  • Item
    Thumbnail Image
    Interobserver Agreement of Electrode to Retina Distance Measurements in a Second-Generation (44-Channel) Suprachoroidal Retinal Prosthesis
    Abbott, CJ ; Baglin, EK ; Kolic, M ; McGuinness, MB ; Titchener, SA ; Young, KA ; Yeoh, J ; Luu, CD ; Ayton, LN ; Petoe, MA ; Allen, PJ (ASSOC RESEARCH VISION OPHTHALMOLOGY INC, 2022-09)
    PURPOSE: The electrode to retina (ER) distance is an important contributory factor to the safety and efficacy of a suprachoroidal retinal prosthesis. Measuring ER distance may be performed by different observers during multisite studies. The aim of this study was to assess the interobserver agreement in measuring ER distance. METHODS: Three independent, trained observers measured ER distance from the center of each suprachoroidal electrode to the inner retinal pigment epithelium in spectral-domain optical coherence tomography (SD-OCT) B-scans. A total of 121 ER distance measurements from 77 B-scans collected over 5 months from one subject implanted with a second-generation 44-channel suprachoroidal retinal prosthesis (NCT03406416) were made by each observer. RESULTS: ER distance ranged from 208 to 509 µm. Pearson's correlation coefficient (ρ) showed agreement of 0.99 (95% confidence interval [CI] = 0.98-0.99) in measuring ER for each pairwise comparison. The mean difference in ER distance between observers ranged from 2.4 to 6.4 µm with pairwise limits of agreement (95% CI) of ±20 µm (5.5% of mean). Intraclass correlation coefficient (ICC) showed agreement of 0.98 (95% CI = 0.97-0.99) between observers. CONCLUSIONS: There is high agreement in measuring ER distances for suprachoroidal retinal prostheses using our systematic approach between multiple, trained observers, supporting the use of a single observer for each image. TRANSLATIONAL RELEVANCE: High interobserver agreement outcomes indicate that multiple, trained observers can be used to take ER measurements across different images in suprachoroidal retinal prosthesis studies. This improves multisite study efficiency and gives confidence in interpreting results relating to the safety and efficacy of suprachoroidal retinal prostheses.
  • Item
    Thumbnail Image
    A Second-Generation (44-Channel) Suprachoroidal Retinal Prosthesis: Long-Term Observation of the Electrode-Tissue Interface
    Titchener, SA ; Nayagam, DAX ; Kvansakul, J ; Kolic, M ; Baglin, EK ; Abbott, CJ ; McGuinness, MB ; Ayton, LN ; Luu, CD ; Greenstein, S ; Kentler, WG ; Shivdasani, MN ; Allen, PJ ; Petoe, MA (ASSOC RESEARCH VISION OPHTHALMOLOGY INC, 2022-06)
    PURPOSE: To report the long-term observations of the electrode-tissue interface and perceptual stability in humans after chronic stimulation with a 44-channel suprachoroidal retinal implant. METHODS: Four subjects (S1-4) with end-stage retinitis pigmentosa received the implant unilaterally (NCT03406416). Electrode impedances, electrode-retina distance (measured using optical coherence tomography imaging), and perceptual thresholds were monitored up to 181 weeks after implantation as the subjects used the prosthesis in the laboratory and in daily life. Stimulation charge density was limited to 32 µC/cm2 per phase. RESULTS: Electrode impedances were stable longitudinally. The electrode-retina distances increased after surgery and then stabilized, and were well-described by an asymptotic exponential model. The stabilization of electrode-retina distances was variable between subjects, stabilizing after 45 weeks for S1, 63 weeks for S2, and 24 weeks for S3 (linear regression; Pgradient > 0.05). For S4, a statistically significant increase in electrode-retina distance persisted (P < 0.05), but by the study end point the rate of increase was clinically insignificant (exponential model: 0.33 µm/wk). Perceptual electrical thresholds were stable in one subject, decreased over time in two subjects (linear model; P < 0.05), and increased slightly in one subject but remained within the predefined charge limits (P = 0.02). CONCLUSIONS: Chronic stimulation with the suprachoroidal retinal prosthesis over 3 years resulted in stable impedances, small individual changes in perceptual electrical thresholds, and no clinically significant increase in electrode-retina distances after a period of settling after surgery. TRANSLATIONAL RELEVANCE: Chronic stimulation with the 44-channel suprachoroidal retinal implant with a charge density of up to 32 µC/cm2 per phase is suitable for long-term use in humans.
