    Speech perception in implanted children: influence of preoperative residual hearing on outcomes [Abstract]
    Cowan, R. S. C. ; Barker, E. J. ; Pegg, P. ; Dettman, S. ; Rennie, M. ; Galvin, K. ; Meskin, T. ; Rance, G. ; Cody, K. ; Sarant, J. ; Larratt, M. ; Latus, K. ; HOLLOW, RODNEY ; Rehn, C. ; Dowell, R. C. ; Pyman, B. ; Gibson, W. P. R. ; Clark, Graeme M. ( 1998)
    Since the first child was implanted with the Nucleus 22-channel prosthesis in Melbourne in 1985, several thousand children world-wide have now benefitted from this technology. More effective paediatric assessment and management procedures have now been developed, allowing cochlear implants to be offered to children under the age of 2 years. Improvements in speech processing strategy have also been implemented in the Nucleus implant system, resulting in increased mean speech perception benefits for implanted adults. Although a range of performance on formal measures of hearing, speech or language has been reported for children using implants, results from the first decade of implant experience consistently show that significant benefits are available to children receiving their implant at an early age. Reported speech perception results for implanted children show that a considerable proportion (60%) of paediatric patients in the Melbourne and Sydney clinics are able to understand some open-set speech using electrical stimulation alone. These results, and the upward trend of mean speech perception benefits shown for postlinguistically deafened adults have raised questions as to whether severely, or severely-to-profoundly deaf children currently using hearing aids would in fact benefit more from a cochlear implant. To investigate the potential influence of the degree of preoperative residual hearing on postoperative speech perception, results for all implanted children in the Melbourne and Sydney cochlear implant programs were analysed. Results showed that as a group, children with higher levels of preoperative residual hearing were consistently more likely to achieve open-set speech perception benefits. Potential factors in this finding could be higher levels of ganglion cell survival or greater patterning of the auditory pathways using conventional hearing aids prior to implantation. Conversely, children with the least preoperative residual hearing were less predictable, with some children achieving open-set perception, and others showing more limited closed-set benefits to perception. For these children, it is likely that preoperative residual hearing is of less significance than other factors in outcomes.
  • Item
    Thumbnail Image
    Comparison of the SPEAK (Spectral Maxima) and multipeak speech processing strategies and improved speech perception in background noise
    Clark, Graeme M. ; Whitford, L. ; Van Hoesel, R. ; McKay, C. M. ; McDermott, H. D. ; Seligman, P. ; Vandali, A. ; Pyman, B. C. ; Cowan, R. C. ( 1995)
    As more is known about speech processing for Cochlear Implant patients, results should continue to improve. It now appears possible that Cochlear Implant patients may, in some instances, reach performance levels that are better than those obtained by most severely deaf people who use hearing aids.
  • Item
    Thumbnail Image
    Speech perception in children using the advanced Speak speech-processing strategy
    Cowan, R. S. C. ; Brown, C. ; Whitford, L. A. ; Galvin, K. L. ; Sarant, J. Z. ; Barker, E. J. ; Shaw, S. ; King, A. ; Skok, M. ; Seligman, P. M. ; Dowell, R. C. ; Everingham, C. ; Gibson, W. P. R. ; Clark, Graeme M. ( 1995)
    The Speak speech-processing strategy, developed by the University of Melbourne and commercialized by Cochlear Pty Limited for use in the new Spectra 22 speech processor, has been shown to provide improved speech perception for adults in both quiet and noisy situations. The present study evaluated the ability of children experienced in the use of the Multipeak (Mpeak) speech-processing strategy (implemented in the Nucleus Minisystem-22 cochlear implant) to adapt to and benefit from the advanced Speak speech-processing strategy (implemented in the Nucleus Spectra 22 speech processor). Twelve children were assessed using Mpeak and Speak over a period of 8 months. All of the children had over 1 year's previous experience with Mpeak, and all were able to score significantly on open-set word and sentence tests using the cochlear implant alone. Children were assessed with both live-voice and recorded speech materials, including Consonant-Nucleus-Consonant monosyllabic words and Speech Intelligibility Test sentences. Assessments were made in both quiet and in noise. Assessments were made at 3-week intervals to investigate the ability of the children to adapt to the new speech-processing strategy. For most of the children, a significant advantage was evident when using the Speak strategy as compared with Mpeak. For 4 of the children, there was no decrement in speech perception scores immediately following fitting with Speak. Eight of the children showed a small (10% to 20%) decrement in speech perception scores for between 3 and 6 weeks following the changeover to Speak. After 24 weeks' experience with Speak, 11 of the children had shown a steady increase in speech perception scores, with final Speak scores higher than for Mpeak. Only 1 child showed a significant decrement in speech perception with Speak, which did not recover to original Mpeak levels.
