    Effects of chronic electrical stimulation on spiral ganglion neuron survival and size in deafened kittens
    Araki, Susumu ; Kawano, Atsushi ; Seldon, H. Lee ; Shepherd, Robert K. ; Funasaka, Sotaro ; Clark, Graeme M. ( 1998)
    We have studied spiral ganglion cell (SGC) survival and soma size in neonatally pharmacologically deafened kittens. They were implanted with a four-electrode array in the left cochlea at 100 to 180 or more days of age. Eight animals were chronically stimulated approximately 1000 hours over approximately 60 days with charge-balanced, biphasic current pulses; three were unstimulated controls. Using three-dimensional computer-aided reconstruction of the cochlea, the SGC position and cross-sectional area were stored. SGC position was mapped to the organ of Corti by perpendicular projections, starting from the basal end. The basal region of the cochlea was divided into three 4-mm segments. SGC survival (number per 0.1 mm of the length of the organ of Corti) and soma size for stimulated cochleae were compared statistically with implanted but unstimulated cochleae. There was no evidence of an effect of electrical stimulation on SGC survival under this protocol and with this duration. On the other hand, the cell size on the stimulated side was significantly larger than the control side in the middle segment (4 to 8 mm from the basal end). SGCs undergo a reduction in size after prolonged auditory deprivation; however, these changes may be partially moderated after chronic intracochlear electrical stimulation.
  • Item
    Thumbnail Image
    Cochlear implants in children: the value of cochleostomy seals in the prevention of labyrinthitis following pneumococcal otitis media
    Dahm, M. C. ; Webb, R. L. ; Clark, Graeme M. ; Franz, B. K-H. ; Shepherd, R. K. ; Burton, M. J. ; ROBINS-BROWNE, R. ( 1995)
    Cochlea implantation at an early age is important in rehabilitating profoundly hearing impaired children. Given the incidence of pneumococcal otitis media in young children, there has been concern that cochlear implantation could increase the possibility of otitis media, leading to labyrinthitis in this age group. Clinical experience has not indicated an increase in the frequency of otitis media and labyrinthitis in implanted adults or children over two years. However, labyrinthitis has occurred in implanted animals with otitis media. In order to assess the impact of cochlear implants on the occurrence of labyrinthitis, pneumococcal otitis media was induced in 21 kittens. Thirty-two kitten cochleas were implanted, of which 9 had a fascial graft and 9 a Gelfoam® graft. Nine control cochleas were unimplanted. Labyrinthitis occurred in 44% of unimplanted controls. 50% of implanted ungrafted cochleas, and 6% of implanted grafted cochleas. There was no statistically significant difference between the incidence of labyrinthitis in the implanted cochleas and the unimplanted controls. However there was a statistically significant difference between the ungrafted and grafted cochleas, but not between the two types of graft.
  • Item
    Thumbnail Image
    Cochlear implantation in young children: histological studies on head growth, leadwire design, and electrode fixation in the monkey model
    Burton, M. J. ; Shepherd, R. K. ; Xu, S. A. ; Xu, J. ; Franz, B. K-H. G. ; Clark, Graeme M. ( 1994)
    For safe cochlear implantation in children under 2 years of age, the implant assembly must not adversely affect adjacent tissues or compromise head growth. Furthermore, growth changes and tissue responses should not impair the function of the device. Dummy receiver-stimulators, interconnect plugs, and leadwire-lengthening systems were implanted for periods of 36 months in the young monkey to effectively model the implantation of the young child. The results show that implanting a receiver-stimulator package has no adverse effects on skull growth or the underlying central nervous system. The system for fixing the electrode at the fossa incudis proved effective. There was marked osteoneogenesis in the mastoid cavity, resulting in the fixation of the leadwire outside the cochlea. This study provides evidence for the safety of cochlear implantation in young subjects.
  • Item
    Thumbnail Image
    A multiple-electrode intracochlear implant for children
    Clark, Graeme M. ; Blamey, Peter J. ; Busby, Peter A. ; Dowell, Richard C. ; Franz, Burkhard K-H. ; Musgrave, Gaye Nicholls ; Nienhuys, Terry G. ; Pyman, Brian C. ; Roberts, Susan A. ; Tong, Yit C. ; Webb, Robert L. ; Kuzma, Januz A. ; Money, David K. ; Patrick, James F. ; Seligman, Peter M. ( 1987)
    A multiple-electrode intracochlear implant that provides 21 stimulus channels has been designed for use in young children. It is smaller than the adult version and has magnets to facilitate the attachment of the headset. It has been implanted in two children aged 5 and 10 years. The two children both lost hearing in their third year, when they were still learning language. Following implantation, it was possible to determine threshold and comfortable listening levels for each electrode pair. This was facilitated in the younger child by prior training in scaling visual and electrotactile stimuli. Both children are regular users of the implant, and a training and assessment program has been commenced.
