Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 10 of 35
  • Item
    Thumbnail Image
    A multiple-electrode cochlear implant
    Clark, Graeme M. ; Tong, Y. C. ; Bailey, Q. R. ; Black, R. C. ; Martin, L. F. ; Millar, J. B. ; O'Loughlin B. J. ; Patrick, J. F. ; Pyman, B. C. ( 1978)
    Interest in artificially stimulating the auditory nerve electrically for sensori-neural deafness was first sparked off by Volta in the 18th century. Count Volta, who was the first to develop the electric battery, connected up a number of his batteries to two metal rods which he inserted into his ears. Having placed the rods in his ears he pressed the switch and received "une secousse dans la tete" and perceived a noise like "the boiling of thick soup".
  • Item
    Thumbnail Image
    Hearing restoration with the multichannel auditory brainstem implant
    Briggs, R. J. S. ; Kaye, A. H. ; Dowell, R. C. ; Hollow, R. D. ; Clark, Graeme M. ( 1997)
    Restoration of useful hearing is now possible in patients with bilateral acoustic neuromas by direct electrical stimulation of the cochlear nucleus. Our first experience with the Multichannel Auditory Brainstem Implant is reported. A forty four year old female with bilateral acoustic neuromas and a strong family history of Neurofibromatosis Type II presented with profound bilateral hearing impairment. Translabyrinthine removal of the right tumour was performed with placement of the Nucleus eight electrode Auditory Brainstem Implant. Intraoperative electrically evoked auditory brainstem response monitoring successfully confirmed placement over the cochlear nucleus. Postoperatively, auditory responses were obtained on stimulation of all electrodes with minimal non-auditory sensations. The patient now receives useful auditory sensations using the "SPEAK" speech processing strategy. Auditory brainstem Implantation should be considered for patients with Neurofibromatosis Type II in whom hearing preservation tumour removal is not possible.
  • Item
    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve at high stimulus rates: a physiological and histopathological study
    XU, JIN ; Shepherd, Robert K. ; Millard, Rodney E. ; Clark, Graeme M. ( 1997)
    A major factor associated with recent improvements in the clinical performance of cochlear implant patients has been the development of speech-processing strategies based on high stimulation rates. While these processing strategies show clear clinical advantage, we know little of their long-term safety implications. The present study was designed to evaluate the physiological and histopathological effects of long-term intracochlear electrical stimulation using these high rates. Thirteen normal-hearing adult cats were bilaterally implanted with scala tympani electrode arrays and unilaterally stimulated for periods of up to 2100 h using either two pairs of bipolar or three monopolar stimulating electrodes. Stimuli consisted of short duration (25-50 µs/phase) charge-balanced biphasic current pulses presented at 1000 pulses per second (pps) per channel for monopolar stimulation, and 2000 pps/channel for bipolar stimulation. The electrodes were shorted between current pulses to minimize any residual direct current, and the pulse trains were presented using a 50% duty cycle (500 ms on; 500 ms oft) in order to simulate speech. Both acoustic (ABR) and electrical (EABR) auditory brainstem responses were recorded periodically during the chronic stimulation program, All cochleas showed an increase in the click-evoked ABR threshold following implant surgery; however, recovery to near-normal levels occurred in approximately half of the stimulated cochleas 1 month post-operatively. The use of frequency-specific stimuli indicated that the most extensive hearing loss generally occurred in the high-frequency basal region of the cochlea (12 and 24 kHz) adjacent to the stimulating electrode. However, thresholds at lower frequencies (2, 4 and 8 kHz), appeared at near-normal levels despite long-term electrode implantation and electrical stimulation. Our longitudinal EABR results showed a statistically significant increase in threshold in nearly 40% of the chronically stimulated electrodes evaluated; however, the gradient of the EABR input/output (I/O) function (evoked potential response amplitude versus stimulus current) generally remained quite stable throughout the chronic stimulation period. Histopathological examination of the cochleas showed no statistically significant difference in ganglion cell densities between cochleas using monopolar and bipolar electrode configurations (P = 0.67), and no evidence of cochlear damage caused by high-rate electrical stimulation when compared with control cochleas. Indeed, there was no statistically significant relationship between spiral ganglion cell density and electrical stimulation (P = 0.459), or between the extent of loss of inner (IHC, P = 0.86) or outer (OHC, P=0.30) hair cells and electrical stimulation. Spiral ganglion cell loss was, however, influenced by the degree of inflammation (P=0.016) and electrode insertion trauma. These histopathological findings were consistent with the physiological data. Finally, electrode impedance, measured at completion of the chronic stimulation program, showed close correlation with the degree of tissue response adjacent to the electrode array. These results indicated that chronic intracochlear electrical stimulation, using carefully controlled charge-balanced biphasic current pulses at stimulus rates of up to 2000 pps/channel, does not appear to adversely affect residual auditory nerve elements or the cochlea in general. This study provides an important basis for the safe application of improved speech-processing strategies based on high-rate electrical stimulation.
  • Item
    Thumbnail Image
    Acoustic and electric forward-masking of the auditory nerve compound action potential: evidence for linearity of electro-mechanical transduction
    McAnally, Ken I. ; Brown, Mel ; Clark, Graeme M. ( 1997)
    We investigated electro-mechanical transduction within the cochlea by comparing masking of the auditory nerve compound action potential (CAP) by acoustical and electrical maskers. Forward-masking of the CAP reflects the response to the masker of the cochlear location tuned to the probe. Electrical stimulation was delivered through bipolar stimulating electrodes within the basal turn of the scala tympani. The growth of masking of high-frequency probes which excite cochlear locations close to the stimulating electrodes was similar for both acoustic and electrical maskers, suggesting a linear transduction of electrical energy to mechanical energy. Exposure to intense acoustic stimulation caused an equal loss of sensitivity to acoustic and electrical maskers. Masking of lower-frequency probes by electrical maskers increased rapidly with masker current, suggesting the direct electrical stimulation of neural elements. This masking was reduced by the administration of strychnine suggesting a contribution by the efferents towards masking of these low-frequency probes.
