    Meningitis after cochlear implantation: the risk is low, and preventive measures can reduce this further
    Wei, Benjamin P. C. ; Clark, Graeme M. ; O'Leary, Stephen J. ; Shepherd, Robert K. ; Robins-Browne, Roy M. ( 2007)
    Since the 1980s, more than 80 000 people have received cochlear implants worldwide. These implants are designed to enable people who are severely or profoundly deaf to experience sound and speech. Since 1990, implantation has become standard treatment for people who cannot communicate effectively despite well fitted hearing aids. Children who are deaf when they are born can perceive sound and learn to speak if they receive cochlear implants at a young age (ideally under 18 months). The use of cochlear implants has been thought to be safe. But since 2002 the number of patients with meningitis related to cochlear implantation has increased worldwide. Mortality and neurological complications after meningitis are high. We need to investigate the reasons for this and look at measures to reduce them.
  • Item
    Thumbnail Image
    Threshold shift: effects of cochlear implantation on the risk of pneumococcal meningitis
    Wei, Benjamin P. C. ; Shepherd, Robert K. ; Robins-Browne, Roy M. ; Clark, Graeme M. ; O'Leary, Stephen J. ( 2007)
    Unavailable due to copyright.
  • Item
    Thumbnail Image
    Effects of inner ear trauma on the risk of pneumococcal meningitis
    Wei, Benjamin P. C. ; Shepherd, Robert K. ; Robins-Browne, Roy M. ; Clark, Graeme M. ; O'LEARY, STEPHEN ( 2007)
    Objective: To examine the risk of pneumococcal meningitis in healthy rats that received a severe surgical trauma to the modiolus and osseous spiral lamina or the standard insertion technique for acute cochlear implantation. Design: Interventional animal studies. Subjects: Fifty-four otologically normal adult Hooded- Wistar rats. Interventions: Fifty-four rats (18 of which received a cochleostomy alone; 18, a cochleostomy and acute cochlear implantation using standard surgical techniques; and 18, a cochleostomy followed by severe inner ear trauma) were infected 4 weeks after surgery with Streptococcus pneumoniae via 3 different routes (hematogenous, middle ear, and inner ear) to represent all potential routes of bacterial infection from the upper respiratory tract to the meninges in cochlear implant recipients with meningitis. Results: Severe trauma to the osseous spiral lamina and modiolus increased the risk of pneumococcal meningitis when the bacteria were given via the middle or inner ear (Fisher exact test, P<.05). However, the risk of meningitis did not change when the bacteria were given via the hematogenous route. Acute electrode insertion did not alter the risk of subsequent pneumococcal meningitis for any route of infection. Conclusions: Severe inner ear surgical trauma to the osseous spiral lamina and modiolus can increase the risk of pneumococcal meningitis. Therefore, every effort should be made to ensure that cochlear implant design and insertion technique cause minimal trauma to the bony structures of the inner ear to reduce the risk of pneumococcalmeningitis.
  • Item
    Thumbnail Image
    Assessment of the protective effect of pneumococcal vaccination in preventing meningitis after cochlear implantation
    Wei, Benjamin P. C. ; Robins-Browne, Roy M. ; Shepherd, Robert K. ; AZZOPARDI, KRISTY ; Clark, Graeme M. ; O'Leary, Stephen J. ( 2007)
    Objectives: To examine if a 23-valent pneumococcal capsular polysaccharide vaccine (PPV23) reduces the risk of meningitis in healthy rats after cochlear implantation. Design: Interventional animal study. Interventions: Thirty-six rats (18 immunized and 18 unimmunized) received cochlear implantations and were then infected with Streptococcus pneumoniae via 3 different routes (hematogenous, middle ear, and inner ear) in numbers sufficient to induce meningitis. Results: The rats with implants that received PPV23 were protected from meningitis when the bacteria were delivered via the hematogenous and middle-ear routes (Fisher exact test P<.05). However, the protective effect of the vaccine in the rats with implants was only moderate when the bacteria were inoculated directly into the inner ear. Conclusions: Our animal model clearly demonstrates that immunization can protect healthy rats with a cochlear implant from meningitis caused by a vaccine-covered serotype. This finding supports the notion that all current and future implant recipients should be vaccinated against S pneumoniae.
