Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 10 of 27
  • Item
    Thumbnail Image
    Cognitive processing in children using cochlear implants: the relationship between visual memory, attention, and executive functions and developing language skills
    Surowiecki, Vanessa N. ; SARANT, JULIA ; MARUFF, PAUL ; Blamey, Peter J. ; Busby, Peter A. ; Clark, Graeme M. ( 2002)
    We performed this study to determine whether children using a cochlear implant performed differently from age- and gender-matched hearing aid users on 8 neuropsychological measures of visual memory, attention, and executive functioning. The study also examined whether differences in cognitive skills could account for some of the observed variance in speech perception, vocabulary, and language abilities of hearing-impaired children. In contrast to previous studies, our results revealed no significant cognitive differences between children who use a cochlear implant and children who use hearing aids. Partial correlation analysis indicated that the children’s visual memory skills, i.e., their recognition memory, delayed recall, and paired associative learning memory skills, correlated significantly with their language skills. When examined at a significance level of .01, attention and executive functioning skills did not relate to the children’s developing speech perception, vocabulary, or language skills. The results suggested that differences in visual memory skills may account for some of the variance seen in the language abilities of children using implants and children using hearing aids.
  • Item
    Thumbnail Image
    Speech perception outcomes in older children who use multichannel cochlear implants: Older is not always poorer
    Dowell, RC ; Dettman, SJ ; Hill, K ; Winton, E ; Barker, EJ ; Clark, GM (ANNALS PUBL CO, 2002-05)
    Speech perception outcomes for early-deafened children who undergo implantation as teenagers or young adults are generally reported to be poorer than results for young children. It is important to provide appropriate expectations when counseling adolescents and their families to help them make an informed choice regarding cochlear implant surgery. The considerable variation of results in this group makes this process more difficult. This study considered a number of factors in a group of 25 children who underwent implantation in Melbourne between the ages of 8 and 18 years. Each subject completed open-set speech perception testing with Bamford-Kowal-Bench sentences before and after implantation and preoperative language testing with the Peabody Picture Vocabulary Test. Data were collected regarding the type of hearing loss, age at implantation, age at hearing aid fitting, audiometric details, and preoperative and postoperative communication mode. Results were submitted to a stepwise multiple linear regression analysis with postoperative open-set sentence scores as the dependent variables. The analysis suggested that 3 factors have a significant predictive value for speech perception after implantation: preoperative open-set sentence score, duration of profound hearing loss, and equivalent language age. These 3 factors accounted for 66% of the variance in this group. The results of this study suggest that children who have useful speech perception before implantation, and higher age-equivalent scores on language measures, would be expected to do well with a cochlear implant. Consistent with other studies, a shorter duration of profound hearing loss is also advantageous. The mean sentence score for this group, 47%, was not significantly different from the mean result across all children in the Melbourne program.
  • Item
    Thumbnail Image
    Speech perception in children using cochlear implants: prediction of long-term outcomes.
    Dowell, RC ; Dettman, SJ ; Blamey, PJ ; Barker, EJ ; Clark, GM (Informa UK Limited, 2002-03)
    A group of 102 children using the Nucleus multichannel cochlear implant were assessed for open-set speech perception abilities at six-monthly intervals following implant surgery. The group included a wide range of ages, types of hearing loss, ages at onset of hearing loss, experience with implant use and communication modes. Multivariate analysis indicated that a shorter duration of profound hearing loss, later onset of profound hearing loss, exclusively oral/aural communication and greater experience with the implant were associated with better open-set speech perception. Developmental delay was associated with poorer speech perception and the SPEAK signal coding scheme was shown to provide better speech perception performance than previous signal processors. Results indicated that postoperative speech perception outcomes could be predicted with an accuracy that is clinically useful.
  • Item
    Thumbnail Image
    Cochlear implants: a personal scientific journey [Abstract]
    Clark, Graeme M. ( 2002)
    Electrical stimulation of the auditory system to reproduce hearing commenced through academic curiosity, and the hope of helping deaf people. It received direction from neurophysiology, and later psychophysics and speech science. In the 1960s and 1970s there were many questions requiring answers before cochlear implants could become a practical reality. Key concerns were: (1) the cochlea was too complex for electrical stimulation to reproduce the coding of sound; (2) multiple electrodes inserted into the cochlea for the place coding of frequency could damage the auditory nerves to be stimulated; (3) speech was too complex to be reproduced by electrical stimulation; and (4) children born deaf would not develop the appropriate neural connectivity for speech understanding. The first questions were addressed on the experimental animal. Speech research on patients was only possible with the advent of silicon chip technology allowing the development of an implantable receiver-stimulator package. Initial research established proof of principle that connected discourse was possible with multiple electrode stimulation of the auditory nerve in severely and profoundly deaf people. The research has been developed industrially for the benefits to be provided on a widespread basis through clinics worldwide. Further research has resulted in continuing improvements so that the average profoundly deaf person can hear as well as someone with severe hearing loss using a hearing aid. There is still much research required to achieve high fidelity sound, hearing in noise, and totally implantable devices.