  • Item
    Thumbnail Image
    Electrical stimulation of the auditory nerve with a cochlear implant and the temporal coding of sound frequencies: a brief review
    Clark, Graeme M. ( 1997)
    There is considerable evidence that the brain translates (encodes) the frequency of a sound into both place of excitation (place encoding), and the pattern of intervals between action potentials (temporal encoding). Furthermore, temporal encoding is now thought to be due to a temporal as well as spatial pattern of action potentials in a small group of neurons. This pattern needs to be reproduced with a cochlear implant for improved speech processing. Our recent research has also demonstrated that the timing of excitatory postsynaptic potentials seen with intracellular recordings from brain cells, rather than extracellularly recorded action potentials, correlates better with the frequency of sound. These excitatory postsynaptic potentials are likely to be the link between the patterns of action potentials arriving at nerve cells and the biomolecular activity in the cell. This response also needs to be replicated with improved speech processing strategies.
  • Item
    Thumbnail Image
    Effects of chronic electrical stimulation on spiral ganglion neuron survival and size in deafened kittens
    Araki, Susumu ; Kawano, Atsushi ; Seldon, H. Lee ; Shepherd, Robert K. ; Funasaka, Sotaro ; Clark, Graeme M. ( 1998)
    We have studied spiral ganglion cell (SGC) survival and soma size in neonatally pharmacologically deafened kittens. They were implanted with a four-electrode array in the left cochlea at 100 to 180 or more days of age. Eight animals were chronically stimulated approximately 1000 hours over approximately 60 days with charge-balanced, biphasic current pulses; three were unstimulated controls. Using three-dimensional computer-aided reconstruction of the cochlea, the SGC position and cross-sectional area were stored. SGC position was mapped to the organ of Corti by perpendicular projections, starting from the basal end. The basal region of the cochlea was divided into three 4-mm segments. SGC survival (number per 0.1 mm of the length of the organ of Corti) and soma size for stimulated cochleae were compared statistically with implanted but unstimulated cochleae. There was no evidence of an effect of electrical stimulation on SGC survival under this protocol and with this duration. On the other hand, the cell size on the stimulated side was significantly larger than the control side in the middle segment (4 to 8 mm from the basal end). SGCs undergo a reduction in size after prolonged auditory deprivation; however, these changes may be partially moderated after chronic intracochlear electrical stimulation.
  • Item
    Thumbnail Image
    Electrophonically driven single unit responses of the anteroventral cochlear nucleus in cat [Abstract]
    Morrison, N. A. ; Brown, M. ; Clark, Graeme M. ( 1996)
    Electrical stimulation of the cochlea results in both direct and electrophonic excitation of auditory nerve fibres. It has been proposed that electrophonic stimulation results from the creation of a mechanical disturbance on the basilar membrane which has properties similar those resulting from acoustic stimuli. Auditory nerve compound action potential (CAP) forward masking studies1 show the level of frequency specific electrophonic stimulation is highly correlated with the spectral energy of the electrical stimulus waveform. The level of spectral energy in pulsatile biphasic electrical stimuli decreases toward low frequencies suggesting the level of electrophonic stimulation will be diminished in the low frequency region of the cochlea.
  • Item
    Thumbnail Image
    Discharge rate-level functions from dorsal cochlear nucleus single units in response to acoustic and electrical stimulation of the auditory nerve
    O'Leary, S. J. ; Clark, Graeme M. ; Tong, Y. C. ( 1995)
    Discharge rate-level (I/O) functions possessed by dorsal cochlear nucleus (DCN) units were examined, in response to bipolar electrical stimulation of the cochlea of the barbiturate-anesthetized cat. Spontaneously active units usually possessed nonmonotonic functions with a minimum, and spontaneously inactive units usually possessed monotonic functions or nonmonotonic functions with a maximum (NM+). In response to acoustic high-pass filtered noise, the function relating discharge rate and cut off frequency resembled the same unit's I/O function to electrical stimulation. The I/O functions to acoustic characteristic tones were usually monotonic or NM+. These results suggest that in the DCN, a prerequisite for the generation of acoustic-like responses with an electrical stimulus may be the matching of the cochlear place and spatial extent activated by each stimulus.
  • Item
    Thumbnail Image
    Electrical stimulation of residual hearing in the implanted cochlea
    Clark, Graeme M. ; McAnally, K. I. ; Black, R. C. ; Shepherd, R. K. ( 1995)
    The average profoundly deaf person using a cochlear implant can now understand more speech than some severely to profoundly deaf people who use a hearing aid. For this reason there will be an increasing need to consider implanting people with residual hearing. In many of these people there could be significant hearing in the operated ear, as a majority of severely to profoundly deaf people are likely to have a symmetrical hearing loss. When three frequency average hearing thresholds were measured on 219 pensioners from the Australian National Acoustic Laboratories (H. Dillon, unpublished findings), 64% had less than a 10-dB difference between thresholds in each ear.
  • Item
    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve at high stimulus rates: preliminary results
    Shepherd, R. K. ; Xu, J. ; Millard, R. E. ; Clark, Graeme M. ( 1994)
    The present preliminary report describes the electrophysiological response of the cochlea during long-term stimulation. The data indicate that electrical stimulation at a rate of 1000 pulses per second does not appear to adversely affect the implanted cochlea.