    Decrement in auditory nerve function following acute high rate stimulation in guinea pigs [Abstract]
    Tykocinski, M. ; Shepherd, R. K. ; Clark, Graeme M. ( 1995)
    Cochlear implants have been shown to successfully provide profoundly deaf patients with auditory cues for speech discrimination. Psychophysical studies suggested that speech processing strategies based on stimulus rates of up to 1000 pulses per second (pps) may lead to an improvement in speech perception, due to a better representation of the rapid variations in the amplitude of speech. However, "neural fatigue" has been known to occur following brief periods of electrical stimulation at rates high enough to ensure that stimuli occur within the neurons relative refractory period, and has been shown to depend on stimulus duration and rate of the evoked neural activity. Prolonged electrical stimulation at these high stimulus rates could, therefore, have an adverse effect on the neurons metabolism and result in cellular energy depletion.
  • Item
    Thumbnail Image
    Investigation of curved intracochlear electrode arrays [Abstract]
    Xu, Shi-Ang ; Xu, J. ; Seldon, H Lee. ; Shepherd, R. K. ; Clark, G. M. ( 1992)
    It has been demonstrated that the Melbourne/Cochlear multi-channel cochlear implant is safe and effective for use in profoundly-totally deaf patients. Recent studies have highlighted the importance of deaf insertion and placing the electrodes closer to the spiral ganglion neurons. In order to improve the electrode insertion depth and proximity to the modiolus, we have investigated curved electrode arrays. Prototypes of such arrays and their accessory inserter have been made. Trial insertions were performed on skeletonized cochleae of human temporal bones. The preliminary results showed that, when compared with conventional straight electrode arrays, the curved arrays could be inserted deeper and located closer to the modiolus. These findings indicate that the curved --.~ electrodes currently under investigation should result in a reduction in stimulus threshold and improve pitch perception and may also result in the use of more channels of stimulation.
  • Item
    Thumbnail Image
    Chronic electrical stimulation of the auditory nerve in cats
    Shepherd, R. K. ; Clark, Graeme M. ; Black, R. C. ; Patrick, J. F. ( 1982)
    One requirement for the success of a cochlear hearing prosthesis is that long-term electrical stimulation must not have adverse effects on the residual spiral ganglion cell population. Electrochemically 'safe' stimulation regimes have been defined for the cortex (Brummer &Turner, 1977). However, few investigators have examined the effects of long-term intracochlear electrical stimulation. Walsh et al (1980), stimulating with current densities greater than the 'safe' limits defined by Brummer &Turner (1977), for periods of up to 800 hours at current levels of 4.0-8.0 mA, recorded slight local neural degeneration adjacent to the electrodes.