    Chronic electrical stimulation of the auditory nerve using non-charge-balanced stimuli
    Shepherd, Robert K. ; Linahan, N. ; Xu, J. ; Clark, Graeme M. ; Araki, S. ( 1999)
    This study was designed to evaluate the pathophysiological response of the cochlea following long-term intracochlear electrical stimulation using a poorly charge-balanced stimulus regime, leading to direct current (DC) levels >0.1 µA. Four normal-hearing adult cats were bilaterally implanted with scala tympani electrode arrays and unilaterally stimulated for periods up to 2200 h. Stimuli consisted of 50 µs monophasic current pulses presented at 2000 pulses per second (pps) per channel, and resulted in DC levels of 0.4-2.8 µA. Both acoustic and electrical (EABR) evoked potentials were periodically recorded during the stimulation program. Frequency-specific stimuli indicated that an extensive and widespread hearing loss occurred over the 4-24 KHz region in all stimulated cochleae, although the 2 KHz region exhibited thresholds close to normal in some animals, despite long-term implantation and chronic stimulation. Longitudinal EABRs showed a statistically significant increase in threshold for three of the four animals. Histopathological evaluation of the cochleae revealed a highly significant reduction in ganglion cell density in stimulated cochleae compared with their controls. Spiral ganglion cell loss was significantly correlated with the degree of inflammation, duration of electrical stimulation, and the level of DC. In conclusion, the present study highlights the potential for neural damage following stimulation using poorly charge-balanced stimuli.
  • Item
    Thumbnail Image
    Cochlear pathology following chronic electrical stimulation using non charge balanced stimuli
    Shepherd, Robert K. ; Matsushima, Jun-Ichi ; Millard, R. E. ; Clark, Graeme M. ( 1991)
    During the course of a chronic intracochlear electrical stimulation study using charge balanced biphasic current pulses, one animal inadvertently received a short period of direct current (DC) stimulation at a level of approximately 1 µA. Subsequent, the animal was chronically stimulated using a poorly charge balanced waveform that produced a DC level of approximately 2 µA. Extensive pathological changes were observed within the cochlea. These changes included widespread spiral ganglion cell loss and new bone growth that extended throughout all turns of the cochlea. Significant changes in the morphology of the electrically evoked auditory brainstem response (EABR) were associated with these pathological changes. EABRs recorded prior to the DC stimulation exhibited a normal waveform morphology. However, responses recorded during the course of the DC stimulation were dominated by a short latency response believed to be vestibular in origin. The response thresholds were also significantly higher than levels recorded before the DC stimulation. In contrast, the contralateral cochlea, stimulated using charge balanced stimuli, showed no evidence of adverse pathological changes. Furthermore, EABRs evoked from this cochlea remained stable throughout the chronic stimulation period. Although preliminary, the present results illustrate the adverse nature of poorly charge balanced electrical stimuli. These results have important implications for both the design of neural prostheses and the use of DC stimuli to suppress tinnitus in patients.