  • Item
    Thumbnail Image
    A Second-Generation (44-Channel) Suprachoroidal Retinal Prosthesis: Interim Clinical Trial Results
    Petoe, MA ; Titchener, SA ; Kolic, M ; Kentler, WG ; Abbott, CJ ; Nayagam, DAX ; Baglin, EK ; Kvansakul, J ; Barnes, N ; Walker, JG ; Epp, SB ; Young, KA ; Ayton, LN ; Luu, CD ; Allen, PJ (ASSOC RESEARCH VISION OPHTHALMOLOGY INC, 2021-12)
    PURPOSE: To report the initial safety and efficacy results of a second-generation (44-channel) suprachoroidal retinal prosthesis at 56 weeks after device activation. METHODS: Four subjects, with advanced retinitis pigmentosa and bare-light perception only, enrolled in a phase II trial (NCT03406416). A 44-channel electrode array was implanted in a suprachoroidal pocket. Device stability, efficacy, and adverse events were investigated at 12-week intervals. RESULTS: All four subjects were implanted successfully and there were no device-related serious adverse events. Color fundus photography indicated a mild postoperative subretinal hemorrhage in two recipients, which cleared spontaneously within 2 weeks. Optical coherence tomography confirmed device stability and position under the macula. Screen-based localization accuracy was significantly better for all subjects with device on versus device off. Two subjects were significantly better with the device on in a motion discrimination task at 7, 15, and 30°/s and in a spatial discrimination task at 0.033 cycles per degree. All subjects were more accurate with the device on than device off at walking toward a target on a modified door task, localizing and touching tabletop objects, and detecting obstacles in an obstacle avoidance task. A positive effect of the implant on subjects' daily lives was confirmed by an orientation and mobility assessor and subject self-report. CONCLUSIONS: These interim study data demonstrate that the suprachoroidal prosthesis is safe and provides significant improvements in functional vision, activities of daily living, and observer-rated quality of life. TRANSLATIONAL RELEVANCE: A suprachoroidal prosthesis can provide clinically useful artificial vision while maintaining a safe surgical profile.
  • Item
    Thumbnail Image
    Functional Vision in the Real-World Environment With a Second-Generation (44-Channel) Suprachoroidal Retinal Prosthesis
    Karapanos, L ; Abbott, CJ ; Ayton, LN ; Kolic, M ; McGuinness, MB ; Baglin, EK ; Titchener, SA ; Kvansakul, J ; Johnson, D ; Kentler, WG ; Barnes, N ; Nayagam, DAX ; Allen, PJ ; Petoe, MA (ASSOC RESEARCH VISION OPHTHALMOLOGY INC, 2021-08)
    PURPOSE: In a clinical trial (NCT03406416) of a second-generation (44-channel) suprachoroidal retinal prosthesis implanted in subjects with late-stage retinitis pigmentosa (RP), we assessed performance in real-world functional visual tasks and emotional well-being. METHODS: The Functional Low-Vision Observer Rated Assessment (FLORA) and Impact of Vision Impairment-Very Low Vision (IVI-VLV) instruments were administered to four subjects before implantation and after device fitting. The FLORA contains 13 self-reported and 35 observer-reported items ranked for ease of conducting task (impossible-easy, central tendency given as mode). The IVI-VLV instrument quantified the impact of low vision on daily activities and emotional well-being. RESULTS: Three subjects completed the FLORA for two years after device fitting; the fourth subject ceased participation in the FLORA after fitting for reasons unrelated to the device. For all subjects at each post-fitting visit, the mode ease of task with device ON was better or equal to device OFF. Ease of task improved over the first six months with device ON, then remained stable. Subjects reported improvements in mobility, functional vision, and quality of life with device ON. The IVI-VLV suggested self-assessed vision-related quality of life was not impacted by device implantation or usage. CONCLUSIONS: Subjects demonstrated sustained improved ease of task scores with device ON compared to OFF, indicating the device has a positive impact in the real-world setting. TRANSLATIONAL RELEVANCE: Our suprachoroidal retinal prosthesis shows potential utility in everyday life, by enabling an increased environmental awareness and improving access to sensory information for people with end-stage RP.