  • Item
    Thumbnail Image
    Psychophysical and speech perception studies: a case report on a binaural cochlear implant subject
    van Hoesel, R. J. M. ; Tong, Y. C. ; Hollow, R. D. ; Clark, Graeme M. ( 1993)
    Abstract not available due to copyright.
  • Item
    Thumbnail Image
    Speech processing for cochlear implants
    Tong, Y. C. ; Millar, J. B. ; Blamey, P. J. ; Clark, Graeme M. ; Dowell, R. C. ; Patrick, J. F. ; Seligman, P. M. (JAI Press Ltd, 1992)
    The cochlear implant is a hearing prosthesis designed to replace the function of the ear. The operation of the prosthesis can be described as a sequence of four functions: the processing of the acoustic signal received by a microphone; the transfer of the processed signal through the skin; the creation of neural activity in the auditory nerve; and the integration of the experience of this neural activity into the perceptual and cognitive processing of the implantee.
  • Item
    Thumbnail Image
    An analysis of high rate speech processing strategies using the Nucleus 24 cochlear implant [Abstract]
    Vandali, A. E. ; Grayden, D. B. ; Whitford, L. A. ; Plant, K. L. ; Clark, Graeme M. ( 1998)
    The effects of varying the stimulation rate on speech perception was evaluated in five postlinguistically deaf adult users of the Nucleus 24 Cochlear Implant System. Three different rates of electrical stimulation, 250, 807, and 1615 pulses per second per channel were employed. For the high stimulus rate the analysis frequency was the same as for the medium rate condition. The study investigated the effect of varying rate of stimulation when using the electrode selection technique of the SPEAK strategy. The study used a repeated ABC experimental design, in order to account for learning effects and to minimize ordering effects. Speech perception was evaluated using both monosyllabic words (open-sets of CNC words in quiet) and sentence materials (open-sets of CUNY sentences at signal-to-noise ratios from +20 to 0dB). In addition, the subjects' perception of closed-sets of 19 vowels and 24 consonants, presented in the H/VID and A/C/A context, were also investigated. The recognition and perception of distinctive features were assessed across strategies and patients. Preliminary speech perception results have shown no statistically significant difference in performance between the low and medium stimulation rates. However, significantly poorer results were observed for the high rate condition for some tests with some individuals. Individual differences may be explained by the effects of rate of stimulation on speech features.
  • Item
    Thumbnail Image
    An analysis of high rate speech processing strategies using the Nucleus 24 cochlear implant [Abstract]
    Vandali, A. E. ; Grayden, D. B. ; Whitford, L. A. ; Plant, K. L. ; Clark, Graeme M. ( 1998)
    Speech comprehension for a group of five users of the Nucleus 24 Cochlear Implant system was explored at three rates of electrical stimulation, 250, 807, and 1615 pulses per second per channel. For the high stimulus rate the analysis frequency was the same as for the medium rate condition. The study investigated the effect of varying rate of stimulation when using the electrode selection technique of the SPEAK strategy. This has been undertaken using a repeated ABC experimental design to account for learning and minimize ordering effects. Speech perception was assessed using open-sets of CNC words in quiet and open-sets of CLTNY sentences at signal-to-noise ratios from +20 to 0dB. Closed-sets of 19 vowels and 24 consonants were also presented, in the H/V/D and A/C/A context. The recognition and perception of distinctive features were assessed across strategies and patients. Preliminary speech perception results have shown no statistically significant difference in performance between the low and medium stimulation rates. However, significantly poorer results were observed for the high rate condition for some tests. Individual differences may be explained by the effects of rate of stimulation on speech features.
