Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 7 of 7
  • Item
    Thumbnail Image
    Issues in long-term management of children with cochlear implants and tactile devices [Abstract]
    COWAN, ROBERT ; DOWELL, RICHARD ; Barker, Elizabeth ; GALVIN, KARYN ; DETTMAN, SHANI ; SARANT, JULIA ; RANCE, GARY ; Hollow, Rod ; BLAMEY, PETER ; Clark, Graeme M. ( 1994)
    For many children with severe and profound hearing losses, conventional hearing aids are unable to provide sufficient amplification to ensure good oral communication and/or in the case of very young children, development of speech and language. Traditionally a number of these children have opted for the use of sign language alone or in Total Communication approaches as a primary means of communication. The advent of multiple channel cochlear implants for children and the continuing development of multiple channel speech processing tactile devices provide auditory approaches to resolving communication difficulties for these children. The successful use of such devices depends on a number of factors including the information provided through the aid; the ease of use, convenience and reliability of the aid; the individual communication needs of the child; and the habilitation and management program used with the device. Long-term data has shown that children continue to show increased speech perception benefits from improvements in speech processing and from further experience with these devices. Habilitation and management programs must therefore be geared to meet the changing needs of children as they progress and of families as children mature and face new challenges. Habilitation must address specific individual needs in speech perception and in speech production. For very young children, benefits of improved speech perception should have an impact on the development of speech and language, and habilitation and management must emphasise the need for language growth.
  • Item
    Thumbnail Image
    Combined cochlear implant and speech processing hearing aid for implant users with a severe to profound hearing loss in the contralateral ear [Abstract]
    BLAMEY, PETER ; Parisi, Elvira ; Dooley, Gary ( 1994)
    The bimodal device was developed for cochlear implant users who simultaneously wear a hearing aid in the opposite ear having residual hearing of a severe to profound degree. The aim was to create a single device to provide both input signals in a more compatible manner and thus maximise use of the individual's total hearing capabilities. The acoustic component of the bimodal device is very flexible and can implement various speech processing strategies with speed, ease and precision. The Frequency Response Tailoring strategy utilises three filters to fit a frequency gain curve to within 1-2 dB of that desired. Modifications at discrete frequencies, ranges or slopes can be readily made. The Peak Sharpening or Spectral Enhancement strategy amplifies the formant peaks in speech for potential improvement of formant resolution and speech perception in the presence of background noise. The Resynthesis strategy presents specifically selected components of speech in selected combinations and includes the ability to transpose higher frequency information to lower frequency ranges for individuals with no aidable high frequency hearing levels. Different fits can be quickly and easily interchanged for comparison and evaluation and subsequent modifications indicated can be readily effected. Any combination of acoustic and implant speech processing strategy can be presented to optimise speech perception for the individual user.
  • Item
    Thumbnail Image
    Psychophysics of electrical stimulation of the auditory nerve: implications for coding of sound and speech processing for cochlear implants [Keynote address]
    Clark, Graeme M. ( 1994)
    Psychophysical studies on electrical stimulation of the auditory nerve have contributed to our understanding of the coding of sound and speech signals. Those studies have also helped establish speech processing strategies for multiple-electrode cochlear implant patients. The first studies were on temporal coding of frequency and pitch perception to help determine whether a single or multiple electrode implant would be preferable for the coding of speech frequencies. Temporal frequency coding was initially studied in the experimental animal by measuring difference limens for frequency of stimulus rate. The results showed that rate coding occurs for low frequencies up to 200 or even 600 pulses per second. It was concluded that higher speech frequencies cannot be conveyed by variations in stimulus rate but require multiple-electrode stimulation. These studies in experimental animals were essentially confirmed in the human.
  • Item
    Thumbnail Image
    Evaluation of a new Spectral Peak coding strategy for the Nucleus 22 channel cochlear implant system
    Skinner, Margaret W. ; Clark, Graeme M. ; Whitford, Lesley A. ; Seligman, Peter M. ; Staller, Steven J. ; Shipp, David B. ; Shallop, Jon K. ; Everingham, Colleen ; Menapace, Christine M. ; Arndt, Patti L. ; Antogenelli, Trisha ; Brimacombe, Judith A. ; Pijl, Sipke ; Daniels, Paulette ; George, Catherine R. ; McDermott, Hugh J. ; Beiter, Anne L. ( 1994)
    Sixty-three postlinguistically deaf adults from four English-speaking countries participated in a 17-week field study of performance with a new speech coding strategy, Spectral Peak (SPEAK), and the most widely used strategy, Multipeak (MPEAK), both of which are implemented on wearable speech processors of the Nucleus 22 Channel Cochlear Implant System; MPEAK is a feature-extraction strategy, whereas SPEAK is a filterbank strategy. Subjects' performance was evaluated with an experimental design in which use of each strategy was reversed and replicated (ABAB). Average scores for speech tests presented sound-only at 70 dB SPL were higher with the SPEAK strategy than with the MPEAK strategy. For tests in quiet, mean scores for medial vowels were 74.8 percent versus 70.1 percent; for medial consonants, 68.6 percent versus 56.6 percent; for monosyllabic words, 33.8 percent versus 24.6 percent; and for sentences, 77.5 percent versus 67.4 percent. For tests in noise, mean scores for Four-Choice Spondees at +10 and +5 dB signal-to-noise ratio (S/N) were 88.5 percent versus 73.6 percent and 80.1 percent versus 62.3 percent, respectively; and for sentences at +15 dB, +10, and +5 dB S/N, 66.5 percent versus 43.4 percent, 61.5 percent versus 37.1 percent, and 60.4 percent versus 31.7 percent, respectively. Subjects showed marked improvement in recognition of sentences in noise with the new SPEAK filterbank strategy. These results agree closely with subjects' responses to a questionnaire on which approximately 80 percent reported they heard best with the SPEAK strategy for everyday listening situations.