  • Item
    Thumbnail Image
    Oculomotor Responses to Dynamic Stimuli in a 44-Channel Suprachoroidal Retinal Prosthesis
    Titchener, SA ; Kvansakul, J ; Shivdasani, MN ; Fallon, JB ; Nayagam, DAX ; Epp, SB ; Williams, CE ; Barnes, N ; Kentler, WG ; Kolic, M ; Baglin, EK ; Ayton, LN ; Abbott, CJ ; Luu, CD ; Allen, PJ ; Petoe, MA (ASSOC RESEARCH VISION OPHTHALMOLOGY INC, 2020-12)
    PURPOSE: To investigate oculomotor behavior in response to dynamic stimuli in retinal implant recipients. METHODS: Three suprachoroidal retinal implant recipients performed a four-alternative forced-choice motion discrimination task over six sessions longitudinally. Stimuli were a single white bar ("moving bar") or a series of white bars ("moving grating") sweeping left, right, up, or down across a 42″ monitor. Performance was compared with normal video processing and scrambled video processing (randomized image-to-electrode mapping to disrupt spatiotemporal structure). Eye and head movement was monitored throughout the task. RESULTS: Two subjects had diminished performance with scrambling, suggesting retinotopic discrimination was used in the normal condition and made smooth pursuit eye movements congruent to the moving bar stimulus direction. These two subjects also made stimulus-related eye movements resembling optokinetic reflex (OKR) for moving grating stimuli, but the movement was incongruent with stimulus direction. The third subject was less adept at the task, appeared primarily reliant on head position cues (head movements were congruent to stimulus direction), and did not exhibit retinotopic discrimination and associated eye movements. CONCLUSIONS: Our observation of smooth pursuit indicates residual functionality of cortical direction-selective circuits and implies a more naturalistic perception of motion than expected. A distorted OKR implies improper functionality of retinal direction-selective circuits, possibly due to retinal remodeling or the non-selective nature of the electrical stimulation. TRANSLATIONAL RELEVANCE: Retinal implant users can make naturalistic eye movements in response to moving stimuli, highlighting the potential for eye tracker feedback to improve perceptual localization and image stabilization in camera-based visual prostheses.
  • Item
    Thumbnail Image
    Advances in implantable bionic devices for blindness: a review
    Lewis, PM ; Ayton, LN ; Guymer, RH ; Lowery, AJ ; Blamey, PJ ; Allen, PJ ; Luu, CD ; Rosenfeld, JV (WILEY, 2016-09)
    Since the 1950s, vision researchers have been working towards the ambitious goal of restoring a functional level of vision to the blind via electrical stimulation of the visual pathways. Groups based in Australia, USA, Germany, France and Japan report progress in the translation of retinal visual prosthetics from the experimental to clinical domains, with two retinal visual prostheses having recently received regulatory approval for clinical use. Regulatory approval for cortical visual prostheses is yet to be obtained; however, several groups report plans to conduct clinical trials in the near future, building upon the seminal clinical studies of Brindley and Dobelle. In this review, we discuss the general principles of visual prostheses employing electrical stimulation of the visual pathways, focusing on the retina and visual cortex as the two most extensively studied stimulation sites. We also discuss the surgical and functional outcomes reported to date for retinal and cortical prostheses, concluding with a brief discussion of novel developments in this field and an outlook for the future.
