dc.contributor.authorClark, Graeme M.en_US
dc.contributor.authorPyman, Brian C.en_US
dc.contributor.authorBailey, Quentin R.en_US
dc.date.accessioned2014-05-21T21:57:33Z
dc.date.available2014-05-21T21:57:33Z
dc.date.issued1979en_US
dc.identifier.citationClark, G. M., Pyman, B. C., & Bailey, Q. R. (1979). The surgery for multiple-electrode cochlear implantations. Journal of Laryngology and Otology, 93(3), 215-223.en_US
dc.identifier.urihttp://hdl.handle.net/11343/28718
dc.descriptionThis is a publisher’s version of an article published in The Journal of Laryngology & Otology © 1979 Cambridge University Press. www.cambridge.org/en_US
dc.description.abstractThe multiple-electrode hearing prosthesis designed in the Departments of Otolaryngology and Electrical Engineering (UMDOLEE) at the University of Melbourne (Clark et al., 1977) has been miniaturized with hybrid circuitry so that, if design changes are necessary as a result of initial patient testing, they can be made at minimal cost. This results, however, in increased package dimensions which makes its placement and the design of the surgery more critical. This problem is increased by the fact that we have considered it important to be able to remove the package and replace it with another without disturbing the implanted electrode array, should the first receiver-stimulator fail or an improved design become available. This has meant the design of a special connector (Patrick, 1977; Clark et al., 1978) which adds to the dimensions of the implanted unit. In addition the placement of the coils for transmitting power and information has to be considered. Not only is it desirable to site the coils at a convenient location behind the ear to facilitate the placement and wearing of the external transmitter, but there should also be no relative movement between the coils and the electronic package. These design considerations have led to the sitting of the coils on top of the hermetically-sealed box, and further increased the height of the package. The dimensions of the package shown in Fig. 1 are length 42 mm, width 32 mm, height of connector 8.5 mm, height of receiver-stimulato unit 13 mm. The surgical considerations discussed are the result of a number of temporal bone and cadaver dissections, and the surgical implantation at The Royal Victorian Eye and Ear Hospital of the UMDOLEE unit in a specially-selected patient.en_US
dc.publisherCambridge University Pressen_US
dc.relation.ispartofScientific publications, vol.2, 1978-1983, no.83en_US
dc.subjectcochlear implanten_US
dc.subjecthearing prosthesisen_US
dc.subjectotolaryngologyen_US
dc.titleThe surgery for multiple-electrode cochlear implantationsen_US
dc.typeJournal Articleen_US
melbourne.source.titleJournal of Laryngology and Otologyen_US
melbourne.source.volume93en_US
melbourne.source.issue3en_US
melbourne.source.pages215-223en_US
melbourne.elementsidNA
melbourne.contributor.authorClark, Graeme
melbourne.contributor.authorPYMAN, BRIAN
melbourne.accessrightsOpen Access


Files in this item

Thumbnail

This item appears in the following Collection(s)