    The surgery for multiple-electrode cochlear implantations
    Clark, Graeme M. ; Pyman, Brian C. ; Bailey, Quentin R. (Cambridge University Press, 1979)
    The multiple-electrode hearing prosthesis designed in the Departments of Otolaryngology and Electrical Engineering (UMDOLEE) at the University of Melbourne (Clark et al., 1977) has been miniaturized with hybrid circuitry so that, if design changes are necessary as a result of initial patient testing, they can be made at minimal cost. This results, however, in increased package dimensions which makes its placement and the design of the surgery more critical. This problem is increased by the fact that we have considered it important to be able to remove the package and replace it with another without disturbing the implanted electrode array, should the first receiver-stimulator fail or an improved design become available. This has meant the design of a special connector (Patrick, 1977; Clark et al., 1978) which adds to the dimensions of the implanted unit. In addition the placement of the coils for transmitting power and information has to be considered. Not only is it desirable to site the coils at a convenient location behind the ear to facilitate the placement and wearing of the external transmitter, but there should also be no relative movement between the coils and the electronic package. These design considerations have led to the sitting of the coils on top of the hermetically-sealed box, and further increased the height of the package. The dimensions of the package shown in Fig. 1 are length 42 mm, width 32 mm, height of connector 8.5 mm, height of receiver-stimulato unit 13 mm. The surgical considerations discussed are the result of a number of temporal bone and cadaver dissections, and the surgical implantation at The Royal Victorian Eye and Ear Hospital of the UMDOLEE unit in a specially-selected patient.
  • Item
    Thumbnail Image
    A multiple-electrode cochlear implant
    Clark, Graeme M. ; Tong, Y. C. ; Bailey, Q. R. ; Black, R. C. ; Martin, L. F. ; Millar, J. B. ; O'Loughlin B. J. ; Patrick, J. F. ; Pyman, B. C. ( 1978)
    Interest in artificially stimulating the auditory nerve electrically for sensori-neural deafness was first sparked off by Volta in the 18th century. Count Volta, who was the first to develop the electric battery, connected up a number of his batteries to two metal rods which he inserted into his ears. Having placed the rods in his ears he pressed the switch and received "une secousse dans la tete" and perceived a noise like "the boiling of thick soup".
  • Item
    Thumbnail Image
    The development of the Melbourne/Cochlear multiple-channel cochlear implant for profoundly deaf children
    Clark, Graeme M. ; Busby, Peter A. ; Dowell, Richard C. ; Dawson, Pamella W. ; Pyman, Brian C. ; Webb, Robert L. ; Staller, Steven J. ; Beiter, Anne L. ; Brimacombe, Judith A. ( 1992)
    In 1978-79, a speech processing strategy which extracted the voicing (FO) and second formant (F2) frequencies and presented these as rate and place of stimulation respectively to residual auditory nerve fibres was developed for the University of Melbourne's prototype multiple-channel receiver-stimulator (Clark et aI1977, Clark et a11978, Tong et aI1980). This speech processing strategy was shown to provide post linguistically deaf adults with some open-set speech comprehension using electrical stimulation alone, and considerable help when used in combination with lipreading (Clark et al 1981).
  • Item
    Thumbnail Image
    Surgical complications with the cochlear multiple-channel intracochlear implant: experience at Hannover and Melbourne
    Webb, Robert L. ; Lehnhardt, Ernst ; Clark, Graeme M. ; Laszig, Roland ; Pyman, Brian C. ; Franz, Burkhard K-H. G. ( 1991)
    The surgical complications for the first 153 multi-channel cochlear implant operations carried out at the Medizinische Hochschule in Hannover and the first 100 operations at The University of Melbourne Clinic, The Royal Victorian Eye and Ear Hospital, are presented. In the Hannover experience the major complications were wound breakdown, wound infection, electrode tie erosion through the external auditory canal, electrode slippage, a persistent increase in tinnitus, and facial nerve stimulation. The incidence of wound breakdown requiring removal of the package was 0.6% in Hannover and 1.0% in Melbourne. The complications for the operation at both clinics were at acceptable levels. It was considered that wound breakdown requiring implant removal could be kept to a minimum by making a generous incision and suturing the flap without tension.