  • Item
    Thumbnail Image
    Development of a surgical procedure for implantation of a prototype suprachoroidal retinal prosthesis
    Saunders, AL ; Williams, CE ; Heriot, W ; Briggs, R ; Yeoh, J ; Nayagam, DAX ; McCombe, M ; Villalobos, J ; Burns, O ; Luu, CD ; Ayton, LN ; McPhedran, M ; Opie, NL ; McGowan, C ; Shepherd, RK ; Guymer, R ; Allen, PJ (WILEY, 2014)
    BACKGROUND: Current surgical techniques for retinal prosthetic implantation require long and complicated surgery, which can increase the risk of complications and adverse outcomes. METHOD: The suprachoroidal position is known to be an easier location to access surgically, and so this study aimed to develop a surgical procedure for implanting a prototype suprachoroidal retinal prosthesis. The array implantation procedure was developed in 14 enucleated eyes. A full-thickness scleral incision was made parallel to the intermuscular septum and superotemporal to the lateral rectus muscle. A pocket was created in the suprachoroidal space, and the moulded electrode array was inserted. The scleral incision was closed and scleral anchor point sutured. In 9 of the 14 eyes examined, the device insertion was obstructed by the posterior ciliary neurovascular bundle. Subsequently, the position of this neurovascular bundle in 10 eyes was characterized. Implantation and lead routing procedure was then developed in six human cadavers. The array was tunnelled forward from behind the pinna to the orbit. Next, a lateral canthotomy was made. Lead fixation was established by creating an orbitotomy drilled in the frontal process of the zygomatic bone. The lateral rectus muscle was detached, and implantation was carried out. Finally, pinna to lateral canthus measurements were taken on 61 patients in order to determine optimal lead length. RESULTS: These results identified potential anatomical obstructions and informed the anatomical fitting of the suprachoroidal retinal prosthesis. CONCLUSION: As a result of this work, a straightforward surgical approach for accurate anatomical suprachoroidal array and lead placement was developed for clinical application.
  • Item
    Thumbnail Image
    First-in-Human Trial of a Novel Suprachoroidal Retinal Prosthesis
    Ayton, LN ; Blamey, PJ ; Guymer, RH ; Luu, CD ; Nayagam, DAX ; Sinclair, NC ; Shivdasani, MN ; Yeoh, J ; McCombe, MF ; Briggs, RJ ; Opie, NL ; Villalobos, J ; Dimitrov, PN ; Varsamidis, M ; Petoe, MA ; McCarthy, CD ; Walker, JG ; Barnes, N ; Burkitt, AN ; Williams, CE ; Shepherd, RK ; Allen, PJ ; Mori, K (PUBLIC LIBRARY SCIENCE, 2014-12-18)
    UNLABELLED: Retinal visual prostheses ("bionic eyes") have the potential to restore vision to blind or profoundly vision-impaired patients. The medical bionic technology used to design, manufacture and implant such prostheses is still in its relative infancy, with various technologies and surgical approaches being evaluated. We hypothesised that a suprachoroidal implant location (between the sclera and choroid of the eye) would provide significant surgical and safety benefits for patients, allowing them to maintain preoperative residual vision as well as gaining prosthetic vision input from the device. This report details the first-in-human Phase 1 trial to investigate the use of retinal implants in the suprachoroidal space in three human subjects with end-stage retinitis pigmentosa. The success of the suprachoroidal surgical approach and its associated safety benefits, coupled with twelve-month post-operative efficacy data, holds promise for the field of vision restoration. TRIAL REGISTRATION: Clinicaltrials.gov NCT01603576.