Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 10 of 38
  • Item
    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve at high stimulus rates: a physiological and histopathological study
    XU, JIN ; Shepherd, Robert K. ; Millard, Rodney E. ; Clark, Graeme M. ( 1997)
    A major factor associated with recent improvements in the clinical performance of cochlear implant patients has been the development of speech-processing strategies based on high stimulation rates. While these processing strategies show clear clinical advantage, we know little of their long-term safety implications. The present study was designed to evaluate the physiological and histopathological effects of long-term intracochlear electrical stimulation using these high rates. Thirteen normal-hearing adult cats were bilaterally implanted with scala tympani electrode arrays and unilaterally stimulated for periods of up to 2100 h using either two pairs of bipolar or three monopolar stimulating electrodes. Stimuli consisted of short duration (25-50 µs/phase) charge-balanced biphasic current pulses presented at 1000 pulses per second (pps) per channel for monopolar stimulation, and 2000 pps/channel for bipolar stimulation. The electrodes were shorted between current pulses to minimize any residual direct current, and the pulse trains were presented using a 50% duty cycle (500 ms on; 500 ms oft) in order to simulate speech. Both acoustic (ABR) and electrical (EABR) auditory brainstem responses were recorded periodically during the chronic stimulation program, All cochleas showed an increase in the click-evoked ABR threshold following implant surgery; however, recovery to near-normal levels occurred in approximately half of the stimulated cochleas 1 month post-operatively. The use of frequency-specific stimuli indicated that the most extensive hearing loss generally occurred in the high-frequency basal region of the cochlea (12 and 24 kHz) adjacent to the stimulating electrode. However, thresholds at lower frequencies (2, 4 and 8 kHz), appeared at near-normal levels despite long-term electrode implantation and electrical stimulation. Our longitudinal EABR results showed a statistically significant increase in threshold in nearly 40% of the chronically stimulated electrodes evaluated; however, the gradient of the EABR input/output (I/O) function (evoked potential response amplitude versus stimulus current) generally remained quite stable throughout the chronic stimulation period. Histopathological examination of the cochleas showed no statistically significant difference in ganglion cell densities between cochleas using monopolar and bipolar electrode configurations (P = 0.67), and no evidence of cochlear damage caused by high-rate electrical stimulation when compared with control cochleas. Indeed, there was no statistically significant relationship between spiral ganglion cell density and electrical stimulation (P = 0.459), or between the extent of loss of inner (IHC, P = 0.86) or outer (OHC, P=0.30) hair cells and electrical stimulation. Spiral ganglion cell loss was, however, influenced by the degree of inflammation (P=0.016) and electrode insertion trauma. These histopathological findings were consistent with the physiological data. Finally, electrode impedance, measured at completion of the chronic stimulation program, showed close correlation with the degree of tissue response adjacent to the electrode array. These results indicated that chronic intracochlear electrical stimulation, using carefully controlled charge-balanced biphasic current pulses at stimulus rates of up to 2000 pps/channel, does not appear to adversely affect residual auditory nerve elements or the cochlea in general. This study provides an important basis for the safe application of improved speech-processing strategies based on high-rate electrical stimulation.
