    Electrode discrimination by early-deafened subjects using the Cochlear Limited multiple electrode cochlear implant
    Busby, P. A. ; Clark, Graeme M. ( 2000)
    Objective: The aims of this study were to determine whether electrode discrimination by early-deafened subjects using the Cochlear Limited prosthesis varied at different locations on the electrode array, was influenced by the effects of auditory deprivation and experience with electric stimulation, and was related to speech perception. Design: Difference limens for electrode discrimination were measured in 16 early-deafened subjects at three positions on the array: electrodes 18 (apical), 14 (mid), and 8 (basal). Electrodes were stimulated using random variations in current level to minimize the influence of loudness cues. Assessed were correlations between the difference limens, subject variables related to auditory deprivation (age at onset of deafness, duration of deafness, and age at implantation) and auditory experience (duration of implant use and the total time period of auditory experience), and speech perception scores from two closed-set and two open-set tests. Results: The average difference limens across the three positions were less than two electrodes for 75%, of subjects, with average limens between 2 and 6.5 electrodes for the remaining 25% of subjects. Significant differences across the three positions were found for 69% of subjects. The average limens and those at the basal position positively correlated with variables related to auditory deprivation, with larger limens for subjects implanted at a later age and with a longer duration of deafness. The average limens and those at the apical position negatively correlated with closed-set speech perception scores, with lower scores for subjects with larger limens, but not with open-set scores. Speech scores also negatively correlated with variables related to auditory deprivation. Conclusions: These findings showed that early-deafened subjects were generally successful in electrode discrimination although performance varied across the array for over half the subjects. Discrimination performance was influenced by the effects of auditory deprivation, and both electrode discrimination and variables related to auditory deprivation influenced closed-set speech perception.
  • Item
    Thumbnail Image
    Pitch estimation by early-deafened subjects using a multiple-electrode cochlear implant
    Busby, P. A. ; Clark, Graeme M. ( 2000)
    Abstract not available due to copyright.
  • Item
    Thumbnail Image
    Short-term auditory memory in children using cochlear implants and its relevance to receptive language
    Dawson, P. W. ; Busby, P. A. ; McKay, C. M. ; Clark, Graeme M. ( 2002)
    The aim of this study was to assess auditory sequential, short-term memory (SSTM) performance in young children using cochlear implants (CI group) and to examine the relationship of this performance to receptive language performance. Twenty-four children, 5 to 11 years old, using the Nucleus 22-electrode cochlear implant, were tested on a number of auditory and visual tasks of SSTM. The auditory memory tasks were designed to minimize the effect of auditory discrimination ability. Stimuli were chosen that children with cochlear implants could accurately identify with a reaction time similar to that of a control group of children with normal hearing (NH group). All children were also assessed on a receptive language test and on a nonverbal intelligence scale. As expected, children using cochlear implants demonstrated poorer auditory and visual SSTM skills than their hearing peers when the stimuli were verbal or were pictures that could be readily labelled. They did not differ from their peers with normal hearing on tasks where the stimuli were less likely to be verbally encoded. An important finding was that the CI group did not appear to have a sequential memory deficit specific to the auditory modality. The difference scores (auditory minus visual memory performance) for the CI group were not significantly different from those for the NH group. SSTM performance accounted for significant variance in the receptive language performance of the CI group. However, a forward stepwise regression analysis revealed that visual spatial memory (one of the subtests of the nonverbal IQ test) was the main predictor of variance in the language scores of the children using cochlear implants.
  • Item
    Thumbnail Image
    Historia del implante cochlear nucleus
    Clark, Graeme M. (Masson, 2002)
  • Item
    Thumbnail Image
    A comparison of a new prototype Tickle Talker with a Tactaid 7
    Galvin, Karyn L. ; Ginis, Jan ; Cowan, Robert S. C. ; Blamey, Peter J. ; Clark, Graeme M. ( 2001)
    This study compared the speech perception enhancement provided by two multichannel tactile aids: a new version of the Tickle TalkerT™ and the Tactaid 7. The subjects' impression of benefit was also examined. In an AB pattern, six adults with hearing impairment used each device daily for approximately 18 weeks and attended 12 training sessions. When tactile information was provided, the group demonstrated a significant enhancement for the perception of words (mean 17.2%) and phonemes (mean 12.9%) in monosyllabic word lists, words in sentences (mean 14.2%) and speech tracking (mean 7.7 wpm). The Tactaid 7 provided a significantly greater enhancement for the perception of words (21 % versus 13.4%), phonemes (16.7% versus 9.1%) and some speech features in monosyllabic word lists. Subjective ratings were slightly higher for the Tactaid 7, and four subjects preferred this device. Either device may be suitable for those not able or willing to have a cochlear implant.
  • Item
    Thumbnail Image
    Cochlear implants: climbing new mountains (The Graham Fraser memorial lecture)
    Clark, Graeme M. ( 2001)
    This 7th annual lecture was given as a special tribute to Graham Fraser (Figure 1). His enthusiasm, drive and constant search to find new ways to help deaf people, particularly with cochlear implants, has been most impressive. I remember with affection the visit Graham and his wife, Pat, made to Melbourne in 1992. There were lively discussions the past and future of cochlear implants. He would have had much to say about the new directors for the next decade. I hope to summarize some of those possibilities, and will refer in particular to research at our centres in Melbourne.
