dc.identifier.citationXu, S., McAnally, K. I., Xu, J., & Clark, G. M. (1993). Comparison of half-band and full-band electrodes for intracochlear electrical stimulation. Annals of Otology, Rhinology and Laryngology, May, 102(5), 363-367.en_US
dc.identifier.urihttp://hdl.handle.net/11343/27371
dc.descriptionThis is a publisher’s version of an article published in Annals of Otology, Rhinology & Laryngology published by Annals Publishing Company. This version is reproduced with permission from Annals Publishing Company. http://www.annals.com/en_US
dc.description.abstractIt was hypothesized that intracochlear stimulating electrodes oriented toward the modiolus would require a lower stimulus current to elicit a threshold neural response than longitudinal band electrodes. Electrically evoked auditory brain stem responses (EABRs) and electrode impedances were recorded with full-band and oriented half-band scala tympani electrodes in anesthetized, deafened cats. To elicit a threshold EABR, the stimulus current required for stimulation through half-band electrodes oriented toward the modiolus was not significantly different from the current required for stimulation through full-band electrodes. The impedances of full-band electrodes were significantly lower than those of half-band electrodes. Considering the significantly higher impedance and current density of half band electrodes in comparison to full-band electrodes, as well as the critical orientation of half-band electrodes during implantation, we believe that full-band electrodes have advantages over half-band electrodes for scala tympani implantation.en_US
dc.relation.ispartofScientific publications, vol. 6, 1991-1992 no. 430en_US
dc.subjectcochlear implanten_US
dc.subjectelectrically evoked auditory brain stem responseen_US
dc.subjecthalf-band electrodeen_US
dc.titleComparison of half-band and full-band electrodes for intracochlear electrical stimulationen_US
dc.typeJournal Articleen_US
melbourne.source.titleAnnals of Otology, Rhinology and Laryngologyen_US
melbourne.source.monthMayen_US
melbourne.source.volume102en_US
melbourne.source.issue5en_US
melbourne.source.pages363-367en_US