    Surgical implications of perimodiolar cochlear implant electrode design: avoiding intracochlear damage and scala vestibuli insertion.
    Briggs, RJ ; Tykocinski, M ; Saunders, E ; Hellier, W ; Dahm, M ; Pyman, B ; Clark, GM (Informa UK Limited, 2001-09)
    OBJECTIVE: To review the mechanisms and nature of intracochlear damage associated with cochlear implant electrode array insertion, in particular, the various perimodiolar electrode designs. Make recommendations regarding surgical techniques for the Nucleus Contour electrode to ensure correct position and minimal insertion trauma. BACKGROUND: The potential advantages of increased modiolar proximity of intracochlear multichannel electrode arrays are a reduction in stimulation thresholds, an increase in dynamic range and more localized neural excitation. This may improve speech perception and reduce power consumption. These advantages may be negated if increased intracochlear damage results from the method used to position the electrodes close to the modiolus. METHOD: A review of the University of Melbourne Department of Otolaryngology experience with temporal bone safety studies using the Nucleus standard straight electrode array and a variety of perimodiolar electrode array designs; comparison with temporal bone insertion studies from other centres and postmortem histopathology studies reported in the literature. Review of our initial clinical experience using the Nucleus Contour electrode array. RESULTS: The nature of intracochlear damage resulting from electrode insertion trauma ranges from minor, localized, spiral ligament tear to diffuse organ of Corti disruption and osseous spiral lamina fracture. The type of damage depends on the mechanical characteristics of the electrode array, the stiffness, curvature and size of the electrode in relation to the scala, and the surgical technique. The narrow, flexible, straight arrays are the least traumatic. Pre-curved or stiffer arrays are associated with an incidence of basilar membrane perforation. The cochleostomy must be correctly sited in relation to the round window to ensure scala tympani insertion. A cochleostomy anterior to the round window rather than inferior may lead to scala media or scala vestibuli insertion. CONCLUSION: Proximity of electrodes to the modiolus can be achieved without intracochlear damage provided the electrode array is a free fit within the scala, of appropriate size and shape, and accurate scala tympani insertion is performed.
  • Item
    Thumbnail Image
    Surgery
    Clark, Graeme M. ; Pyman, Brian C. ; Webb, Robert L. (Singular Publishing, 1997)
    Cochlear implant surgery should be undertaken only after the cochlear implant team has established that the child is not achieving useful communication with a hearing aid. This can be difficult because of poor language development in deaf children in this age group or because the child is at a preverbal stage and too young for the use of formal assessment tests. The child's unaided and aided thresholds, however, are important for assessment, as are his or her communication skills. These need to be evaluated by an experienced paediatric audiologist.
  • Item
    Thumbnail Image
    Preoperative medical evaluation
    Clark, Graeme M. ; Pyman, Brian C. (Singular Publishing, 1997)
    The aim of the medical assessment of infants and children is to determine the cause, severity and duration of any hearing loss as well as the presence of any medical conditions that may influence their management with a cochlear implant. There should also be an initial assessment of the child's communication skills and the parental expectations for his or her education.