Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 6 of 6
  • Item
    Thumbnail Image
    Chronic middle ear disease and cochlear implantation
    Donnelly, M. J. ; Pyman, B. C. ; Clark, Graeme M. ( 1995)
    Profound or total hearing loss can occur in the setting of chronic suppurative otitis media (CSOM), either coincidentally or secondary to the disease process. Obviously, inserting a foreign body through a potentially infected field into a space that communicates intracranially presents a challenging management problem. This paper presents the experience from the Melbourne Cochlear Implant Clinic (CIC) in implanting patients with bilateral CSOM. This is certainly not a common problem, as there have been only 3 cases from 121 implanted adults. However, we feel that it is an important issue with potentially devastating consequences. In addition, there are many countries in which bilateral CSOM is a more common problem and cause of profound or total hearing loss.
  • Item
    Thumbnail Image
    Surgical considerations for the placement of the new Cochlear Pty Limited micro-multiple-channel cochlear implant for research studies
    Clark, Graeme M. ; Pyman, B. C. ; Webb, R. L. ( 1995)
    A new micro-receiver-stimulator for research studies on very young children as well as adults has been developed by Cochlear Pty Limited. The dimensions of the device are length 58 mm and width at front 18 mm. The width starts to increase 19 mm from the front, and the back section, which has the receiver coil and magnet, has a maximum diameter of 33 mm. The depth or thickness of the front portion of the electronic package is 5.7mm, and the remainder of the package 4.0 mm. The antenna section is 3.5 mm thick. The other dimensions of the front section of the electronic package are length 9.5 mm and width 13.7 mm. There is an angle of 160° between the titanium electronic package and the receiver coil. The dimensions of the receiver-stimulator were arrived at after anatomic studies on the temporal bones of children ranging in age from 2 to 11 months.
  • Item
    Thumbnail Image
    The biologic safety of the Cochlear Corporation multiple-electrode intracochlear implant
    Webb, Robert L. ; Clark, Graeme M. ; Shepherd, Robert K. ; Franz, Burkhard K-H. ; Pyman, Brian C. ( 1988)
    Studies have been undertaken to confirm the biologic safety of the Cochlear Corporation multi-electrode intracochlear implant. The materials used are biocompatible. The electrode array is flexible: it can be inserted with minimal or no trauma, providing the insertion is stopped when resistance is first felt. An atraumatic insertion is facilitated if a good view is obtained along the scala tympani of the basal turn of the cochlea by drilling through the crista fenestrae. The passage of the electrode around the cochlea can be facilitated if the electrode is rotated during insertion (clockwise for the left and anticlockwise for the right cochlea). The electrode can be explanted and another one reinserted with minimal or no trauma. A seal established around the electrode after an implantation period of 2 weeks can prevent infection extending from the middle to the inner ear. The electrical stimulus parameters produced by the Nucleus receiver-stimulator cause no loss of spiral ganglion cells or corrosion of the platinum band electrodes. Long-term stimulation has been carried out for up to 8 years in patients without affecting their clinical performance.
  • Item
    Thumbnail Image
    Surgery for the safe insertion and reinsertion of the banded electrode array
    Clark, Graeme M. ; Pyman, B. C. ; Webb, R. L. ; Franz, B. K-H. G. ; Redhead, T. J. ; Shepherd, R. K. ( 1987)
    Adhering to the surgical technique outlined in the protocol for the Nucleus implant has resulted in over 100 patients worldwide obtaining significant benefit from multichannel stimulation. A detailed analysis of the results in 40 patients shows that it improves their awareness of environmental sounds and their abilities in understanding running speech when combined with lipreading. In addition, one third to one half of the patients also understand significant amounts of running speech without lipreading and some can have interactive conversations over the telephone. It is clear that any insertion trauma is not significant, which is confirmed by the excellent clinical results.
  • Item
    Thumbnail Image
    Preliminary results for the Cochlear Corporation multielectrode intracochlear implant in six prelingually deaf patients
    Clark, Graeme M. ; Busby, Peter A. ; Roberts, Susan A. ; Dowell, Richard C. ; Blamey, Peter J. ; Mecklenburg, Dianne J. ; Webb, Robert L. ; Pyman, Brian C. ; Franz, Burkhard K. ( 1987)
    The preliminary results from this study indicate that some prelingually deaf patients may get worthwhile help from a multiple-electrode cochlear implant that uses a formant-based speech processing strategy. It is encouraging that these improvements can occur in young adults and teenagers. The results for two children are also encouraging. A 10-year-old child obtained significant improvement on some speech perception tests. It was easy to set thresholds and comfortable listening levels on a 5-year-old child, and he is now a regular user of the device. There are, however, considerable variations in performance among the prelingual patients, which may be related to the following factors: whether they have had some hearing after birth, the method of education used, the motivation of the patient, and age at implantation.
  • Item
    Thumbnail Image
    Surgery for an improved multiple-channel cochlear implant
    Clark, Graeme M. ; Pyman, Brian C. ; Webb, Robert L. ; Bailey, Quentin E. ; Shepherd, Robert K. ( 1984)
    An improved multiple-channel cochlear implant has been developed. The titanium container with enclosed electronics, the receiver coil and the connector are embedded in medical-grade Silastic. The upper half of the implant has a diameter of 35 mm and a height of 4.5 mm. and the lower half a diameter of 23 mm and a height of.5 mm. The electrode array has also been designed to reduce the possibility of breakage due to repeated movements over many years. The surgery involves drilling a bed in the mastoid bone for the receiver-stimulator, and fixing the proximal electrode under the mastoid cortex. Gentle insertion of the electrode array through the round window and along the seala tympani is achieved with a specially designed microclaw.