Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 10 of 10
  • Item
    Thumbnail Image
    A comparison of a new prototype Tickle Talker with a Tactaid 7
    Galvin, Karyn L. ; Ginis, Jan ; Cowan, Robert S. C. ; Blamey, Peter J. ; Clark, Graeme M. ( 2001)
    This study compared the speech perception enhancement provided by two multichannel tactile aids: a new version of the Tickle TalkerT™ and the Tactaid 7. The subjects' impression of benefit was also examined. In an AB pattern, six adults with hearing impairment used each device daily for approximately 18 weeks and attended 12 training sessions. When tactile information was provided, the group demonstrated a significant enhancement for the perception of words (mean 17.2%) and phonemes (mean 12.9%) in monosyllabic word lists, words in sentences (mean 14.2%) and speech tracking (mean 7.7 wpm). The Tactaid 7 provided a significantly greater enhancement for the perception of words (21 % versus 13.4%), phonemes (16.7% versus 9.1%) and some speech features in monosyllabic word lists. Subjective ratings were slightly higher for the Tactaid 7, and four subjects preferred this device. Either device may be suitable for those not able or willing to have a cochlear implant.
  • Item
    Thumbnail Image
    A clinical report on speech production of cochlear implant users
    Dawson, P. W. ; Blamey, P. J. ; Dettman, S. J. ; Rowland, L. C. ; Barker, E. J. ; Tobey, E. A. ; Busby, P. A. ; Cowan, R. C. ( 1995)
    Objective: The aim was to assess articulation and speech intelligibility over time in a group of cochlear implant users implanted at 8 yr or over. The hypothesis was that the postoperative speech production performance would be greater than the preoperative performance. Design: A test of intelligibility using sentences and an articulation test measuring non-imitative elicited speech were administered to 11 and 10 subjects, respectively, who were implanted with the 22-electrode cochlear implant. Nine subjects received both tests. Age at implantation ranged from 8 yr to 20 yr and implant use ranged from 1 yr to 4 yr 5 mo. Results: For both the intelligibility and articulation tests roughly half of the subjects showed significant improvements over time and group mean postoperative performance significantly exceeded preoperative performance. Improvements occurred for front, middle, and back consonants; for stops, fricatives, and glides and for voiceless and voiced consonants. Conclusions: Despite being deprived of acoustic speech information for many childhood years, roughly half of the patients assessed showed significant gains in speech intelligibility and articulation postimplantation. The lack of a control group of non-implanted patients means that we cannot separate out the influence of the implant on speech production from other influences such as training and tactile-kinaesthetic feedback.
  • Item
    Thumbnail Image
    Issues in the development of multichannel tactile devices for hearing-impaired children and adults
    Cowan, Robert S. C. ; Galvin, Karyn L. ; Sarant, Julia Z. ; Blamey, Peter J. ; Clark, Graeme M. ( 1995)
    Levitt, Pickett and Houde (1980), in their landmark monograph, noted that the history of tactile aid development has been characterized by periodic bursts of enthusiasm and research, often culminating in identification of new avenues to be explored for improving tactile perception of speech. While several research groups have maintained long-term interest in tactile research (Boothroyd, 1985; Oller, Payne, & Gavin, 1980; Saunders, 1985), there was a marked increase in reports of new multichannel tactile devices during the 1980s (reviewed in McGarr, 1989). This upsurge may have been spurred in part by the rapid increase world-wide in the number of hearing-impaired children and adults using cochlear implants as everyday communication devices, and the perceived need for a non-surgical approach to assisting hearing-impaired children. Despite this increase in tactile research, no tactile device has yet achieved widespread commercial use by the hearing-impaired community. It is, therefore, of interest to question why cochlear implants have been more widely accepted than tactile devices.