  • Item
    Thumbnail Image
    Cochlear implants for adults and children
    Clark, Graeme M. (Martin Dunitz, 2002)
    Cochlear implants which use multiple-electrode speech-processing strategies are now established clinical entity for children and adults, as a result, preoperative selection and (re)habilitation are key issues. It is hard to realize that it was only in the 1960s and 1970s when many scientists and clinicians said that successful cochlear implants were not possible in the foreseeable future. The questions that had to be addressed by a multi disciplinary research effort are discussed, and the solutions achieved from the University of Melbourne's perspective are presented. However, the main aim of this chapter is to focus on preoperative selection, and (re)habilitation, including the results obtained. These issues are discussed primarily with reference to data from the University of Melbourne's Cochlear Implant Clinic at the Royal Victorian Eye and Ear Hospital. As this is a book on audiological medicine only, an overview of surgical principles is presented. The surgical management of the patient is, of course, very important, so for more details the reader is referred elsewhere. Cochlear implantation has also been the subject of quite intense ethical debate, particularly over its use for children. For this reason, a discussion of ethical issues is included. Finally, the chapter concludes with a vision of research in the next Millennium.
  • Item
    Thumbnail Image
    Histopathology of the binaural cochlear implant subject [Abstract]
    Yukawa, K. ; O'Leary, S. J. ; Clark, Graeme M. ( 2001)
    Binaural hearing improves speech reception in noise, and is necessary for sound localisation. Normal hearing subjects use both interaural time, and intensity, differences to localise sound. This study investigates why sound localisation in bilateral cochlear implantees is insensitive to interaural time differences (Hoesel 1993). We looked for evidence of neural degeneration in the auditory brainstem involved in binaural sound localisation, since this may have degraded the neural circuitry required to accurately code interaural time delays. Method: The brainstem of a bilateral cochlear implantee was prepared for light microscopy by embedding it in paraffin, sectioning at 10 mm and staining sections with thionine or Luxol fast blue (LFB). The histological sections were digitised with NIH Image and 3-dimensional reconstructions made of the cochlear nucleus (CN) and superior olivary complex (SOC) with AnalysePC. Within the CN and the SOC, cell number and size were estimated by the physical dissector technique following thionine staining, and myelination of the nerve fibres was estimated using the optical density method following LFB staining. Results: A reduction in cell size (from thionine staining) and myelination (from LFB staining) was seen in both the CN and the SOC. Conclusions: These finding are consistent with neural degeneration within the auditory pathways. This may have lead to a degradation of the neural circuitry required to accurately detect interaural time delays.
  • Item
    Thumbnail Image
    The role of radiographic phase-contrast imaging in the development of intracochlear electrode arrays
    XU, JIN ; Stevenson, Andrew W. ; Gao, Dachao ; TYKOCINSKI, MICHAEL ; LAWRENCE, DAVID ; Wilkins, Stephen W. ; Clark, Graeme M. ; Saunders, Elaine ; Cowan, Robert S. ( 2001)
    Objective: This study describes the application of a new radiographic imaging modality, phase-contrast radiography, to in vitro human temporal bone imaging and investigates it use in the development of new electrode arrays for cochlear implants. Background: The development of perimodiolar electrode arrays for cochlear implants requires detailed information from postoperative radiologic assessment on the position of the array in relation to the cochlear structures. Current standard radiographic techniques provide only limited details. Materials and Methods: Nucleus standard electrode arrays and perimodiolar Contour electrode arrays were implanted into the scala tympani of 11 human temporal bones. Both conventional and phase-contrast radiographs were taken of each temporal bone for comparative purposes. Results: Phase-contrast imaging provides better visulization of anatomic details of the inner ear and of the structure of the intracochlear electrode array, and better definition of electrode location in relation to cochlear walls. Conclusion: Phase-contrast radiography offers significant improvement over conventional radiography in images of in vitro human temporal bones. It seems to be a valuable tool in the development of intracochlear electrode arrays and cochlear implant research. However, this new radiographic technique still requires certain computational and physics challenges to be addressed before its clinical use can be established.