  • Item
    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve using non-charge-balanced stimuli
    Shepherd, Robert K. ; Linahan, N. ; Xu, J. ; Clark, Graeme M. ; Araki, S. ( 1999)
    This study was designed to evaluate the pathophysiological response of the cochlea following long-term intracochlear electrical stimulation using a poorly charge-balanced stimulus regime, leading to direct current (DC) levels >0.1 µA. Four normal-hearing adult cats were bilaterally implanted with scala tympani electrode arrays and unilaterally stimulated for periods up to 2200 h. Stimuli consisted of 50 µs monophasic current pulses presented at 2000 pulses per second (pps) per channel, and resulted in DC levels of 0.4-2.8 µA. Both acoustic and electrical (EABR) evoked potentials were periodically recorded during the stimulation program. Frequency-specific stimuli indicated that an extensive and widespread hearing loss occurred over the 4-24 KHz region in all stimulated cochleae, although the 2 KHz region exhibited thresholds close to normal in some animals, despite long-term implantation and chronic stimulation. Longitudinal EABRs showed a statistically significant increase in threshold for three of the four animals. Histopathological evaluation of the cochleae revealed a highly significant reduction in ganglion cell density in stimulated cochleae compared with their controls. Spiral ganglion cell loss was significantly correlated with the degree of inflammation, duration of electrical stimulation, and the level of DC. In conclusion, the present study highlights the potential for neural damage following stimulation using poorly charge-balanced stimuli.
  • Item
    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve at high rates: I. Effect on residual hearing [Abstract]
    Xu, J. ; Shepherd, R. K. ; Clark, Graeme M. ( 1996)
    In addition to direct excitation of auditory nerve fibres, cochlear implant patients with small amounts of residual hearing may receive important additional auditory cues via electrophonic activation of hair cells 1. Before incorporating electrophonic hearing into speech processing strategies, the extent of hair cell survival following cochlear implantation must first be determined. We have recently demonstrated widespread survival of hair cells apical to electrode arrays implanted for periods of up to three years, the present report describes the effects of chronic electrical stimulation on hair cell survival.
  • Item
    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve at high rates: II. Cochlear pathophysiology [Abstract]
    Shepherd, R. K. ; Xu, J. ; Clark, Graeme M. ( 1996)
    A major factor in the improved performance of cochlear implant patients has been the use of high stimulus rate speech processing strategies. While these strategies show clear clinical advantage, we know little of their long-term safety. Indeed, recent studies have indicated that high stimulus rates at intensities above clinical limits, can result in neural damage as a result of prolonged neuronal hyperactivity. The present study was designed to evaluate the effects of chronic electrical stimulation of the auditory nerve at high rates, using intensities within clinical limits.
  • Item
    Thumbnail Image
    Spatial representation of the cochlea within the inferior colliculus of neonatally deafened kittens following chronic electrical stimulation of the auditory nerve [Abstract]
    Shepherd, R. K. ; Martin, R. L. ; Brown, M. ; Clark, Graeme M. ( 1995)
    The orderly tonotopic representation of the cochlea is accurately reproduced within the central auditory system of normal hearing animals. Any degradation of this representation as a result of a neonatal hearing loss or chronic electrical stimulation during development could have important implications for the use of multichannel cochlear implants in young children. In the present study we have used 2-deoxyglucose autoradiography (2-00) to examine the topographic representation of the cochlea within the inferior colliculus (IC) of neonatally deafened kittens following periods of chronic intracochlear electrical stimulation.
  • Item
    Thumbnail Image
    Cochlear histopatholgic characteristics following long-term implantation: safety studies in the young monkey
    Burton, Martin J. ; Shepherd, Robert K. ; Clark, Graeme M. ( 1996)
    Objective: To evaluate the safety of cochlear implantation in children 2 years of age or younger using a non-human primate model.
  • Item
    Thumbnail Image
    Cochlear pathology following chronic electrical stimulation using non charge balanced stimuli
    Shepherd, Robert K. ; Matsushima, Jun-Ichi ; Millard, R. E. ; Clark, Graeme M. ( 1991)
    During the course of a chronic intracochlear electrical stimulation study using charge balanced biphasic current pulses, one animal inadvertently received a short period of direct current (DC) stimulation at a level of approximately 1 µA. Subsequent, the animal was chronically stimulated using a poorly charge balanced waveform that produced a DC level of approximately 2 µA. Extensive pathological changes were observed within the cochlea. These changes included widespread spiral ganglion cell loss and new bone growth that extended throughout all turns of the cochlea. Significant changes in the morphology of the electrically evoked auditory brainstem response (EABR) were associated with these pathological changes. EABRs recorded prior to the DC stimulation exhibited a normal waveform morphology. However, responses recorded during the course of the DC stimulation were dominated by a short latency response believed to be vestibular in origin. The response thresholds were also significantly higher than levels recorded before the DC stimulation. In contrast, the contralateral cochlea, stimulated using charge balanced stimuli, showed no evidence of adverse pathological changes. Furthermore, EABRs evoked from this cochlea remained stable throughout the chronic stimulation period. Although preliminary, the present results illustrate the adverse nature of poorly charge balanced electrical stimuli. These results have important implications for both the design of neural prostheses and the use of DC stimuli to suppress tinnitus in patients.