    EORTC PET response criteria are more influenced by reconstruction inconsistencies than PERCIST but both benefit from the EARL harmonization program

    Thumbnail
    Download
    Published version (1.266Mb)

    Citations
    Scopus
    Altmetric
    7
    Author
    Lasnon, C; Quak, E; Le Roux, P-Y; Robin, P; Hofman, MS; Bourhis, D; Callahan, J; Binns, DS; Desmonts, C; Salaun, P-Y; ...
    Date
    2017-05-30
    Source Title
    EJNMMI Physics
    Publisher
    SPRINGEROPEN
    University of Melbourne Author/s
    Hofman, Michael; Hicks, Rodney
    Affiliation
    Sir Peter MacCallum Department of Oncology
    Medicine and Radiology
    Metadata
    Show full item record
    Document Type
    Journal Article
    Citations
    Lasnon, C., Quak, E., Le Roux, P. -Y., Robin, P., Hofman, M. S., Bourhis, D., Callahan, J., Binns, D. S., Desmonts, C., Salaun, P. -Y., Hicks, R. J. & Aide, N. (2017). EORTC PET response criteria are more influenced by reconstruction inconsistencies than PERCIST but both benefit from the EARL harmonization program. EJNMMI PHYSICS, 4 (1), https://doi.org/10.1186/s40658-017-0185-4.
    Access Status
    Open Access
    URI
    http://hdl.handle.net/11343/259422
    DOI
    10.1186/s40658-017-0185-4
    Abstract
    BACKGROUND: This study evaluates the consistency of PET evaluation response criteria in solid tumours (PERCIST) and European Organisation for Research and Treatment of Cancer (EORTC) classification across different reconstruction algorithms and whether aligning standardized uptake values (SUVs) to the European Association of Nuclear Medicine acquisition (EANM)/EARL standards provides more consistent response classification. MATERIALS AND METHODS: Baseline (PET1) and response assessment (PET2) scans in 61 patients with non-small cell lung cancer were acquired in protocols compliant with the EANM guidelines and were reconstructed with point-spread function (PSF) or PSF + time-of-flight (TOF) reconstruction for optimal tumour detection and with a standardized ordered subset expectation maximization (OSEM) reconstruction known to fulfil EANM harmonizing standards. Patients were recruited in three centres. Following reconstruction, EQ.PET, a proprietary software solution was applied to the PSF ± TOF data (PSF ± TOF.EQ) to harmonize SUVs to the EANM standards. The impact of differing reconstructions on PERCIST and EORTC classification was evaluated using standardized uptake values corrected for lean body mass (SUL). RESULTS: Using OSEMPET1/OSEMPET2 (standard scenario), responders displayed a reduction of -57.5% ± 23.4 and -63.9% ± 22.4 for SULmax and SULpeak, respectively, while progressing tumours had an increase of +63.4% ± 26.5 and +60.7% ± 19.6 for SULmax and SULpeak respectively. The use of PSF ± TOF reconstruction impacted the classification of tumour response. For example, taking the OSEMPET1/PSF ± TOFPET2 scenario reduced the apparent reduction in SUL in responding tumours (-39.7% ± 31.3 and -55.5% ± 26.3 for SULmax and SULpeak, respectively) but increased the apparent increase in SUL in progressing tumours (+130.0% ± 50.7 and +91.1% ± 39.6 for SULmax and SULpeak, respectively). Consequently, variation in reconstruction methodology (PSF ± TOFPET1/OSEMPET2 or OSEM PET1/PSF ± TOFPET2) led, respectively, to 11/61 (18.0%) and 10/61 (16.4%) PERCIST classification discordances and to 17/61 (28.9%) and 19/61 (31.1%) EORTC classification discordances. An agreement was better for these scenarios with application of the propriety filter, with kappa values of 1.00 and 0.95 compared to 0.75 and 0.77 for PERCIST and kappa values of 0.93 and 0.95 compared to 0.61 and 0.55 for EORTC, respectively. CONCLUSION: PERCIST classification is less sensitive to reconstruction algorithm-dependent variability than EORTC classification but harmonizing SULs within the EARL program is equally effective with either.

    Export Reference in RIS Format     

    Endnote

    • Click on "Export Reference in RIS Format" and choose "open with... Endnote".

    Refworks

    • Click on "Export Reference in RIS Format". Login to Refworks, go to References => Import References


    Collections
    • Minerva Elements Records [52443]
    • Medicine and Radiology - Research Publications [3310]
    • Sir Peter MacCallum Department of Oncology - Research Publications [1029]
    Minerva AccessDepositing Your Work (for University of Melbourne Staff and Students)NewsFAQs

    BrowseCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects
    My AccountLoginRegister
    StatisticsMost Popular ItemsStatistics by CountryMost Popular Authors