  • Item
    Thumbnail Image
    Signal processing for multichannel cochlear implants: past, present and future [Abstract]
    DOWELL, RICHARD ; SELIGMAN, PETER ; MCDERMOTT, HUGH ; Whitford, Lesley ; BLAMEY, PETER ; Clark, Graeme M. ( 1994)
    Since the late 1970's, many groups have worked on developing effective signal processing for multichannel cochlear implants. The main aim of such schemes has been to provide the best possible speech perception for those using the device. Secondary aims of providing awareness and discrimination of environmental sounds and appreciation of music have also been considered. Early designs included some that attempted to simulate the normal cochlea. The application of such complex processing schemes was limited by the technology of the times. In some cases, researchers reverted to the use of single channel systems which could be controlled reliably with the existing technology. In other cases, as with the Australian implant, a simple multichannel processing scheme was devised that allowed a reliable implementation with available electronics. Over the next 15 years, largely due to the improvements in integrated circuit technology, the signal processors have slowly become more complex. Further psychophysical research has shown how additional information can be transferred effectively to implant users via electrical stimulation of the cochlea. This has lead to rapid improvement in the speech perception abilities of adults using cochlear implants. Some of the main developments in signal processing over the last 15 years will be discussed along with the latest speech perception results obtained with the new SPEAK processing scheme for the Australian 22-channel cochlear implant. Initial results for SPEAK show mean scores of 70% (equivalent to 85-90% phoneme scores) for open set monosyllabic word testing for experienced adult users. Although there remains a large range of performance for all users of cochlear implants, average speech perception scores for all implanted adults have also improved significantly with the developments in signal processing. It appears likely that multichannel cochlear implants will be a viable alternative for the treatment of severe hearing loss in the future.
  • Item
    Thumbnail Image
    Comparison of current speech coding strategies
    Whitford, L. A. ; Seligman, P. M. ; Blamey, Peter J. ; McDermott, H. J. ; Patrick, J. F. ( 1993)
    This paper reports on two studies carried out at the University of Melbourne jointly with Cochlear Pty Ltd. The studies demonstrated substantial speech perception improvements over the current Multipeak strategy in background noise.
  • Item
    Thumbnail Image
    Multichannel cochlear implantation in children: a summary of current work at The University of Melbourne
    Dowell, Richard C. ; Dawson, Pam W. ; Dettman, Shani J. ; Shepherd, Robert K. ; Whitford, Lesley A. ; Seligman, Peter M. ; Clark, Graeme M. ( 1991)
    This paper summarizes research work relating to multichannel cochlear implantation in children at the University of Melbourne. Ongoing safety studies relating to the implantation of young children are discussed. Results of these studies suggest that special design considerations are necessary for a prosthesis to be implanted in children under the age of 2 years. Results of clinical assessment of implanted children and adolescents are also discussed in terms of speech perception, speech production, and language development, and some possible predictive factors are suggested. Preliminary data suggests that a high proportion of young children can achieve open-set speech perception with the cochlear implant given appropriate training and support. Initial results with adults using new speech processing hardware and a new coding scheme are also presented. These suggest that improved speech perception in quiet and competing noise is possible with the new system.
  • Item
    Thumbnail Image
    Direct current measurements in cochlear implants: an in vivo and in vitro study
    Huang, Christie Q. ; Carter, Paul M. ; Shepherd, Robert K. ; Seligman, Peter M. ; Tabor, Bruce ; Clark, Graeme M. ( 1998)
    Direct current (DC) was measured both in vivo and in vitro in cochlear implant electrodes with stimulation at moderate to high pulse rates in monopolar and bipolar modes. In vivo DC was approximately 2-3 times higher than that measured in vitro. In vivo DC levels were <100 nA even at very high rates, although DC levels increased as a function of stimulus rate and charge intensity. DC levels were lower: in the monopolar than in the bipolar stimulation condition. Stimulation with a monopolar capacitively coupled extracochlear electrode showed even lower DC levels in the intracochlear .electrodes. Our results indicated that the Nucleus electrode shorting system is able to maintain a low level of DC during very high rate stimulation for both monopolar and bipolar modes.
  • Item
    Thumbnail Image
    Electrical stimulus induced changes in excitability of the auditory nerve
    Huang, C. O. ; Shepherd, Robert K. ; Seligman, P. M. ; Clark, Graeme M. ( 1997)
    High rate electrical stimulation of the auditory nerve using stimulus intensities well above the clinical limits can induce a significant reduction in the excitability of the auditory nerve as measured by a decrement in the amplitude of the electrically evoked auditory brainstem response (EABR). Two potential mechanisms may be associated with this stimulus induced reduction in activity: 1) stimulus induced prolonged neuronal hyperactivity; and 2) the generation of adverse electrochemical productions from the electrode surface. The purpose of the present study was to assess the extent to which adverse electrochemical damage contributes to the stimulus induced reduction in auditory nerve excitability. Twenty-six adult guinea pigs anaesthetized with ketamine (40 mg/kg i.p.) and xylazine (4 mglkg i.p.), were bilaterally implanted and unilaterally stimulated for two hours using a stimulus intensity of two or four times EABR threshold. Stimulus rates of 200, 400, or 1000 pulses/s (pps) were delivered via a standard platinum scala tympani electrode or large surface area ("high Q") platinum electrode.
  • Item
    Thumbnail Image
    Cochlear implants: high rate stimulation studies and the effect of electrode position [Abstract]
    Shepherd, R. K. ; Huang, C. Q. ; Xu, J. ; Tykocinski, M. ; Seligman, P. M. ; Clark, Graeme M. ( 1996)
    This paper summarizes our recent findings investigating the safety of high rate electrical stimulation, and reviews the effects of electrode position on auditory excitability. These studies used charge balanced biphasic pulses and electrode shorting between stimuli to minimize any residual charge or direct current. High rate (400-1000 pulses/s) electrical stimulation of the auditory nerve can result in significant stimulus induced reductions in auditory nerve excitability at stimulus levels well above those used clinically (1). The extent of this reduction was dependent on stimulus rate, intensity and duty cycle, implying that such changes were related to the degree of evoked activity.
  • Item
    Thumbnail Image
    Latest results and future directions in speech processing for the Nucleus multichannel cochlear prosthesis [Abstract]
    Dowell, R. ; Whitford, L. ; Seligman, P. ; Vandali, A. ; Hollow, R. ; Clark, Graeme M. ( 1995)
    The past two years has seen the introduction of the Speak speech encoding scheme for most patients using the Nucleus 22-channel cochlear prosthesis. This scheme, based on the Spectral Maxima Speech Processor (SMSP) developed at the University of Melbourne, uses a bank of 20 band-pass filters to present detailed spectral information to the intracochlear electrode array. Clinical trials of this speech processor have shown highly significant improvements over the previous Multipeak scheme in English, German, French and Japanese speaking patients. The largest improvements were evident for open-set testing in background noise, which represents a more realistic measure of everyday benefit than testing in quiet. The latest results for adults who have changed from Multipeak to Speak will be presented, along with results over time for newly-implanted patients using the Speak scheme. New research aimed at improving the speech processing in both the spectral and temporal domains will also be discussed.