  • Item
    Thumbnail Image
    Reduction in excitability of the auditory nerve following electrical simulation at high stimulus rates. II. Comparison of fixed amplitude with amplitude modulated stimuli
    TYKOCINSKI, MICHAEL ; Shepherd, Robert K. ; Clark, Graeme M. ( 1997)
    We have previously shown that acute electrical stimulation of the auditory nerve using charge-balanced biphasic current pulses presented continuously can lead to a prolonged decrement in auditory nerve excitability (Tykocinski et al., Hear. Res. 88 (1995), 124-142). This work also demonstrated a reduction in electrically evoked auditory brainstem response (EABR) amplitude decrement when using an otherwise equivalent pulse train with a 50% duty cycle. In the present study we have extended this work in order to compare the effects of electrical stimulation using both fixed amplitude electrical pulse trains and amplitude modulated (AM) pulse trains that more accurately model the dynamic stimulus paradigms used in cochlear implants. EABRs were recorded from guinea pigs following acute stimulation using AM trains of charge-balanced biphasic current pulses. The extent of stimulus-induced reductions in the EABR were compared with our previous results using either fixed amplitude continuous, or 50% duty cycle pulse trains operating at 0.34 µC/phase (2 mA, 170 µs/phase) at 400 or 1000 pulses/s (Tykocinski et al., Hear. Res. 88 (1995) 124-142). The AM pulse train, operating at the same rates, was based on a I-s sequence of the most extensively activated electrode of a Nucleus Mini-22 cochlear implant using the SPEAK speech processing strategy exposed to 4-talker babble, and delivered the same total charge as the fixed amplitude 50% duty cycle pulse train. Two hours of continuous stimulation induced a significant, rate-dependent reduction in auditory nerve excitability, and showed only a slight post-stimulus recovery for monitoring periods of up to 6 hours. Following 2 or 4 h of stimulation using an otherwise equivalent pulse train with a 50% duty cycle or the AM pulse train, significantly less reduction in the EABR was observed, and recovery to pre-stimulus levels was generally rapid and complete. These differences in the extent of the recovery between the continuous waveform and both the 50% duty cycle and AM waveforms were statistically significant for both 400 and 1000 pulses/s stimuli. Consistent with our previous results, the stimulus changes observed using AM pulse trains were rate dependent, with higher rate stimuli evoking more extensive stimulus-induced changes. The present findings show that while stimulus-induced reductions in neural excitability are dependent on the extent of stimulus-induced neuronal activity, the use of an AM stimulus paradigm further reduces post-stimulus neural fatigue.
  • Item
  • Item
    Thumbnail Image
    Effects of chronic electrical stimulation on spiral ganglion neuron survival and size in deafened kittens
    Araki, Susumu ; Kawano, Atsushi ; Seldon, H. Lee ; Shepherd, Robert K. ; Funasaka, Sotaro ; Clark, Graeme M. ( 1998)
    We have studied spiral ganglion cell (SGC) survival and soma size in neonatally pharmacologically deafened kittens. They were implanted with a four-electrode array in the left cochlea at 100 to 180 or more days of age. Eight animals were chronically stimulated approximately 1000 hours over approximately 60 days with charge-balanced, biphasic current pulses; three were unstimulated controls. Using three-dimensional computer-aided reconstruction of the cochlea, the SGC position and cross-sectional area were stored. SGC position was mapped to the organ of Corti by perpendicular projections, starting from the basal end. The basal region of the cochlea was divided into three 4-mm segments. SGC survival (number per 0.1 mm of the length of the organ of Corti) and soma size for stimulated cochleae were compared statistically with implanted but unstimulated cochleae. There was no evidence of an effect of electrical stimulation on SGC survival under this protocol and with this duration. On the other hand, the cell size on the stimulated side was significantly larger than the control side in the middle segment (4 to 8 mm from the basal end). SGCs undergo a reduction in size after prolonged auditory deprivation; however, these changes may be partially moderated after chronic intracochlear electrical stimulation.
  • Item
    Thumbnail Image
    Cochlear implants in children: the value of cochleostomy seals in the prevention of labyrinthitis following pneumococcal otitis media
    Dahm, M. C. ; Webb, R. L. ; Clark, Graeme M. ; Franz, B. K-H. ; Shepherd, R. K. ; Burton, M. J. ; ROBINS-BROWNE, R. ( 1995)
    Cochlea implantation at an early age is important in rehabilitating profoundly hearing impaired children. Given the incidence of pneumococcal otitis media in young children, there has been concern that cochlear implantation could increase the possibility of otitis media, leading to labyrinthitis in this age group. Clinical experience has not indicated an increase in the frequency of otitis media and labyrinthitis in implanted adults or children over two years. However, labyrinthitis has occurred in implanted animals with otitis media. In order to assess the impact of cochlear implants on the occurrence of labyrinthitis, pneumococcal otitis media was induced in 21 kittens. Thirty-two kitten cochleas were implanted, of which 9 had a fascial graft and 9 a Gelfoam® graft. Nine control cochleas were unimplanted. Labyrinthitis occurred in 44% of unimplanted controls. 50% of implanted ungrafted cochleas, and 6% of implanted grafted cochleas. There was no statistically significant difference between the incidence of labyrinthitis in the implanted cochleas and the unimplanted controls. However there was a statistically significant difference between the ungrafted and grafted cochleas, but not between the two types of graft.
