Graeme Clark Collection

Permanent URI for this collection

Search Results

Now showing 1 - 2 of 2
  • Item
    Thumbnail Image
    The multi-channel cochlear implant and the relief of severe-to-profound deafness.
    Clark, G (Informa UK Limited, 2012-05)
    This personal reflection outlines the discoveries at the University of Melbourne leading to the multi-channel cochlear implant, and its development industrially by Cochlear Limited. My earlier experimental electrophysiological research demonstrated temporal coding occurred for only low frequencies, i.e. below 200-500 pulses/second. I was able to confirm these findings perceptually in behaviourally conditioned animals. In addition, these studies showed that temporal discrimination occurred across spatial coding channels. These experimental results correlated with the later conscious experience for electrical stimulation in my implant patients. In addition, the mid-to-high frequencies were coded in part by place of stimulation using bipolar and monopolar stimulation to restrict current spread. Furthermore, place of stimulation had the qualities of sharpness and dullness, and was also experienced as vowels. Owing to the limitation in coding speech with a physiological model due to the overlap of electrical current leading to unpredictable variations in loudness, a speech coding strategy that extracted the most important speech features for transmission through an electro-neural 'bottle-neck' to the brain was explored. Our inaugural strategy, discovered in 1978, extracted the second formant for place of stimulation, voicing for rate of stimulation, and sound pressure for current level. This was the first coding strategy to provide open-set speech understanding, as shown by standard audiological tests, and it became the first clinically successful interface between the world and human consciousness. This strategy was improved with place coding for the third formant or high-frequency spectrum, and then the spectral maxima. In 1989, I operated on our first patient to receive a bilateral implant, and in 1990, the first with a bimodal processor. The psychophysics and speech perception for these showed that the stimuli from each side could be fused into a single image, and localized according to differences in intensity and time of arrival of the stimuli. There were significant improvements for speech perception in noise. In 1985, I implanted our first children with the multi-channel prosthesis and found that speech understanding and spoken language were greatly improved the younger the child at surgery, and especially when younger than 12 months. Speech understanding was strongly related to the development of place coding. In 1990, the US Food and Drug Administration approved the implant for deaf children, the first by any world health regulatory body making it the first major advance in helping deaf children to communicate.
  • Item
    Thumbnail Image
    Midbrain responses to micro-stimulation of the cochlea using high density thin-film arrays
    Allitt, BJ ; Morgan, SJ ; Bell, S ; Nayagam, DAX ; Arhatari, B ; Clark, GM ; Paolini, AG (ELSEVIER SCIENCE BV, 2012-05)
    A broader activation of auditory nerve fibres than normal using a cochlear implant contributes to poor frequency discrimination. As cochlear implants also deliver a restricted dynamic range, this hinders the ability to segregate sound sources. Better frequency coding and control over amplitude may be achieved by limiting current spread during electrical stimulation of the cochlea and positioning electrodes closer to the modiolus. Thin-film high density microelectrode arrays and conventional platinum ring electrode arrays were used to stimulate the cochlea of urethane-anaesthetized rats and responses compared. Neurophysiological recordings were taken at 197 multi-unit clusters in the central nucleus of the inferior colliculus (CIC), a site that receives direct monaural innervation from the cochlear nucleus. CIC responses to both the platinum ring and high density electrodes were recorded and differences in activity to changes in stimulation intensity, thresholds and frequency coding of neural activation were examined. The high density electrode array elicited less CIC activity at nonspecific frequency regions than the platinum ring electrode array. The high density electrode array produced significantly lower thresholds and larger dynamic ranges than the platinum ring electrode array when positioned close to the modiolus. These results suggest that a higher density of stimulation sites on electrodes that effectively 'aim' current, combined with placement closer to the modiolus would permit finer control over charge delivery. This may equate to improved frequency specific perception and control over amplitude when using future cochlear implant devices.