Diurnal Intraocular Pressure Fluctuations with Self-tonometry in Glaucoma Patients and Suspects: A Clinical Trial

Download
Author
Huang, J; Katalinic, P; Kalloniatis, M; Hennessy, MP; Zangerl, BDate
2018-02-01Source Title
Optometry and Vision SciencePublisher
LIPPINCOTT WILLIAMS & WILKINSUniversity of Melbourne Author/s
Kalloniatis, MichaelAffiliation
Anatomy and NeuroscienceMetadata
Show full item recordDocument Type
Journal ArticleCitations
Huang, J., Katalinic, P., Kalloniatis, M., Hennessy, M. P. & Zangerl, B. (2018). Diurnal Intraocular Pressure Fluctuations with Self-tonometry in Glaucoma Patients and Suspects: A Clinical Trial. OPTOMETRY AND VISION SCIENCE, 95 (2), pp.88-95. https://doi.org/10.1097/OPX.0000000000001172.Access Status
Open AccessAbstract
SIGNIFICANCE: This article shows that self-tonometry can provide robust measures of diurnal intraocular pressure (IOP) and also detect changes to IOP in response to treatment within a short period of monitoring. These advances in IOP monitoring may contribute to improved management of glaucoma patients and suspects. PURPOSE: The aim of this study was to prospectively investigate the utility of rebound self-tonometry performed over several weeks for detecting diurnal IOP fluctuations in glaucoma patients and suspects and also initial response to topical treatment in glaucoma patients. METHODS: Forty patients were recruited following glaucoma-specific examination. Subsequent to successful training with the iCare HOME tonometer, patients were instructed to measure IOP, in a sitting position, four times a day over 4 to 6 weeks. Date, time, laterality, and IOP downloaded from the tonometer and clinical examination data, including applanation IOP and corneal thickness, were analyzed. A user satisfaction survey was also administered at study completion. t Test and analysis of variance were used to compare groups and IOP across days. Pearson correlation was used to compare measurements to Goldmann applanation tonometry and IOP measurements from the first day/s to the overall mean IOP. RESULTS: Twenty-seven patients (18 suspects and 9 glaucoma patients) completed data collection. Patients self-measured IOP on 118 (±29) occasions for 40 (±7.4) days. Two dominant patterns of fluctuation were revealed: peak IOP upon awakening (n = 11) and at midday (n = 13). Diurnal IOP measured in the first 7 days showed strong correlation to diurnal IOP across the entire study period (r = 0.82, P < .0001). Within 24 hours of treatment commencement (latanoprost 0.005% ophthalmic solution), IOP reduced from 23.9 (±5.2) to 16.1 (±2.6) mmHg. Overall, patients rated the instrument as easy to use, although difficulties with correct alignment were expressed. CONCLUSIONS: Rebound self-tonometry demonstrated utility for measuring diurnal IOP fluctuations in most patients, hence enhancing management of patient with or at risk of developing glaucoma.
Export Reference in RIS Format
Endnote
- Click on "Export Reference in RIS Format" and choose "open with... Endnote".
Refworks
- Click on "Export Reference in RIS Format". Login to Refworks, go to References => Import References