  • Item
    Thumbnail Image
    The Contour electrode array: Safety study and initial patient trials of a new perimodiolar design
    Tykocinski, M ; Saunders, E ; Cohen, LT ; Treaba, C ; Briggs, RJS ; Gibson, P ; Clark, GM ; Cowan, RSC (LIPPINCOTT WILLIAMS & WILKINS, 2001-01)
    OBJECTIVE: The aim of these studies was to investigate the insertion properties and safety of a new intracochlear perimodiolar electrode array design (Contour). BACKGROUND: An electrode array positioned close to the neural elements could be expected to reduce stimulation thresholds and might potentially reduce channel interaction. METHODS: Two sequential studies were conducted. In study 1, the Contour electrode array was inserted in 12 human temporal bones. After cochlear surface preparation, the position of the array was noted and the basilar membrane was examined for insertion damage. On the basis of the outcome of this temporal bone study, study 2 investigated the Contour array, mounted on a Nucleus CI-24 M device and implanted in three adult patients. RESULTS: Study I showed that in 10 temporal bones, the Contour array was positioned close to the modiolus, and the basilar membrane was intact. In the two remaining bones, the arrays had pierced the basilar membrane and were positioned in the scala vestibuli apical to the penetration. Statistical analysis showed an equivalent probability of insertion-induced damage of the two array designs. In study 2, image analysis indicated that the Contour electrodes were positioned closer to the modiolus than the standard Nucleus straight array. Lower T and C levels, but higher impedance values, were recorded from electrodes close to the modiolus. Initial speech perception data showed that all patients gained useful open-set speech perception, two patients achieving scores of 100% on sentence material 3 months postoperatively. CONCLUSIONS: The temporal bone studies showed the Contour electrode array to be generally positioned closer to the modiolus than the standard Nucleus straight array, and to have an equivalent probability of causing insertion-induced damage.
  • Item
    Thumbnail Image
    Variation in speech perception scores among children with cochlear implants
    Sarant, JZ ; Blamey, PJ ; Dowell, RC ; Clark, GM ; Gibson, WPR (LIPPINCOTT WILLIAMS & WILKINS, 2001-02)
    OBJECTIVE: The objective of this study was to identify common factors affecting speech perception scores in children with cochlear implants. DESIGN: Speech perception data for 167 implanted children were collected at two cochlear implant centres in Melbourne and Sydney. The data comprised audition-alone scores on open-set word and sentence tests. Children were selected on the basis that they had a Nucleus 22-electrode cochlear implant. The average age of the children was 5 yr. Information was also collected about 12 factors that may have influenced speech perception scores for each child. Analysis of covariance was used to identify factors that significantly affected speech perception scores. Pearson pairwise correlation coefficients were also calculated for all factors analyzed. RESULTS: The analyses in this study identified factors that accounted for 51%, 34%, and 45% of the variance in phoneme, word and sentence perception scores. Scores decreased by 1.4 to 2.4% per year of profound deafness prior to implantation. Children who normally use oral communication scored significantly higher than children normally using sign or simultaneous oral and sign communication. Children implanted in Sydney scored higher on average than children implanted in Melbourne. CONCLUSIONS: The results show that a significant part of the variation in speech perception scores is systematically related to audiological and environmental factors for each child. The reasons for significant differences between children using different communication modes or from different clinics were not identified.
  • Item
    Thumbnail Image
    Psychophysics of a prototype peri-modiolar cochlear implant electrode array
    Cohen, LT ; Saunders, E ; Clark, GM (ELSEVIER SCIENCE BV, 2001-05)
    Psychophysical measurements were performed in three hearing-impaired adult subjects implanted with a CI22 cochlear prosthesis (Cochlear Ltd.) fitted with a developmental peri-modiolar electrode array. The array was manufactured with a curvature approximating that of the inner wall of the scala tympani but, after straightening and insertion, lay on average about half way between the inner and outer walls of the scala. All subjects were tested with bipolar stimulation; two were also tested with monopolar, employing the most basal electrode as the return. Maximum comfortable level and threshold reduced with decreasing distance of electrode from the modiolus, whereas dynamic range increased. The linearity of the loudness growth function did not vary significantly with electrode position but the function was more non-linear for lower maximum comfortable levels. Current level discrimination, normalized with respect to dynamic range, improved with decreasing distance of electrode from the modiolus in two subjects. Pitch varied regularly with insertion depth of the stimulated electrode for bipolar stimulation in two subjects and also for monopolar stimulation in one subject. Electrode discrimination was enhanced by closeness to the modiolus. Whereas the forward masking patterns for bipolar stimulation of electrodes close to the modiolus had a sharp double-peaked structure, those for monopolar stimulation were flatter and had a single peak.