  • Item
    Thumbnail Image
    Electrical stimulation of residual hearing in the implanted cochlea
    Clark, Graeme M. ; McAnally, K. I. ; Black, R. C. ; Shepherd, R. K. ( 1995)
    The average profoundly deaf person using a cochlear implant can now understand more speech than some severely to profoundly deaf people who use a hearing aid. For this reason there will be an increasing need to consider implanting people with residual hearing. In many of these people there could be significant hearing in the operated ear, as a majority of severely to profoundly deaf people are likely to have a symmetrical hearing loss. When three frequency average hearing thresholds were measured on 219 pensioners from the Australian National Acoustic Laboratories (H. Dillon, unpublished findings), 64% had less than a 10-dB difference between thresholds in each ear.
  • Item
    Thumbnail Image
    Cochlear implantation: osteoneogenesis, electrode-tissue impedance, and residual hearing
    Clark, Graeme M. ; Shute, S. A. ; Shepherd, R. K. ; Carter, T. D. ( 1995)
    This study was undertaken to find out how new bone is produced in the implanted cochlea, and the effects of fibrous tissue and new bone growth on electrode-tissue impedance. This knowledge is essential, as bone and fibrous tissue in the cochlea could account for variations in patients' speech perception performance. The study was also carried out to examine the effects of implantation on residual hearing. This information is also important, as cochlear implant speech perception results in profoundly deaf people are now better on average than severely or profoundly deaf people obtain with a hearing aid. Consequently, more people will need to be considered for cochlear implantation in ears with some residual hearing. In this case we need to know to what extent residual hearing is affected by implantation. (From Introduction)
  • Item
    Thumbnail Image
    Temporal coding of frequency: neuron firing probabilities for acoustic and electric stimulation of the auditory nerve
    Clark, Graeme M. ; Carter, T. D. ; Maffi, C. L. ; Shepherd, R. K. ( 1995)
    A better understanding of the temporal coding of frequency, and its application to electrical stimulation of auditory nerve fibers, should lead to advances in cochlear implant speech processing. Past research studies have suggested that the intervals between nerve action potentials are important in the temporal coding of frequency. For sound frequencies up to approximately 500 Hz, the shortest or predominant intervals between the nerve action potentials are usually the same as the periods of the sound waves. The intervals between each nerve action potential can be plotted as an interval histogram. Although there is evidence that the intervals between spikes are important in the temporal coding of frequency, it is not known up to what frequency this applies. It is also not known whether the information transmitted along individual fibers or an ensemble of fibers is important, to what extent the coding of frequency is interrelated with the coding of intensity, the relative importance of temporal and place coding for different frequencies, and finally, how well electrical stimulation can simulate the temporal coding of sound.
