Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 4 of 4
  • Item
    Thumbnail Image
    Speech perception using a two-formant 22-electrode cochlear prosthesis in quiet and in noise
    Dowell, Richard C. ; Seligman, Peter M. ; Blamey, Peter J. ; Clark, Graeme M. ( 1987)
    A new speech-processing strategy has been developed for the Cochlear Pty. Ltd. 22-electrode cochlear prosthesis which codes an estimate of the first formant frequency in addition to the amplitude. voice pitch and second formant frequencies. Two groups of cochlear implant patients were tested 3 months after implant surgery, one group (n= 13) having used the old (F0F2) processing strategy and the other (n=9) having used the new (F0FIF2) strategy. All patients underwent similar postoperative training programs. Results indicated significantly improved speech recognition for the F0FIF2 group particularly on open set tests with audition alone. Additional testing with a smaller group of patients was carried out with competing noise (speech babble). Results for a closed set spondee test showed that patient performance was significantly degraded at a signal-to-noise ratio of 10 dB when using the F0F2 strategy, but was not significantly affected with the F0FIF2 strategy.
  • Item
    Thumbnail Image
    A multi-channel hearing prosthesis for profound-to-total hearing loss
    Money, D. K. ; Clark, Graeme M. ; Tong, Y. C. ; Patrick, J. F. ; Seligman, P. M. ; Crosby, P. A. ; Kuzma, J. A. ( 1984)
    A multi-channel cochlear implant hearing prosthesis providing 22 separate channels of stimulation has been developed. The electronics for the implantable receiver-stimulator have been incorporated on a single chip, using digital circuits and employing CMOS technology. The chip is enclosed in a titanium capsule with platinum/ceramic electrode feed-throughs. A pocket-sized speech processor and directional microphone extract the following speech parameters: signal amplitude, fundamental frequency and formant frequency. The fundamental frequency is coded as electric pulse rate, and formant frequency by electrode position. The speech processor has been realized using hybrid circuits and CMOS gate arrays. The multi-channel prosthesis has undergone a clinical trial on four postlingually deaf patients with profound-total hearing losses. The speech perception results indicate that they were able to obtain open-set speech recognition scores for phonetically balanced words, CID sentences and spondees. In all cases the tests showed significant improvements when using the cochlear prosthesis combined with lipreading compared to lipreading alone.
  • Item
    Thumbnail Image
    A multi-channel cochlear prosthesis for profound-to-total hearing loss
    Money, D. K. ; Clark, Graeme M. ; Tong, Y. C. ; Patrick, J. F. ; Seligman, P. M. ; Crosby, P. A. ; Kuzma, J. A. ( 1984)
    A multi-channel cochlear prosthesis for profound-total hearing loss has been developed by the University of Melbourne and Nucleus Limited. Clinical trials have shown that the prosthesis provides significant help for postlingually deaf adult patients (lost hearing after normal language patterns have been established). The prosthesis helps the patients understand running speech when combined with lipreading, and a proportion obtain significant open-set speech scores for electrical stimulation alone. The patients with these open-set score can use the device in situations where lipreading is not possible, for example, to converse over the telephone. The prosthesis consists of an externally worn, pocket-sized speech-processor, a headset and an implanted receiver/stimulator and electrode array. The headset contains an .ear-level directional microphone that picks up the speech signal. The speech processor encodes the speech as a series of electrical pulses on the electrode array. The data describing these pulses and the power required to produce them, are sent to the receiver/stimulator from .a radio-frequency coil mounted on the headset. The receiver/stimulator decodes the data and delivers the speech signal as a series of biphasic electrical pulses to the 22 electrodes which have been gently passed along the scala tympani during implantation.
  • Item
    Thumbnail Image
    Clinical trial of a multi-channel cochlear prosthesis: results on 10 postlingually deaf patients
    Clark, Graeme M. ; Dowell, R. C. ; Pyman, B. C. ; Brown, A. M. ; Webb, R. L. ; Tong, Y. C. ; Bailey, Q. ; Seligman, P. M. ( 1984)
    The clinical trial of a multi-channel cochlear prosthesis has been carried out on 10 profoundly-totally deaf adult patients. Speech perception tests have shown that all the patients received significant benefit from the device. They obtained improvements in understanding running speech from 47% to 550% when using the device in conjunction with lipreading compared to lipreading alone. With an open-set CID sentence test, three patients obtained scores showing an ability to understand speech without the need to lipread, and a further three patients had scores indicating they could also receive useful information without lipreading. In two patients, very limited open-set scores for electrical stimulation alone were obtained. This was most probably due to the fact that only a few channels of stimulation were possible due to cochlear disease and they were therefore receiving information more like a single-channel device. The prosthesis has also been found to provide considerable help in hearing and recognizing everyday sounds.