  • Item
    Thumbnail Image
    Multi-channel cochlear implants for children: the Melbourne Program
    Dawson, Pam W. ; Blamey, Peter J. ; Dettman, Shani J. ; Rowland, L.C. ; Brown, A. M. ; Dowell, Richard C. ; Pyman, B. C. ; Webb, R. L. ( 1991)
    Although there have been 300 years of deaf education, profoundly-totally deaf children today on average are not able to reach the same level of achievement as their normally hearing peers (Geers & Moog, 1989). This failure of deaf children to develop their true potential is largely due to the difficulty they have in communicating with normally hearing people. During the last 300 years there have been basically two different methods of education used (The New Encyclopaedia Britannica, 1983). Firstly, one which maximises auditory and lip reading cues (auditory/oral), advocated by Juan Pablo Bonet (1620), and one which uses a series of signs to convey meaning (signing), developed by Charles-Michel (1712-89). In addition, there is a method which endeavours to combine both auditory/oral and signing approaches called total communication. In practice, however, children taught by total communication tend to receive speech more predominantly by one or other of these methods.
  • Item
    Thumbnail Image
    The histopathology of the human temporal bone and auditory central nervous system following cochlear implantation in a patient: correlation with psychophysics and speech perception results
    Clark, Graeme M. ; Shepherd, Robert K. ; Franz, Burkhard K.-H. ; Dowell, Richard C. ; Tong, Yit C. ; Blamey, Peter J. ; Webb, Robert L. ; Pyman, Brian C. ; McNaughton, Judy ; Bloom, David M. ; Kakulas, Byron A. ; Siejka, Stan ( 1988)
    Cochlear implantation has become a recognised surgical procedure for the management of a profound-total hearing loss, especially in patients who have previously had hearing before going deaf (postlingual deafness). Nevertheless, it is important for progress in the field that patients who have had a cochlear implant, bequeath their temporal bones for research. This will then make it possible to further assess the safety of the procedure, and the factors that are important for its effectiveness. Biological safety has been assessed in a number of studies on animals, in particular, the biocompatibility of the materials used (1,2), the histopathological effects of long-term implantation on the cochlea (3, 4, 5, 6, 7, 8), and the effects of chronic electrical stimulation on the viability of spiral ganglion cells (9, 10, 11, 12). In studying the temporal bones of deceased cochlear implant patients it is possible to help establish that the animal experimental results are applicable to Man. Surgical trauma has been most frequently evaluated by inserting electrodes into cadaver temporal bones. It is important, however, to examine bones that have been previously implanted surgically to ensure that the cadaver findings are applicable to operations on patients. The effectiveness of cochlear implantation can be studied by correlating the histopathological findings, the dendrite and spiral ganglion cell densities, in particular, with the psychophysical and speech perception results. Other benefits also accrue, for example, establishing the accuracy of preoperative X-rays and electrical stimulation of the promontory in predicting cochlear pathology and spiral ganglion cell numbers. For the above reasons it has been especially interesting to examine both the temporal bones and central nervous system from one of our patients (patient 13) who participated in the initial clinical trial of the Cochlear Proprietary Limited (a member of the Nucleus group) multiple-electrode cochlear prosthesis, and who died due to a myocardial infarction following coronary bypass surgery.