  • Item
    Thumbnail Image
    Loudness growth characteristics of cochlear implantees using the Spectral Maxima Sound Processor [Abstract]
    MCDERMOTT, HUGH ; MCKAY, COLETTE ( 1994)
    The study of perceptual characteristics of subjects with cochlear implants can lead to improvements in the design of speech processors. One important aspect of speech processing which has received little attention in the past is the conversion acoustic signal amplitudes into appropriate levels of electrical stimulation. The optimum conversion would provide implantees with loudness growth characteristics that mimic those of normal hearing. To investigate how implantees using the Spectral Maxima Sound Processor (SMSP) perceive changes in loudness, an experiment involving production of fixed loudness ratios was conducted. Ten subjects participated: five users of the Mini System 22 cochlear implant, and five normally-hearing subjects. In the experiment, the subjects were required to adjust the loudness of two stimuli (white noise and speech-weighted noise) to equal half or twice that of a reference. The reference was presented at various levels over a range of 25 to 75 dBA. The results for three of the implantees were similar to those of all the normally-hearing subjects, who produced an average level change of 10.8 dB for the task. The remaining subjects, who had the largest electrical dynamic ranges, produced larger level changes (up to 20 dB) which were constrained by the limited electrical dynamic range of the processor (46 dB). The SMSP utilises an amplitude conversion function by which the stimulus level (in dB) is directly proportional to the input sound level (in dB). The experimental results suggest that the shape of this function is satisfactory, though not necessarily optimum, for these implantees.
  • Item
    Thumbnail Image
    Speech perception benefits for children using an advanced cochlear implant speech processing strategy in quiet and in noise [Abstract]
    Dettman, Shani J. ; Skok, Marissa ; Dowell, Richard C. ; Clark, Graeme M. ; Cowan, Robert S. C. ; Whitford, Lesley A. ; Sarant, Julia Z. ; Galvin, Karyn L. ; Barker, Elizabeth J. ; King, Alison ( 1994)
    A new speech processing strategy (SPEAK) has been developed by the University of Melbourne and Cochlear Pty Ltd for use with the Nucleus 22-channel electrode array. In this strategy, 20 programmable filters are repetitively scanned at an average rate of 250Hz and the largest spectral components or maxima are selected from the incoming speech signal. This new speech processing strategy has been shown to provide significantly improved benefits in adult implant patients, particularly in the presence of background noise. This report presents data of a preliminary paediatric clinical trial of the new SPEAK speech processing strategy.
  • Item
    Thumbnail Image
    Physiological and histopathological response of the cochlea to chronic electrical stimulation of the auditory nerve at high stimulus rates [Abstract]
    Shepherd, R. K. ; Xu, J. ; Clark, Graeme M. ( 1994)
    Previous research has shown that chronic electrical stimulation of the auditory nerve using charge balanced biphasic current pulses at rates of up to 500 pulses per second (pps) does not adversely affect the adjacent spiral ganglion population. More recently, a number of clinical trials have suggested that speech processing strategies based on high pulse rates (e.g. 1000 pps), can further improve speech perception. In the present study we evaluated the physiological and histopathological response of the cochlea following long-term stimulation using rates of 1000 pps. Thirteen normal hearing cats were bilaterally implanted with scala tympani electrodes and unilaterally stimulated using 25-50 �s per phase charge balanced biphasic current pulses presented at 1000 pps. Additional charge balance was achieved by shorting the electrodes between current pulses. Each animal was stimulated for periods ranging from 700 - 2100 hours at current levels within its dynamic range. Auditory brainstem responses to both acoustic (ABR) and electrical (EABR) stimuli were periodically recorded throughout the chronic stimulation program. At completion of the program the cochleas were prepared for histological examination. While all animals exhibited an increase in acoustic thresholds following surgery, click evoked ABR's returned to near normal levels in half the animals. Frequency specific stimuli indicated that the most extensive hearing loss occurred adjacent to the array (>12 kHz) while lower frequency thresholds appeared at or near normal Our EABR data showed that the majority of animals exhibited slight increases in threshold, although response amplitudes remained very stable for the duration of the stimulus program. The physiological data reported here will be correlated with cochlear histopathology. These initial findings suggest that chronic intracochlear electrical stimulation at high pulse rates, using a carefully designed charge balanced stimulator, does not appear to adversely affect the implanted cochlea.