  • Item
    Thumbnail Image
    Issues in long-term management of children with cochlear implants and tactile devices [Abstract]
    COWAN, ROBERT ; DOWELL, RICHARD ; Barker, Elizabeth ; GALVIN, KARYN ; DETTMAN, SHANI ; SARANT, JULIA ; RANCE, GARY ; Hollow, Rod ; BLAMEY, PETER ; Clark, Graeme M. ( 1994)
    For many children with severe and profound hearing losses, conventional hearing aids are unable to provide sufficient amplification to ensure good oral communication and/or in the case of very young children, development of speech and language. Traditionally a number of these children have opted for the use of sign language alone or in Total Communication approaches as a primary means of communication. The advent of multiple channel cochlear implants for children and the continuing development of multiple channel speech processing tactile devices provide auditory approaches to resolving communication difficulties for these children. The successful use of such devices depends on a number of factors including the information provided through the aid; the ease of use, convenience and reliability of the aid; the individual communication needs of the child; and the habilitation and management program used with the device. Long-term data has shown that children continue to show increased speech perception benefits from improvements in speech processing and from further experience with these devices. Habilitation and management programs must therefore be geared to meet the changing needs of children as they progress and of families as children mature and face new challenges. Habilitation must address specific individual needs in speech perception and in speech production. For very young children, benefits of improved speech perception should have an impact on the development of speech and language, and habilitation and management must emphasise the need for language growth.
  • Item
    Thumbnail Image
    Habilitation issues in the management of children using the cochlear multiple-channel cochlear prosthesis
    Cowan, Robert S. C. ; Barker, Elizabeth J. ; Dettman, Shani J. ; Blamey, Peter J. ; RANCE, GARY ; Sarant, Julia Z. ; Galvin, Karyn L. ; Dawson, Pam W. ; Hollow, Rod ; Dowell, Richard C. ; PYMAN, BRIAN ; Clark, Graeme M. (Wien, 1994)
    Since 1985, a significant proportion of patients seen in the Melbourne cochlear implant clinic have been children. The children represent a diverse population, with both congenital and acquired hearing-impairments, a wide-range of hearing levels pre-implant, and an age range from 2 years to 18 years. The habilitation programme developed for the overall group must be flexible enough to be tailored to the individual needs of each child, and to adapt to the changing needs of children as they progress. Long-term data shows that children are continuing to show improvements after 5-7 years of device use, particularly in their perception of open-set words and sentences. Habilitation programs must therefore be geared to the long-term needs of children and their families. Both speech perception and speech production need to be addressed in the specific content of the habilitation program for any individual child. In addition, for young children, the benefits of improved speech perception should have an impact on development of speech and language, and the focus of the programme for this age child will reflect this difference in emphasis. Specific materials and approaches will vary for very young children, school-age and teenage children. In addition, educational setting will have a bearing on the integration of listening and device use into the classroom environment.
  • Item
    Thumbnail Image
    Cochlear implants for congenitally deaf adolescents: is open-set speech perception a realistic expectation?
    Sarant, J. Z. ; Cowan, R. S. C. ; Blamey, P. J. ; Galvin, K. L. ; Clark, Graeme M. ( 1994)
    The prognosis for benefit from use of cochlear implants in congenitally deaf adolescents, who have a long duration of profound deafness prior to implantation, has typically been low. Speech perception results for two congenitally deaf patients implanted as adolescents at the University of Melbourne/Royal Victorian Eye and Ear Hospital Clinic show that, after 12 months of experience, both patients had significant open-set speech discrimination scores without lipreading. These results suggest that although benefits may in general be low for congenitally deaf adolescents, individuals may attain significant benefits to speech perception after a short period of experience. Prospective patients from this group should therefore be considered on an individual basis with regard to prognosis for benefit from cochlear implantation.
  • Item
    Thumbnail Image
    Role of a multichannel electrotactile speech processor in a cochlear implant program for profoundly hearing-impaired adults
    Cowan, Robert S. C. ; Blamey, Peter J. ; Sarant, Julia Z. ; Galvin, Karyn L. ; Alcantara, Joseph I. ; Whitford, Lesley A. ; Clark, Graeme M. ( 1991)
    Four profoundly hearing-impaired adults who did not meet current selection criteria for implantation at the University of Melbourne were each fitted with a wearable multichannel electrotactile speech processor (Tickle Talker). The subjects were evaluated with a test battery of speech discrimination tests subsequent to training in use of the device. Thresholds for detection of pure tones were lower for the Tickle Talker than for hearing aids across the frequency range 250 to 4000 Hz. Mean speech detection thresholds for the Ling 5-sound test showed that all sounds were detected by users of the electrotactile device at normal conversational speech intensity levels. Mean speech discrimination scores were significantly higher (p < 0.05) in the tactually aided condition as compared with the tactually unaided for identification of vowels and consonants, on open-set words, open-set sentences, and on connected discourse tracking. Mean scores increased by 20% for vowels, 19% for consonants, 30% for openset words, and 25% for open-set sentences when the Tickle Talker was used in a multimodal combination with lipreading or lipreading and hearing aids. Speechtracking rates for three subjects showed increases of from 18 to 28 wpm when the tactile device was used. Comparison of tactually aided versus unaided tracking rates for two subjects with long-term experience shows continuing improvement with additional experience with the device. These results demonstrate that hearing impaired adults not meeting selection criteria for cochlear implantation may benefit from use of an electrotactile speech processor, and highlight the potential benefits from integration of such devices into cochlear implant programs for profoundly hearing-impaired patients.
