, Judith A. ; Pijl, Sipke ; Daniels, Paulette ; George, Catherine R. ; McDermott, Hugh J. ; Beiter, Anne L. ( 1994)
    Sixty-three postlinguistically deaf adults from four English-speaking countries participated in a 17-week field study of performance with a new speech coding strategy, Spectral Peak (SPEAK), and the most widely used strategy, Multipeak (MPEAK), both of which are implemented on wearable speech processors of the Nucleus 22 Channel Cochlear Implant System; MPEAK is a feature-extraction strategy, whereas SPEAK is a filterbank strategy. Subjects' performance was evaluated with an experimental design in which use of each strategy was reversed and replicated (ABAB). Average scores for speech tests presented sound-only at 70 dB SPL were higher with the SPEAK strategy than with the MPEAK strategy. For tests in quiet, mean scores for medial vowels were 74.8 percent versus 70.1 percent; for medial consonants, 68.6 percent versus 56.6 percent; for monosyllabic words, 33.8 percent versus 24.6 percent; and for sentences, 77.5 percent versus 67.4 percent. For tests in noise, mean scores for Four-Choice Spondees at +10 and +5 dB signal-to-noise ratio (S/N) were 88.5 percent versus 73.6 percent and 80.1 percent versus 62.3 percent, respectively; and for sentences at +15 dB, +10, and +5 dB S/N, 66.5 percent versus 43.4 percent, 61.5 percent versus 37.1 percent, and 60.4 percent versus 31.7 percent, respectively. Subjects showed marked improvement in recognition of sentences in noise with the new SPEAK filterbank strategy. These results agree closely with subjects' responses to a questionnaire on which approximately 80 percent reported they heard best with the SPEAK strategy for everyday listening situations.
  • Item
    Thumbnail Image
    The spectral maxima sound processor: recent findings in speech perception and psychophysics
    McKay, Colette M. ; McDermott, Hugh J. ; Vandali, Andrew E. ; Clark, Graeme M. (Wien, 1994)
    The Spectral Maxima Sound Processor (SMSP) was developed at the University of Melbourne for use with the Mini System 22 implant manufactured by Cochlear Pty Ltd. The SMSP has been shown in recent studies to provide improved speech perception to implantees when compared to the currently commercially available processor for this implant (the MSP (MULTIPEAK) processor). In the first of three experiments, the effect on speech perception of increasing the rate of stimulation of the SMSP and of increasing the number of electrodes activated in each stimulation cycle was studied. It was found that these parameter changes made little difference to speech perception in quiet but both changes were advantageous for some subjects when listening in noise. The second and third experiments investigated psychophysically the effects of two aspects of the SMSP strategy which differ from previous processors for this implant. In the second experiment, it was found that concurrent stimulation of two adjacent or nearby electrodes evoked a pitch which was intermediate to that of either electrode. This may explain, in part, the better discrimination of vowel formants by users of the SMSP. In the third experiment, it was found that a pitch related to the modulation frequency was evoked by amplitude-modulating a constant rate stimulus, provided that the rate of stimulation was sufficiently high (four times the modulation frequency) or a multiple of the modulation frequency. This result may explain the equal ability of SMSP and MSP users to perceive speaker differences and intonation patterns, even though the rate of stimulation is constant In the SMSP.
  • Item
    Thumbnail Image
    Loudness growth characteristics of cochlear implantees using the Spectral Maxima Sound Processor [Abstract]
    MCDERMOTT, HUGH ; MCKAY, COLETTE ( 1994)
    The study of perceptual characteristics of subjects with cochlear implants can lead to improvements in the design of speech processors. One important aspect of speech processing which has received little attention in the past is the conversion acoustic signal amplitudes into appropriate levels of electrical stimulation. The optimum conversion would provide implantees with loudness growth characteristics that mimic those of normal hearing. To investigate how implantees using the Spectral Maxima Sound Processor (SMSP) perceive changes in loudness, an experiment involving production of fixed loudness ratios was conducted. Ten subjects participated: five users of the Mini System 22 cochlear implant, and five normally-hearing subjects. In the experiment, the subjects were required to adjust the loudness of two stimuli (white noise and speech-weighted noise) to equal half or twice that of a reference. The reference was presented at various levels over a range of 25 to 75 dBA. The results for three of the implantees were similar to those of all the normally-hearing subjects, who produced an average level change of 10.8 dB for the task. The remaining subjects, who had the largest electrical dynamic ranges, produced larger level changes (up to 20 dB) which were constrained by the limited electrical dynamic range of the processor (46 dB). The SMSP utilises an amplitude conversion function by which the stimulus level (in dB) is directly proportional to the input sound level (in dB). The experimental results suggest that the shape of this function is satisfactory, though not necessarily optimum, for these implantees.
  • Item
    Thumbnail Image
    Pitch ranking with nonsimultaneous dual-electrode electrical stimulation of the cochlea
    McDermott, Hugh J. ; McKay, Colette M. ( 1994)
    Abstract not available due to copyright.