  • Item
    Thumbnail Image
    Surgical implications of perimodiolar cochlear implant electrode design: avoiding intracochlear damage and scala vestibuli insertion.
    Briggs, RJ ; Tykocinski, M ; Saunders, E ; Hellier, W ; Dahm, M ; Pyman, B ; Clark, GM (Informa UK Limited, 2001-09)
    OBJECTIVE: To review the mechanisms and nature of intracochlear damage associated with cochlear implant electrode array insertion, in particular, the various perimodiolar electrode designs. Make recommendations regarding surgical techniques for the Nucleus Contour electrode to ensure correct position and minimal insertion trauma. BACKGROUND: The potential advantages of increased modiolar proximity of intracochlear multichannel electrode arrays are a reduction in stimulation thresholds, an increase in dynamic range and more localized neural excitation. This may improve speech perception and reduce power consumption. These advantages may be negated if increased intracochlear damage results from the method used to position the electrodes close to the modiolus. METHOD: A review of the University of Melbourne Department of Otolaryngology experience with temporal bone safety studies using the Nucleus standard straight electrode array and a variety of perimodiolar electrode array designs; comparison with temporal bone insertion studies from other centres and postmortem histopathology studies reported in the literature. Review of our initial clinical experience using the Nucleus Contour electrode array. RESULTS: The nature of intracochlear damage resulting from electrode insertion trauma ranges from minor, localized, spiral ligament tear to diffuse organ of Corti disruption and osseous spiral lamina fracture. The type of damage depends on the mechanical characteristics of the electrode array, the stiffness, curvature and size of the electrode in relation to the scala, and the surgical technique. The narrow, flexible, straight arrays are the least traumatic. Pre-curved or stiffer arrays are associated with an incidence of basilar membrane perforation. The cochleostomy must be correctly sited in relation to the round window to ensure scala tympani insertion. A cochleostomy anterior to the round window rather than inferior may lead to scala media or scala vestibuli insertion. CONCLUSION: Proximity of electrodes to the modiolus can be achieved without intracochlear damage provided the electrode array is a free fit within the scala, of appropriate size and shape, and accurate scala tympani insertion is performed.
  • Item
    Thumbnail Image
    The development of speech perception in children using cochlear implants: effects of etiologic factors and delayed milestones
    PYMAN, BRIAN ; Blamey, Peter J. ; Lacy, Peter ; Clark, Graeme M. ; DOWELL, RICHARD ( 2000)
    Hypothesis: Speech perception outcomes for cochlear implantation of children vary over a wide range, and it is hypothesized that central pathologic states associated with certain causes of hearing impairment account for a substantial part of the variance. Study Design: A retrospective analysis was carried out to ascertain the relationships between speech perception, etiologic factors, and central pathologic states as indicated by preoperative delayed motor milestones and/or cognitive delays. Setting: Data were obtained from the pre-and postoperative records of patients attending a hospital cochlear implant clinic. Patients: Results for 75 consecutive patients up to age 5 years who underwent implantation were included in the study. Intervention: Patients received a 22-electrode cochlear prosthesis and were seen by the clinic for regular tune-up and assessments. Home-and school-based habilitation was recommended by the clinic. Main Outcome Measures: Speech perception measures were classified on a five-point scale to allow for different evaluation procedures at different ages and developmental stages. Results: The incidence of motor and cognitive delays were fairly evenly spread across etiologic factors, except for cytomegalovirus, which had a much higher than average incidence. Children with motor and/or cognitive delays were significantly slower than other children in the development of speech perception skills after implantation. Etiologic factors did not have a statistically significant effect on speech perception outcome. Conclusions: It is likely that central pathologic states account for a substantial part of the variance among children using cochlear implants. Specific indicators of central pathologic states should be used to assess a child's prognosis in preference to less specific information based on etiologic factors alone.
  • Item
    Thumbnail Image
    The relationship between speech perception and electrode discrimination in cochlear implantees
    Henry, Belinda A. ; McKay, Colette M. ; McDermott, Hugh J. ; Clark, Graeme M. ( 2000)
    Abstract not available due to copyright.
  • Item
    Thumbnail Image
    Electrode discrimination and speech perception in young children using cochlear implants
    Dawson, P. W. ; McKay, C. M. ; Busby, P. A. ; Grayden, D. B. ; Clark, Graeme M. ( 2000)
    Objective: The aim was to determine the efficacy of a child-appropriate procedure to assess electrode discrimination ability in young children using cochlear implants and to investigate the relationship of electrode discrimination ability and speech perception performance in children implanted at a young age. Design: An adaptation of the play audiometry procedure was used to assess electrode discrimination in seventeen 4- to 10-yr-old children. The children were required to respond with a game-like motor response when a repeating stimulation on a reference electrode “changed” to a different electrode. They were also assessed on a speech feature discrimination test, a closed-set word recognition test and a nonverbal intelligence task. Results: Sixty-five percent of subjects demonstrated ability to discriminate adjacent electrodes in mid and apical regions of the cochlea, whilst the remaining subjects needed electrode separations of between two and nine electrodes for successful discrimination. In a forward stepwise regression analysis electrode discrimination ability was found to be the strongest factor in accounting for variance in the speech perception scores. Subject variables such as duration of deafness, nonverbal intelligence and implant experience did not significantly account for further variance in the speech perception scores for this group of children. Conclusions: Electrode discrimination ability was the strongest factor in predicting performance on speech perception measures in a group of children using cochlear implants.