    Clark, Graeme M. ; Pyman, Brian C. ; Webb, R. L. ; Dowell, R. C. ; Staller, S. J. ; Beiter, A. L. ; Brimacombe, J. A. ( 1992)
    The performance of the Nucleus 22 channel cochlear implant has been assessed on 142 English speaking children who have worn their device for at least 12 months. The safety of the device has been evaluated on 309 children. A significant improvement for prosody was observed in 66%, for closed-set words in 63% and open-set words in 46%, using electrical stimulation alone. Performance over time increased, especially for open-set speech tests. Prelinguistically deaf children had similar scores to postlinguistically deaf children from prosody and closed-set word tests, but scores were not as good for open-set word tests. Lipreading enhancement was assessed using the CID sentence test, and the mean lipreading-alone score of 51% increased significantly to 71% when lipreading was combined with electrical stimulation. Speech intelligibility was determined with McGarr material and 63% were significantly more intelligible after 12 months' implant experience. There were 6-8% medical/surgical complications compared to 12% for a comparable group of adults. In 2.6% surgical intervention was required and this was primarily for infection or necrosis of the flap.