  • Item
    Thumbnail Image
    Speech perception in implanted children: effects of speech processing strategy and residual hearing
    Meskin, T. ; Rance, G. ; Cody, K. ; Sarant, J. ; Larratt, M. ; Latus, K. ; Hollow, R. ; Rehn, C. ; Dowell, R.C. ; Pyman, B. ; Gibson, W.P.R. ; Clark, Graeme M. ; Cowan, Robert S. C. ; Barker, E. J. ; Pegg, P. ; Dettman, S. ; Rennie, M. ; Galvin, K. (Mendoza Editor, 1997)
    The ability of implanted children to adapt to different speech processing strategies has been demonstrated for the Nucleus implant system. Children previously experienced with the Multipeak speech processing strategy. were able to gain significant improvements in consonant, word and sentence perception using the Speak speech processing strategy. suggesting some degree of neural plasticity in neural-auditory coding. Of 192 implanted children with different degrees of preoperative residual hearing, 65% were found to obtain significant scores on open-set speech materials using electrical stimulation alone. Those children with more residual hearing had a greater probability of achieving open-set understanding and at a minimum level, perceived high frequency consonant information which would not have been available through conventional hearing aids.
  • Item
    Thumbnail Image
    Speech perception in implanted children: effects of preoperative residual hearing and speech processing strategy [Abstract]
    Meskin, T. ; Rance, G. ; Cody, K. ; Sarant, J. ; Larratt, M. ; Latus, K. ; HOLLOW, RODNEY ; Rehn, C. ; Dowell, R.C. ; Pyman, B. ; Gibson, W.P.R. ; Clark, Graeme M. ; Cowan, Robert S. C. ; Barker, E. J. ; Pegg, P. ; Dettman, S. ; Rennie, M. ; Galvin, K. ( 1997)
    Since the first child was implanted with the Nucleus 22-channel cochlear prosthesis in Melbourne in 1985, the number of implanted children world-wide has rapidly expanded. Over this period, more effective paediatric assessment and management procedures have developed, allowing cochlear implants to be offered to children under the age of 2 years. In addition, a succession of improved speech processing strategies have been implemented in the Nucleus implant system, resulting in increased mean speech perception benefits for implanted adults. Research in the Melbourne and Sydney Cochlear Implant Clinics has also demonstrated that young children can adapt to and benefit from improved speech processing strategies such as the Speak strategy. Reported speech perception results for implanted children show that a considerable proportion (60%) of paediatric patients in the Melbourne and Sydney clinics are able to understand some open set speech using electrical stimulation alone. These results, and the upward trend of speech perception benefits to improve over time with advances in speech processing. have raised questions as to whether severely, or severely-to-profoundly deaf children currently using hearing aids would in fact benefit more from a cochlear implant. To investigate the potential effect of the level of preoperative residual hearing on postoperative speech perception. results for all implanted children in the Melbourne and Sydney cochlear implant programs were analysed. Results showed that as 8 group, children with higher levels of preoperative residual hearing were consistently more likely to achieve open-set speech perception benefits. Potential factors in this finding could be higher levels of ganglion cell survival or greater patterning of the auditory pathways using conventional hearing aids prior to implantation. Conversely, children with the least preoperative residual hearing were less predictable, with some children achieving open-set perception, and others showing more limited closed-set benefits to perception. For these children, it is likely that preoperative residual hearing is of less significance than other factors in outcomes.
  • Item
    Thumbnail Image
    Speech perception benefits for children using an advanced cochlear implant speech processing strategy in quiet and in noise [Abstract]
    Dettman, Shani J. ; Skok, Marissa ; Dowell, Richard C. ; Clark, Graeme M. ; Cowan, Robert S. C. ; Whitford, Lesley A. ; Sarant, Julia Z. ; Galvin, Karyn L. ; Barker, Elizabeth J. ; King, Alison ( 1994)
    A new speech processing strategy (SPEAK) has been developed by the University of Melbourne and Cochlear Pty Ltd for use with the Nucleus 22-channel electrode array. In this strategy, 20 programmable filters are repetitively scanned at an average rate of 250Hz and the largest spectral components or maxima are selected from the incoming speech signal. This new speech processing strategy has been shown to provide significantly improved benefits in adult implant patients, particularly in the presence of background noise. This report presents data of a preliminary paediatric clinical trial of the new SPEAK speech processing strategy.