- Graeme Clark Collection
Graeme Clark Collection
Permanent URI for this collection
Search Results
Now showing
1 - 2 of 2
-
ItemResults for two children using a multiple-electrode intracochlear implantBusby, P. A. ; Tong, Yit C. ; Roberts, S. A. ; Altidis, P. M. ; Dettman, S. J. ; Blamey, Peter J. ; Clark, Graeme M. ; Watson, R. K. ; Rickards, Field W. ( 1989)Abstract not available due to copyright.
-
ItemSurgery for an improved multiple-channel cochlear implantClark, Graeme M. ; Pyman, Brian C. ; Webb, Robert L. ; Bailey, Quentin E. ; Shepherd, Robert K. ( 1984)An improved multiple-channel cochlear implant has been developed. The titanium container with enclosed electronics, the receiver coil and the connector are embedded in medical-grade Silastic. The upper half of the implant has a diameter of 35 mm and a height of 4.5 mm. and the lower half a diameter of 23 mm and a height of.5 mm. The electrode array has also been designed to reduce the possibility of breakage due to repeated movements over many years. The surgery involves drilling a bed in the mastoid bone for the receiver-stimulator, and fixing the proximal electrode under the mastoid cortex. Gentle insertion of the electrode array through the round window and along the seala tympani is achieved with a specially designed microclaw.