  • Item
    Thumbnail Image
    The University of Melbourne/Nucleus cochlear prosthesis
    Clark, Graeme M. ; Blamey, P. J. ; Brown, A. M. ; Busby, P. A. ; Dowell, R. C. ; Franz, B. K-H. ; Millar, J. B. ; Pyman, B. C. ; Shepherd, R. K. ; Tong, Y. C. ; Webb, R. L. ; Brimacombe, J. A. ; Hirshorn, M. S. ; Kuzma, J. ; Mecklenburg, D. J. ; Money, D. K. ; Patrick, J. F. ; Seligman, P. M. ( 1988)
    This is a review of research to develop the University of Melbourne/Nucleus cochlear prosthesis for patients with a profound-total hearing loss. A more complete review can be obtained in Clark et al. A prototype receiver-stimulator and multiple-electrode array developed at the University of Melbourne was first implanted in a postlingually deaf adult patient with a profound-total hearing loss on 1 August 1978. A speech processing strategy which could help this patient understand running speech, especially when combined with lipreading was developed in 1978 following initial psychophysical studies. A prototype wearable speech processor was fabricated in 1979, that could provide significant help for the first two patients in understanding running speech when used in combination with lipreading compared with lipreading alone, and it also enabled them to understand some running speech when using electrical stimulation alone. An implantable receiver-stimulator and wearable speech processor embodying the principles of the prototype devices were then produced for clinical trial by the Australian biomedical firm, Nucleus Ltd, and its subsidiaries, Cochlear Pty Ltd and Cochlear Corporation. This cochlear implant was initially clinically trialled on six patients at The Royal Victorian Eye & Ear Hospital in 1982, and shown to give similar results to those obtained with the prototype device. In view of these findings a clinical trial was carried out for a Premarket Approval Application to the US Food and Drug Administration (FDA), and extended to a number of centres in the US, Canada, and West Germany. This clinical trial confirmed that patients could understand running speech when electrical stimulation was combined with lipreading, and that some patients could also understand running speech when using electrical stimulation alone. Today, more than 600 patients world-wide are using cochlear implants developed from the research described in this paper.
  • Item
    Thumbnail Image
    A multiple-electrode intracochlear implant for children
    Clark, Graeme M. ; Blamey, Peter J. ; Busby, Peter A. ; Dowell, Richard C. ; Franz, Burkhard K-H. ; Musgrave, Gaye Nicholls ; Nienhuys, Terry G. ; Pyman, Brian C. ; Roberts, Susan A. ; Tong, Yit C. ; Webb, Robert L. ; Kuzma, Januz A. ; Money, David K. ; Patrick, James F. ; Seligman, Peter M. ( 1987)
    A multiple-electrode intracochlear implant that provides 21 stimulus channels has been designed for use in young children. It is smaller than the adult version and has magnets to facilitate the attachment of the headset. It has been implanted in two children aged 5 and 10 years. The two children both lost hearing in their third year, when they were still learning language. Following implantation, it was possible to determine threshold and comfortable listening levels for each electrode pair. This was facilitated in the younger child by prior training in scaling visual and electrotactile stimuli. Both children are regular users of the implant, and a training and assessment program has been commenced.
  • Item
    Thumbnail Image
    Banded intracochlear electrode array: evaluation of insertion trauma in human temporal bones
    Shepherd, R. K. ; Clark, Graeme M. ; Pyman, B. C. ; Webb, R. L. ( 1985)
    A banded free-fit scala tympani array was inserted into a basal turn of nine human cochleas to evaluate the trauma produced by the procedure. These nine cochleas, together with five nonimplanted controls, were serially sectioned and examined microscopically for damage to the membranous labyrinth, in particular the spiral ligament, the basilar and Reissner’s membranes, the stria vascularis, and the osseous spiral lamina. The severity and location of any trauma along the cochlear spiral were recorded. The results indicate that the insertion of the banded scala tympani array resulted in minimal mechanical damage, occurring primarily to a localized region of the spiral ligament. This would not result in significant neural degeneration, and therefore would not compromise the efficacy of multichannel cochlear prosthesis.
  • Item
    Thumbnail Image
    The multi-channel cochlear implant
    Webb, R. L. ; Dowell, R. C. ; Clark, Graeme M. ; Pyman, A. M. ; Brown, Y. C. ; Tong, Y. C.. ; Seligman, P. M. ; Blamey, P.J. ; Xu, S. ( 1984/85)
    The multi-channel cochlear implant codes sounds on the bases of rate and place pitch. Experimental studies on animals and patients have shown it is difficult for electrical stimulation to code rate pitch above about 200-400 pulses/second. Therefore to convey as much information about speech as possible it is necessary to produce multi-channel stimulation or place pitch so that the important frequency cues in vowels and consonants can be perceived by the patient.