  • Item
    Thumbnail Image
    Speech processing for multichannel cochlear implants: variations of the Spectral Maxima Sound Processor strategy
    McKay, Colette M. ; Vandali, Andrew E. ; McDermott, Hugh J. ; Clark, Graeme M. ( 1994)
    The Spectral Maxima Sound Processor (SMSP) incorporates a bank of 16 band-pass filters which are assigned to 16 intracochlear electrodes. In each stimulation period 6 electrodes are activated, based on the outputs of the filters with the largest amplitudes. The SMSP has previously been compared with the present MSP (MULTIPEAK) processor and found to improve speech comprehension results. The SMSP speech processing scheme has recently been implemented successfully in a new speech processor, also developed at the University of Melbourne, which utilises digital signal processing techniques. The programming flexibility of this processor has facilitated the investigation of variations of the SMSP strategy which might provide further enhancement of speech perception. Three variations have been investigated: firstly, increasing the constant pulse rate from the usual 250 Hz to 400 Hz; secondly, changing the number of electrodes selected in each stimulation period from 6 to numbers between 4 and 8; thirdly, sharpening the spectral peaks prior to selection of the active electrodes. The results of these studies showed that all three variations had minimal effect on speech perception in quiet, but that increasing the number of electrodes selected for stimulation to 8, or increasing the rate of stimulation, was advantageous for some subjects when listening in background noise.
  • Item
    Thumbnail Image
    Pitch percepts associated with amplitude-modulated current pulse trains in cochlear implantees
    McKay, Colette M. ; McDermott, Hugh J. ; Clark, Graeme M. ( 1994)
    Abstract not available due to copyright.
  • Item
    Thumbnail Image
    A comparison of speech perception of cochlear implantees using the Spectral Maxima Sound Processor (SMSP) and the MSP (Multipeak) processor
    McKay, Colette M. ; McDermott, Hugh J. ; Vandali, Andrew E. ; Clark, Graeme M. ( 1992)
    The Spectral Maxima Sound Processor (SMSP) is a portable speech processor which has recently been developed at the University of Melbourne for use with multiple-electrode cochlear implants. In this processor, the six largest outputs (maxima) of 16 bandpass filters are used to stimulate the cochlea on a place basis at a constant rate. This speech processing strategy has been compared with the MSP(MULTIPEAK) strategy, in which four electrodes are selected for stimulation in every glottal pulse period. The study was undertaken on four postlinguistically deaf adults. The results show that, for this group of subjects, the performance of the SMSP processor was significantly better than that of the MSP(MUL TIPEAK) processor for the recognition of closed-set vowels and consonants, open-set monosyllabic words, and open-set sentences in noise, when using electrical stimulation alone. The SMSP mean scores were: vowels 91.3%, consonants 74.9%, words 57.4%, and sentences in noise 78.7%. The MSP(MULTIPEAK) mean scores were: vowels 76.3%, consonants 59.4%, words 39.9%, and sentences in noise 50.0%.
  • Item
    Thumbnail Image
    Comparison of current speech coding strategies
    Whitford, L. A. ; Seligman, P. M. ; Blamey, Peter J. ; McDermott, H. J. ; Patrick, J. F. ( 1993)
    This paper reports on two studies carried out at the University of Melbourne jointly with Cochlear Pty Ltd. The studies demonstrated substantial speech perception improvements over the current Multipeak strategy in background noise.
  • Item
    Thumbnail Image
    The pitch of amplitude-modulated electrical stimuli in cochlear implantees [Abstract]
    McKay, Colette M. ; McDermott, Hugh J. ; Clark, Graeme M. ( 1993)
    The ability of cochlear implantees to detect amplitude modulation of pulsatile electrical stimulation, suggests that some speech feature information may be conveyed effectively by this means. For example, modulations at the fundamental frequency of speech may provide a voice pitch percept to implantees, particularly in speech processing strategies which generate constant-rate stimulation. The pitch evoked by sinusoidally modulated current pulse trains on a single electrodes has been studied. Modulation frequencies of 100, 150 and 200Hz, and carrier pulse rates varying from 200 to 1200Hz, were used. The results showed that the pitch of the stimulation was related to the modulation frequency, provided that either the carrier rate was a multiple of the modulation frequency, or the carrier rate was sufficiently high (at least four times the modulation frequency for the stimuli studied here). Furthermore, when the modulated stimuli were matched in pitch to non-modulated pulse trains, it was. found that the rate of the matched non-modulated stimuli was close to but somewhat higher than the modulation frequency. This difference depended on the carrier rate and varied among subjects.
  • Item
    Thumbnail Image
    Temporal pitch coding for cochlear implantees: the effects of carrier rate and amplitude modulation of pulsatile electrical stimuli [Abstract]
    McKay, Colette M. ; McDermott, Hugh J. ; Clark, Graeme M. ( 1993)