  • Item
    Thumbnail Image
    Banded intracochlear electrode array: evaluation of insertion trauma in human temporal bones
    Shepherd, R. K. ; Clark, Graeme M. ; Pyman, B. C. ; Webb, R. L. ( 1985)
    A banded free-fit scala tympani array was inserted into a basal turn of nine human cochleas to evaluate the trauma produced by the procedure. These nine cochleas, together with five nonimplanted controls, were serially sectioned and examined microscopically for damage to the membranous labyrinth, in particular the spiral ligament, the basilar and Reissner’s membranes, the stria vascularis, and the osseous spiral lamina. The severity and location of any trauma along the cochlear spiral were recorded. The results indicate that the insertion of the banded scala tympani array resulted in minimal mechanical damage, occurring primarily to a localized region of the spiral ligament. This would not result in significant neural degeneration, and therefore would not compromise the efficacy of multichannel cochlear prosthesis.
  • Item
    Thumbnail Image
    The histopathological effects of chronic electrical stimulation of the cat cochlea
    Shepherd, R. K. ; Clark, Graeme M. ; Black, R. C. ; Patrick, J. F. (Cambridge University Press, 1983)
    The success of a cochlear implant depends on stimulating an adequate number of viable spiral ganglion cells. The effect of chronic electrical stimulation on ganglion cells is therefore an important consideration when assessing the effectiveness and safety of such a device. The histopathological assessment of chronic unstimulated intracochlear electrodes is now well documented (Simmons, 1967; Clark, 1973; Clark et al, 1975; Schindler and Merzenich, 1974; Schindler, 1976; Schindler et al, 1977; Sutton et al, 1980). These experimental studies have used a variety of electrode designs, materials and surgical techniques. However, all have shown that chronic implantation has little effect on the peripheral nerves and the spiral ganglion cells adjacent to an implant, provided the insertion procedure is free of trauma and infection.
  • Item
    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve in cats: physiological and histopathological results
    Shepherd, R. K. ; Clark, Graeme M. ; Black, R. C. ( 1983)
    The ability of spiral ganglion cells to survive long-term electrical stimulation is a precondition for the success of cochlear prostheses. In this study 10 cats were implanted bilaterally with bipolar scala tympani electrodes and stimulated for periods of up to 2029 hours using charge balanced biphasic current pulses. The status of the auditory nerve was monitored periodically by recording electrically evoked auditory brainstem responses. At the conclusion of the stimulation program, spiral ganglion cell survival was assessed for stimulated and control cochleas; comparison of the two groups showed no statistically significant difference. The results of this study indicate that long-term intracochlear electrical stimulation using carefully controlled biphasic pulses does not adversely affect the spiral ganglion cell population.
  • Item
    Thumbnail Image
    Design and fabrication of the banded electrode array
    Clark, Graeme M. ; Shepherd, R. K. ; Patrick, James F. ; Black, R. C. ; Tong, Y. C. ( 1983)
    A multiple-channel electrode array must meet certain design requirements; these are listed in TABLE 1. First, there should be no trauma associated with the surgical insertion, and if there is a need to replace the array, this procedure should also be atraumatic. Second, it should be biologically inert. This means that it should be biocompatible with the tissues. When placed in the cochlea, the array should also not predispose the patient to local infection, and this is particularly important in children, in whom recurrent middle ear infections could spread to the inner ear. There should also be no risk of carcinogenicity with long-term implantation. Third, the electrode array should be designed so that the stimulus current can be localized to discrete groups of nerve fibers, and it should also be possible to stimulate as many groups as possible from the total remaining nerve population. Fourth, with long-term stimulation, there should be no significant corrosion of the electrodes used, and the electrical stimulation should not lead to damage of the tissues in the cochlea, especially the residual nerve fibers. Fifth, the electrode array should be mechanically robust and stable. It should not be prone to break as a result of repeated stress by the acceleration of the head during everyday movements. The array should also be capable of being fixed in place so that it will not shift its position. Sixth, it is desirable that the means of fabrication of the multiple-channel array should be simple and inexpensive. (From Introduction)
  • Item
    Thumbnail Image
    Investigation of curved intracochlear electrode arrays [Abstract]
    Xu, Shi-Ang ; Xu, J. ; Seldon, H Lee. ; Shepherd, R. K. ; Clark, G. M. ( 1992)
    It has been demonstrated that the Melbourne/Cochlear multi-channel cochlear implant is safe and effective for use in profoundly-totally deaf patients. Recent studies have highlighted the importance of deaf insertion and placing the electrodes closer to the spiral ganglion neurons. In order to improve the electrode insertion depth and proximity to the modiolus, we have investigated curved electrode arrays. Prototypes of such arrays and their accessory inserter have been made. Trial insertions were performed on skeletonized cochleae of human temporal bones. The preliminary results showed that, when compared with conventional straight electrode arrays, the curved arrays could be inserted deeper and located closer to the modiolus. These findings indicate that the curved --.~ electrodes currently under investigation should result in a reduction in stimulus threshold and improve pitch perception and may also result in the use of more channels of stimulation.