  • Item
    Thumbnail Image
    Improved sound processing for cochlear implants
    James, C.J. ; Just, Y. ; Knight, M.R. ; Martin, L.F.A. ; McKay, C.M. ; Plant, K.L. ; Tari, S. ; Vandali, A.E. ; Clark, Graeme M. ; Cowan, R.S.C. ; McDermott, H. J. ; Blamey, P. J. ; Dawson, P. ; Fearn, R. A. ; Grayden, D. B. ; Henshall, K. R. ( 2002)
    Four signal processing schemes currently under development aim to improve the perception of sounds/ especially speech, for children and adults using the Nucleus cochlear implant system. The schemes are (1) fast-acting input-signal compression, (2) Adaptive Dynamic Range Optimisation (ADRO), (3) TESM, a scheme that emphasises transients in signals, and (4) DRSP, a strategy that applies different stimulation rates to selected sets of electrodes.
  • Item
    Thumbnail Image
    What factors contribute to successful outcomes for children using cochlear implants
    Cowan, Robert C. ; Clark, Graeme M. ; Dowell, Richard C. ; Dettman, Shani J ; Barker, Elizabeth ; Latus, Katie ; Hollow, Rod ; Blamey, Peter J. ( 2000)
    Long term speech perception data has been collected for 100 children using the Nucleus multichannel cochlear prosthesis in Melbourne. Scores on a number of different assessments are available at approximately six month intervals following implantation for these children. The group represents an unselected sample of cochlear implant users, as all children were included if they had sufficient developmental skills to perform formal speech perception tests. Information was also collected on each child regarding type of hearing loss, age of onset of profound hearing loss, duration of profound hearing loss, age at implantation, pre and post-implant communication mode, developmental delay, speech processing strategy and length of experience with implant use.
  • Item
    Thumbnail Image
    Comparative performance of children using the cochlear 22-channel implant and 8-channel 'Tickle Talker'
    Cowan, Robert S. C. ; Sarant, S. J. ; Dettman, K. L ; Galvin, K. L. ; Blamey, P. J. ; Clark, Graeme M. ( 1992)
    Direct comparison of "Tickle Talker" and Cochlear Implant users is problematic, due to difficulties in matching groups of children for hearing loss, age, duration of deafness, speech perception and language skills, and educational placement. However, two studies were undertaken to compare and contrast potential benefits available from these two devices. In the first study, a number of children from one educational setting were evaluated over a six month period. Half of the children used the multiple channel cochlear implant, while the other half used the multiple channel "Tickle Talker". The number of training sessions, clinicians involved, type of training provided, and overall management philosophy were identical for both devices. Comparison of progress of these two groups of children demonstrates that both devices are effective in improving' communication. However, differences were found with the tactile device being more limited in information provided and speech perception benefits. In the second study, two children who have used both devices were evaluated. These two children initially used the ''Tickle Talker" for periods up to two years, and subsequently had a multiple-channel cochlear implant. Similar habilitation was provided to these children with both devices, and measures of speech perception were taken at similar time periods. Results for one of the children, a prelinguistically deafened adolescent, already show similar benefits in terms of supplementation of lipreading for both devices. However, this patient has also shown some open-set word and sentence perception using the implant-alone. To date, this level of performance has not been achieved with the ''Tickle Talker". The results of these studies suggest a role for a multiple channel tactile device in a cochlear implant clinic as a training device for evaluating the ability of adults or children to integrate speech information presented through different sensory modalities. This could facilitate pre-implant evaluation of the potential for children to benefit from added speech information, which is often difficult to evaluate in congenitally deaf children.