  • Item
    Thumbnail Image
    Cochlear pathology following reimplantation of a multichannel scala tympani electrode array in the macaque
    Shepherd, Robert K. ; Clark, Graeme M. ; Xu, Shi-Ang ; Pyman, Brian C. ( 1995)
    The histopathologic consequence of removing and reimplanting intracochlear electrode arrays on residual auditory nerve fibers is an important issue when evaluating the safety of cochlear prostheses. The authors have examined this issue by implanting multichannel intracochlear electrodes in macaque monkeys. Macaques were selected because of the similarity of the surgical technique used to insert electrodes into the cochlea compared to that in humans, in particular the ability to insert the arrays into the upper basal turn. Five macaques were bilaterally implanted with the Melbourne/Cochlear banded electrode array. Following a minimum implant period of 5 months, the electrode array on one side of each animal was removed and another immediately implanted. The animals were sacrificed a minimum of 5 months following the reinsertion procedure, and the cochleas prepared for histopathologic analysis. Long-term implantation of the electrode resulted in a relatively mild tissue response within the cochlea. Results also showed that inner and outer hair cell survival, although significantly reduced adjacent to the array, was normal in 8 of the 10 cochleas apicalward. Moreover, the electrode reinsertion procedure did not appear to adversely affect this apical hair cell population. Significant new bone formation was frequently observed in both control and reimplanted cochleas close to the electrode fenestration site and was associated with trauma to the endosteum and/or the introduction of bone chips into the cochlea at the time of surgery. Electrode insertion trauma, involving the osseous spiral lamina or basilar membrane, was more commonly observed in reimplanted cochleas. This damage was usually restricted to the lower basal turn and resulted in a more extensive ganglion cell loss. Finally, in a number of cochleas part of the electrode array was located within the scala media or scala vestibuli. These electrodes did not appear to evoke a more extensive tissue response or result in more extensive neural degeneration compared with electrodes located within the scala tympani. In conclusion, the present study has shown that the reimplantation of a multichannel scala tympani electrode array can be achieved with minimal damage to the majority of cochlear structures. Increased insertion trauma, resulting in new bone formation and spiral ganglion cell loss, can occur in the lower basal turn in cases where the electrode entry point is difficult to identify due to proliferation of granulation and fibrous tissue.
  • Item
    Thumbnail Image
    Evaluation of a new Spectral Peak coding strategy for the Nucleus 22 channel cochlear implant system
    Skinner, Margaret W. ; Clark, Graeme M. ; Whitford, Lesley A. ; Seligman, Peter M. ; Staller, Steven J. ; Shipp, David B. ; Shallop, Jon K. ; Everingham, Colleen ; Menapace, Christine M. ; Arndt, Patti L. ; Antogenelli, Trisha ; Brimacombe, Judith A. ; Pijl, Sipke ; Daniels, Paulette ; George, Catherine R. ; McDermott, Hugh J. ; Beiter, Anne L. ( 1994)
    Sixty-three postlinguistically deaf adults from four English-speaking countries participated in a 17-week field study of performance with a new speech coding strategy, Spectral Peak (SPEAK), and the most widely used strategy, Multipeak (MPEAK), both of which are implemented on wearable speech processors of the Nucleus 22 Channel Cochlear Implant System; MPEAK is a feature-extraction strategy, whereas SPEAK is a filterbank strategy. Subjects' performance was evaluated with an experimental design in which use of each strategy was reversed and replicated (ABAB). Average scores for speech tests presented sound-only at 70 dB SPL were higher with the SPEAK strategy than with the MPEAK strategy. For tests in quiet, mean scores for medial vowels were 74.8 percent versus 70.1 percent; for medial consonants, 68.6 percent versus 56.6 percent; for monosyllabic words, 33.8 percent versus 24.6 percent; and for sentences, 77.5 percent versus 67.4 percent. For tests in noise, mean scores for Four-Choice Spondees at +10 and +5 dB signal-to-noise ratio (S/N) were 88.5 percent versus 73.6 percent and 80.1 percent versus 62.3 percent, respectively; and for sentences at +15 dB, +10, and +5 dB S/N, 66.5 percent versus 43.4 percent, 61.5 percent versus 37.1 percent, and 60.4 percent versus 31.7 percent, respectively. Subjects showed marked improvement in recognition of sentences in noise with the new SPEAK filterbank strategy. These results agree closely with subjects' responses to a questionnaire on which approximately 80 percent reported they heard best with the SPEAK strategy for everyday listening situations.